Hybrid Deep Learning and Model-Based Needle Shape Prediction

被引:0
作者
Lezcano, Dimitri A. [1 ]
Zhetpissov, Yernar [2 ]
Bernardes, Mariana C. [3 ]
Moreira, Pedro [3 ]
Tokuda, Junichi [3 ]
Kim, Jin Seob [1 ]
Iordachita, Iulian I. [1 ]
机构
[1] Johns Hopkins Univ, Mech Engn Dept, Baltimore, MD 21201 USA
[2] Johns Hopkins Univ, Mech Engn Dept, Baltimore, MD 21201 USA
[3] Harvard Med Sch, Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
基金
美国国家卫生研究院;
关键词
Needles; Shape; Predictive models; Sensors; Three-dimensional displays; Deep learning; Trajectory; flexible needle; machine learning; medical device; model-based; shape prediction; STEERABLE NEEDLE; PROSTATE BIOPSY; INSERTION; DEFLECTION; TRACKING; TISSUE; ROBOT;
D O I
10.1109/JSEN.2024.3386120
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Needle insertion using flexible bevel-tip needles is a common minimally invasive surgical technique for prostate cancer interventions. Flexible, asymmetric bevel-tip needles enable physicians for complex needle steering techniques to avoid sensitive anatomical structures during needle insertion. For accurate placement of the needle, predicting the trajectory of these needles intraoperatively would greatly reduce the need for frequently needle reinsertions, thus improving patient comfort and positive outcomes. However, predicting the trajectory of the needle during insertion is a complex task that has yet to be solved due to random needle-tissue interactions. In this article, we present and validate, for the first time, a hybrid deep learning and model-based approach to handle the intraoperative needle shape prediction problem through leveraging a validated Lie-group theoretic model for needle shape representation. Furthermore, we present a novel self-supervised learning (SSL) and method in conjunction with the Lie-group shape model for training these networks in the absence of data, enabling further refinement of these networks with transfer learning (TL). Needle shape prediction was performed in single-layer and double-layer homogeneous phantom tissue for C- and S-shaped needle insertions. Our method demonstrates an average root-mean-square prediction error of 1.03 mm over a dataset containing approximately 3000 prediction samples with the maximum prediction steps of 110 mm.
引用
收藏
页码:18359 / 18371
页数:13
相关论文
共 69 条
[51]   A data-driven soft sensor for needle deflection in heterogeneous tissue using just-in-time modelling [J].
Rossa, Carlos ;
Lehmann, Thomas ;
Sloboda, Ronald ;
Usmani, Nawaid ;
Tavakoli, Mahdi .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2017, 55 (08) :1401-1414
[52]  
Secoli R, 2013, IEEE INT CONF ROBOT, P5831, DOI 10.1109/ICRA.2013.6631416
[53]   Learning to Detect Collisions for Continuum Manipulators Without a Prior Model [J].
Sefati, Shahriar ;
Sefati, Shahin ;
Iordachita, Iulian ;
Taylor, Russell H. ;
Armand, Mehran .
MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2019, PT V, 2019, 11768 :182-190
[54]   FBG-Based Position Estimation of Highly Deformable Continuum Manipulators: Model-Dependent vs. Data-Driven Approaches [J].
Sefati, Shahriar ;
Hegeman, Rachel ;
Alambeigi, Farshid ;
Iordachita, Iulian ;
Armand, Mehran .
2019 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2019,
[55]  
Shahriari N, 2016, IEEE INT CONF ROBOT, P4443, DOI 10.1109/ICRA.2016.7487644
[56]  
Su H, 2011, IEEE INT CONF ROBOT, P1583
[57]   Robot-assisted flexible needle insertion using universal distributional deep reinforcement learning [J].
Tan, Xiaoyu ;
Lee, Yonggu ;
Chng, Chin-Boon ;
Lim, Kah-Bin ;
Chui, Chee-Kong .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2020, 15 (02) :341-349
[58]  
Torrey L., 2010, Handbook of Research on Machine Learning Applications and Trends: Algorithms, Methods, and Techniques, P242, DOI [DOI 10.4018/978-1-60566-766-9.CH011, 10.4018/978-1-60566-766-9.ch011]
[59]   Design of an actively controlled steerable needle with tendon actuation and FBG-based shape sensing [J].
van de Berg, Nick J. ;
Dankelman, Jenny ;
van den Dobbelsteen, John J. .
MEDICAL ENGINEERING & PHYSICS, 2015, 37 (06) :617-622
[60]  
Waine M., 2016, J. Med. Robot. Res, V1, P1