Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms

被引:1
作者
Fania, Claudio [1 ]
Giletto, Antonella [1 ]
Niero, Alice [1 ]
Palatini, Paolo [2 ]
机构
[1] Villa Maria Hosp, Dept Med & Surg, Via delle Melette,20, I-35138 Padua, Italy
[2] Univ Padua, Studium Patavinum, Dept Med, Padua, Italy
关键词
blood pressure; blood pressure measuring device validation; device validation; extra-large arms; obese people; MEDICAL INSTRUMENTATION/EUROPEAN SOCIETY; HYPERTENSION/INTERNATIONAL ORGANIZATION; UNIVERSAL STANDARD; CUFF; VALIDATION; ADVANCEMENT; ASSOCIATION;
D O I
10.1097/MBP.0000000000000696
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol. Methods The KOROT P3 Accurate was tested in 37 subjects with upper-arm circumference ranging from >42 to 53 cm using a mercury sphygmomanometer coupled to a 20 x 40 cm tronco-conical cuff as the reference standard. Results The mean BP difference between the device and the observers' reference measurements was 1.2 +/- 2.0 mmHg for systolic BP and 1.0 +/- 2.0 mmHg for diastolic BP. These data were in agreement with criterion 1 of the protocol standard requirements (<= 5 +/- 8 mmHg). Also criterion 2 was satisfied being the standard deviations +/- 1.7 mmHg for systolic BP and +/- 1.6 for diastolic BP, well below the maximum values required by the protocol (+/- 6.84/6.87 mmHg). Scatterplots of device-reference systolic and diastolic BP differences showed similar accuracy across the range of participants' BP, arm circumference and upper-arm slant angle. Conclusions These data show that the KOROT P3 Accurate monitor satisfied the ISO 81060-2:2018 standard requirements in a special population of people with extra-large arms ranging from >42 to 53 cm.
引用
收藏
页码:144 / 148
页数:5
相关论文
共 14 条
[1]  
International Organization for Standardization, ISO 81060-2:2018/AMD 1:2020
[2]  
International Organization for Standardization, 2018, INT ORG STANDARDIZAT
[3]   COMPARISON OF THE CONICAL CUFF AND THE STANDARD RECTANGULAR CUFFS [J].
MAXWELL, GF ;
PRUIJT, JFM ;
ARNTZENIUS, AC .
INTERNATIONAL JOURNAL OF EPIDEMIOLOGY, 1985, 14 (03) :468-472
[4]   Measurement of Blood Pressure in Humans A Scientific Statement From the American Heart Association [J].
Muntner, Paul ;
Shimbo, Daichi ;
Carey, Robert M. ;
Charleston, Jeanne B. ;
Gaillard, Trudy ;
Misra, Sanjay ;
Myers, Martin G. ;
Ogedegbe, Gbenga ;
Schwartz, Joseph E. ;
Townsend, Raymond R. ;
Urbina, Elaine M. ;
Viera, Anthony J. ;
White, William B. ;
Wright, Jackson T., Jr. .
HYPERTENSION, 2019, 73 (05) :E35-E66
[5]  
Ntineri A., 2023, J HYPERTENS, V41, P356, DOI DOI 10.1097/HJH.0000000000003341
[6]   Use of Anthropometric Indices to Identify Appropriate Cuff Shapes for Blood Pressure Measurement: Normative Data for Adults [J].
Palatini, Paolo ;
Fania, Claudio ;
Ermolao, Andrea ;
Battista, Francesca ;
Saladini, Francesca .
AMERICAN JOURNAL OF HYPERTENSION, 2022, 35 (06) :526-532
[7]   Criteria for a suitable reference cuff for validation studies of blood pressure measuring devices in people with arm circumference between 43 and 50 cm [J].
Palatini, Paolo ;
Fania, Claudio ;
Benetti, Elisabetta ;
Saladini, Francesca ;
Battista, Francesca ;
Ermolao, Andrea .
BLOOD PRESSURE MONITORING, 2023, 28 (01) :67-72
[8]   Recommendations for blood pressure measurement in large arms in research and clinical practice: position paper of the European society of hypertension working group on blood pressure monitoring and cardiovascular variability [J].
Palatini, Paolo ;
Asmar, Roland ;
O'Brien, Eoin ;
Padwal, Raj ;
Parati, Gianfranco ;
Sarkis, Josh ;
Stergiou, George .
JOURNAL OF HYPERTENSION, 2020, 38 (07) :1244-1250
[9]   Only troncoconical cuffs can provide accurate blood pressure measurements in people with severe obesity [J].
Palatini, Paolo ;
Benetti, Elisabetta ;
Fania, Claudio ;
Saladini, Francesca .
JOURNAL OF HYPERTENSION, 2019, 37 (01) :37-41
[10]   Cuff and Bladder: Overlooked Components of BP Measurement Devices in the Modern Era? [J].
Palatini, Paolo ;
Frick, Gerhard N. .
AMERICAN JOURNAL OF HYPERTENSION, 2012, 25 (02) :136-138