Performance of a dual-layer scanner for hybrid SPECT/CBCT

被引:11
作者
Dietze, Martijn M. A. [1 ,2 ,3 ]
Kunnen, Britt [1 ,2 ,3 ]
van der Velden, Sandra [1 ,2 ,3 ]
Steenbergen, J. H. Leo [1 ,2 ]
Koppert, Wilco J. C. [1 ,2 ]
Viergever, Max A. [2 ,3 ]
de Jong, Hugo W. A. M. [1 ,2 ,3 ]
机构
[1] Univ Utrecht, Radiol & Nucl Med, POB 85500, NL-3508 GA Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, POB 85500, NL-3508 GA Utrecht, Netherlands
[3] Univ Utrecht, Image Sci Inst, POB 85500, NL-3508 GA Utrecht, Netherlands
关键词
SPECT; CBCT; dual-layer; radioembolization; HEPATIC RADIOEMBOLIZATION;
D O I
10.1088/1361-6560/ab15f6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fluoroscopic procedures involving radionuclides would benefit from interventional nuclear imaging by obtaining real-time feedback on the activity distribution. We have previously proposed a dual-layer detector that offers such procedural guidance by simultaneous fluoroscopic and nuclear planar imaging. Acquisition of single photon computed tomography (SPECT) and cone beam computed tomography (CBCT) could provide additional information on the activity distribution. This study investigates the feasibility and the image quality of simultaneous SPECT/CBCT, by means of phantom experiments and simulations. Simulations were performed to study the obtained reconstruction quality for (i) clinical SPECT/CT, (ii) a dual-layer scanner configured with optimized hardware, and (iii) our (non-optimized) dual-layer prototype. Experiments on an image quality phantom and an anthropomorphic phantom (including extrahepatic depositions with volumes and activities close to the median values encountered in hepatic radioembolization) were performed with a clinical SPECT/CT scanner and with our dual-layer prototype. Nuclear images were visually and quantitatively evaluated by measuring the tumor/non-tumor (T/N) ratio and contrast-to-noise ratio (CNR). The simulations showed that the maximum obtained CNR was 38.8 +/- 0.8 for the clinical scanner, 30.2 +/- 0.9 for the optimized dual-layer scanner, and 20.8 +/- 0.4 for the prototype scanner. T/N ratio showed a similar decline. The phantom experiments showed that performing simultaneous SPECT/CBCT is feasible. The CNR obtained from the SPECT reconstruction of largest sphere in the image quality phantom was 43.1 for the clinical scanner and 28.6 for the developed prototype scanner. The anthropomorphic phantom showed that the extrahepatic depositions were detected with both scanners. A dual-layer detector is able to simultaneously acquire SPECT and CBCT. Both CNR and T/N ratio are worse than that of a clinical system, but the phantom experiments showed that extrahepatic depositions with volumes and activities close to the median values encountered in hepatic radioembolization could be distinguished.
引用
收藏
页数:12
相关论文
共 20 条
  • [1] [Anonymous], SYMB S T SYST SPEC
  • [2] Bailey D L, 2014, EUR J NUCL MED MOL I, V41
  • [3] An Evidence-Based Review of Quantitative SPECT Imaging and Potential Clinical Applications
    Bailey, Dale L.
    Willowson, Kathy P.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (01) : 83 - 89
  • [4] TIGRE: a MATLAB-GPU toolbox for CBCT image reconstruction
    Biguri, Ander
    Dosanjh, Manjit
    Hancock, Steven
    Soleimani, Manuchehr
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (05):
  • [5] 90Y Hepatic Radioembolization: An Update on Current Practice and Recent Developments
    Braat, Arthur J. A. T.
    Smits, Maarten L. J.
    Braat, Manon N. G. J. A.
    van den Hoven, Andor F.
    Prince, Jip F.
    de Jong, Hugo W. A. M.
    van den Bosch, Maurice A. A. J.
    Lam, Marnix G. E. H.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (07) : 1079 - 1087
  • [6] Acceleration of Monte Carlo SPECT simulation using convolution-based forced detection
    de Jong, HWAM
    Slijpen, ETP
    Beekman, FJ
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (01) : 58 - 64
  • [7] Dietze M M A, 2018, RSNA
  • [8] Fast quantitative reconstruction with focusing collimators for liver SPECT
    Dietze, Martijn M. A.
    van der Velden, Sandra
    Lam, Marnix G. E. H.
    Viergever, Max A.
    de Jong, Hugo W. A. M.
    [J]. EJNMMI PHYSICS, 2018, 5
  • [9] Same-day 90Y radioembolization: implementing a new treatment paradigm
    Gabr, Ahmed
    Kallini, Joseph Ralph
    Gates, Vanessa L.
    Hickey, Ryan
    Kulik, Laura
    Desai, Kush
    Thornburg, Bartley
    Marshall, Karen
    Salzig, Krystina
    Williams, Melissa
    del Castillo, Carlene
    Ganger, Daniel
    Hohlastos, Elias
    Baker, Talia
    Lewandowski, Robert J.
    Salem, Riad
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (13) : 2353 - 2359
  • [10] High Impact of Preferential Flow on 99mTc-MAA and 90Y-Loaded Microsphere Uptake Correlation
    Garin, Etienne
    Edeline, Julien
    Rolland, Yan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (11) : 1829 - 1830