Development and experimental evaluation of hybrid K-edge/X-ray fluorescence densitometer for uranium solution measurement

被引:0
|
作者
Zhang, Yan [1 ,2 ]
Fu, Chun-Qing [1 ]
Qiu, Jun [1 ]
Qu, Jin-Hui [1 ]
Jia, Wenbao [2 ]
Cheng, Can [3 ]
Wang, Ren-Bo [1 ,4 ]
Tang, Bin [1 ]
机构
[1] East China Univ Technol, Jiangxi Prov Key Lab Nucl Phys & Technol, Nanchang 330013, Peoples R China
[2] Nanjing Univ Aeronaut & Astronaut, Dept Nucl Sci & Technol, Nanjing 211106, Peoples R China
[3] Nanjing Tech Univ, Coll Elect Engn & Control Sci, Nanjing 211816, Peoples R China
[4] Pan China DetectTechnol Co Ltd, Nanchang 330013, Peoples R China
关键词
Hybrid K-Edge/X-ray fluorescence; densitometer; Key parameters; Performance test; Uranium;
D O I
10.1016/j.radmeas.2024.107267
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The hybrid K-edge/X-ray fluorescence densitometer (HKED) is a combination of K-edge absorption technology (KED) and characteristic X-ray fluorescence (XRF), which has the advantages of direct, fast and non-destructive determination, and is an ideal non-destructive measurement technology for uranium and plutonium concentrations. In this paper, a new HKED was developed, primarily utilizing an X-ray tube from COMET, alongside high-purity germanium (HPGe) and cadmium telluride (CdTe) detectors from AMETEK ORTEC. This manuscript delves into several variables that influence measurement outcomes under predefined experimental conditions and operational prerequisites to pinpoint critical parameters. It was discerned that the adoption of a 160 kV high voltage setting markedly diminishes experimental interferences, while the beam current, optimally set at 2 mA, not only ensures a linear correlation with the count rate but also maximizes the effective count detected. The incorporation of a 2 cm fixed-length iron rod along the trajectory between the sample and the detector, complemented by an additional 3 mm external absorber before the KED detector, effectively mitigates direct X-ray exposure, thereby enhancing transmittance values to attainable extents. Subsequent to the determination of these pivotal parameters, validation of the HKED system's efficacy was conducted via performance evaluation tests on a laboratory-scale HKED setup. Measurements undertaken for both KED and XRF across an interval ranging from 300 to 3000 s fell within the 2 sigma boundary, affirming the system's stability. Repeated measurements of 50 g/L and 150 g/L uranium solutions yielded KED precision rates of 0.56% and 0.19%, respectively. Moreover, linear regression analyses linking transmittance, characteristic X-ray fluorescence, and uranium concentrations across a spectrum of 0-150 g/L underscored the laboratory HKED instrument's robust analytical capabilities. Notably, the relative discrepancy between theoretical predictions and empirical findings for the 150 g/L uranium sample was minimized to a commendable 0.58%.
引用
收藏
页数:8
相关论文
共 13 条
  • [1] A quantitative Monte Carlo modelling of the uranium and plutonium X-ray fluorescence (XRF) response from a hybrid K-edge/K-XRF densitometer
    Berlizoy, A. N.
    Sharikov, D. A.
    Ottmar, H.
    Eberle, H.
    Galy, J.
    Luetzenkirchen, K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 615 (01) : 127 - 135
  • [2] System optimization study of hybrid K-edge/XRF densitometer for uranium-plutonium solution measurement
    Zhang, Yan
    Fu, Chun-Qing
    Qiu, Jun
    Qu, Jin-Hui
    Yang, Yu-Chao
    Tang, Bin
    Wang, Ren-Bo
    APPLIED RADIATION AND ISOTOPES, 2024, 210
  • [3] Development and validation of hybrid K-edge/K-XRF densitometer for assay of nuclear fuel from reprocessing plants
    Bootharajan, M.
    Senthilvadivu, R.
    Sundararajan, K.
    Kumar, R.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2020, 324 (02) : 623 - 633
  • [4] Development and validation of hybrid K-edge/K-XRF densitometer for assay of nuclear fuel from reprocessing plants
    M. Bootharajan
    R. Senthilvadivu
    K. Sundararajan
    R. Kumar
    Journal of Radioanalytical and Nuclear Chemistry, 2020, 324 : 623 - 633
  • [5] Identification of Uranyl Minerals Using Oxygen K-Edge X-Ray Absorption Spectroscopy
    Ward, Jesse D.
    Bowden, Mark
    Resch, C. Tom
    Smith, Steven
    McNamara, Bruce K.
    Buck, Edgar C.
    Eiden, Gregory C.
    Duffin, Andrew M.
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2016, 40 (01) : 135 - 148
  • [6] Energy dispersive X-ray fluorescence determination of cadmium in uranium matrix using Cd Kα line excited by continuum
    Dhara, Sangita
    Misra, N. L.
    Aggarwal, S. K.
    Venugopal, V.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2010, 65 (06) : 461 - 465
  • [7] Oxygen K-Edge X-ray Absorption Spectra of ThO2 and CeO2: Experiment, Interpretation, and Structural Effects
    Amidani, Lucia
    Dumas, Thomas
    Shuh, David K.
    Butorin, Sergei M.
    Sahle, Christoph J.
    Longo, Alessandro
    Kvashnina, Kristina O.
    JOURNAL OF PHYSICAL CHEMISTRY C, 2023, 127 (06) : 3077 - 3084
  • [8] Energy dispersive X-ray fluorescence determination of uranium in different uranates using Rh Kα scattered peaks for matrix correction
    Kumar, Sanjay
    Dhara, Sangita
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2022, 193
  • [9] Development of Methods to Evaluate Several Levels of Uranium Concentrations in Drainage Water Using Total Reflection X-Ray Fluorescence Technique
    Matsuyama, Tsugufumi
    Izumoto, Yukie
    Ishii, Kota
    Sakai, Yasuhiro
    Yoshii, Hiroshi
    FRONTIERS IN CHEMISTRY, 2019, 7
  • [10] Evaluation of total-reflection X-ray fluorescence measurements for the prospective determination of uranium in the decommissioning of the Fukushima Daiichi Nuclear Power Plant reactor
    Yoshii, Hiroshi
    Izumoto, Yukie
    Matsuyama, Tsugufumi
    Ishii, Kota
    Sakai, Yasuhiro
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 148 : 183 - 187