Miniature Fiber-Optic Microflowmeter Based on Fabry-Perot Interferometer and Carbon Nanotubes

被引:1
|
作者
Liu, Weinan [1 ]
Pu, Shengli [1 ,2 ]
Duan, Simiao [1 ]
Hao, Zijian [1 ]
Zhang, Chencheng [1 ]
Wu, Qiang [3 ]
机构
[1] Univ Shanghai Sci & Technol, Coll Sci, Shanghai 200093, Peoples R China
[2] Univ Shanghai Sci & Technol, Shanghai Key Lab Modern Opt Syst, Shanghai 200093, Peoples R China
[3] Northumbria Univ, Dept Math Phys & Elect Engn, Newcastle upon Tyne NE1 8ST, England
基金
上海市自然科学基金; 中国国家自然科学基金;
关键词
Heating systems; Probes; Optical fiber sensors; Sensors; Temperature sensors; Temperature measurement; Optical surface waves; Fabry-Perot interferometer (FPI); fiber sensor; flowmeter; microfluid; SENSOR;
D O I
10.1109/TIM.2024.3449955
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel fiber-optic micro flowmeter based on a Fabry-Perot interferometer (FPI) and carbon nanotubes (CNTs) has been proposed and investigated. The proposed sensor integrates a Fabry-Perot cavity on the tip of a single-mode fiber (SMF), which is used as the sensing probe. The Fabry-Perot cavity comprises an SMF and a short hollow-core fiber (HCF) section. The HCF is partially filled with CNTs-doped polydimethylsiloxane (CNTs-PDMS), which emits heat obviously due to the excellent photothermal effect of CNTs-PDMS when illuminated by a laser. Under a certain flow velocity, the heat would be taken away by a microfluid as it flows over the surface of the Fabry-Perot cavity until the thermal equilibrium is established. Besides, due to the thermal expansion of CNTs-PDMS, the variation of heat will result in a change in the cavity length of FPI. By monitoring the dip wavelength variation of the sensor, the microflow rate will be determined. The experimentally measured maximum flow rate sensitivity is 68.52 nm/(mu L/s) in the 0-2 mu L/s range. The proposed sensing probe is ultracompact, and the footprint is extremely small (the length of the sensing part is only 126 mu m), which has potential in flow rate detection in narrow spaces and in vivo applications in biomedicine.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] A Fiber-Optic Sensor for Simultaneous Temperature and Pressure Measurements Based on a Fabry-Perot Interferometer and a Fiber Bragg Grating
    Fadeev, K. M.
    Larionov, D. D.
    Zhikina, L. A.
    Minkin, A. M.
    Shevtsov, D., I
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2020, 63 (04) : 543 - 546
  • [42] SiC thin-film Fabry-Perot interferometer for fiber-optic temperature sensor
    Cheng, L
    Steckl, AJ
    Scofield, J
    IEEE TRANSACTIONS ON ELECTRON DEVICES, 2003, 50 (10) : 2159 - 2164
  • [43] Influence of the Q point on the extrinsic Fabry-Perot interferometer fiber-optic sensors with a diaphragm
    Wang Wenhua
    Jiang Xinsheng
    SENSORS, MEASUREMENT AND INTELLIGENT MATERIALS, PTS 1-4, 2013, 303-306 : 78 - +
  • [44] Hybrid structured fiber-optic Fabry-Perot interferometer for simultaneous measurement of strain and temperature
    Zhou, Ai
    Qin, Boyang
    Zhu, Zheng
    Zhang, Yaxun
    Liu, Zhihai
    Yang, Jun
    Yuan, Libo
    OPTICS LETTERS, 2014, 39 (18) : 5267 - 5270
  • [45] Miniaturized Fiber-Optic Fabry-Perot Interferometer for Highly Sensitive Refractive Index Measurement
    M Deng and H Li are with the Key Laboratory of Optoelectronic Technology and Systems Chongqing University Chongqing China YJ Rao and T Zhu are with the Key Laboratory of Optoelectronic Technology and Systems Chongqing University Chongqing China and the Key Lab of Broadband Optical Fiber Transmission Communication Networks Technology Ministry of Education of China University of Electronic Science Technology of China Chengdu China
    Journal of Electronic Science and Technology of China, 2008, 6 (04) : 365 - 368
  • [46] ALD thin ZnO layer as an active medium in a fiber-optic Fabry-Perot interferometer
    Jedrzejewska-Szczerska, M.
    Wierzba, P.
    Abou Chaaya, A.
    Bechelany, M.
    Miele, P.
    Viter, R.
    Mazikowski, A.
    Karpienko, K.
    Wrobel, M.
    SENSORS AND ACTUATORS A-PHYSICAL, 2015, 221 : 88 - 94
  • [47] Miniaturized Fiber-Optic Fabry-Perot Interferometer for Highly Sensitive Refractive Index Measurement
    M. Deng and H. Li are with the Key Laboratory of Optoelectronic Technology and Systems
    Journal of Electronic Science and Technology, 2008, 6 (04) : 365 - 368
  • [48] Fiber-optic Fabry-Perot interferometer tip accelerometer fabricated by laser-micromachining
    Ran, Zengling
    Lu, En
    Rao, Yunjiang
    Ni, Min
    Peng, Fei
    Zeng, Dehong
    21ST INTERNATIONAL CONFERENCE ON OPTICAL FIBER SENSORS, 2011, 7753
  • [49] Miniaturized Fiber-Optic Fabry-Perot Interferometer for Highly Sensitive Refractive Index Measurement
    Deng, Ming
    Zhu, Tao
    Rao, Yun-Jiang
    Li, Hong
    APOS: 2008 1ST ASIA-PACIFIC OPTICAL FIBER SENSORS CONFERENCE, 2008, : 19 - 22