Relationships between Size-specific Dose Estimate and Signal to Noise Ratio under Chest CT Examinations with Tube Current Modulation

被引:0
|
作者
Qin, Tian [1 ]
Wang, Jing [1 ]
Wang, Mengting [1 ]
Gu, Ye [1 ]
Xie, Zongyu [2 ]
Liang, Baohui [1 ]
机构
[1] Bengbu Med Coll, Sch Med Imaging, Bengbu, Peoples R China
[2] Bengbu Med Coll, Affiliated Hosp 1, 287 Changhuai Rd, Bengbu 233004, Anhui, Peoples R China
关键词
Chest computed tomography (CT); Size-specific dose estimate; Volume CT dose index; Automatic tube current modulation; Signal-to-noise ratio; ROI; CANCER-RISK ESTIMATION; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; SPECTRUM; HEAD;
D O I
10.2174/0115734056267251231121064259
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Exploring the relationship between the signal-to-noise ratio (SNR) of organs and size-specific dose estimate (SSDE) in tube current modulation (TCM) chest CT examination.Methods: Forty patients who received TCM chest CT scanning were retrospectively collected and divided into four groups according to the tube voltage and sexes. We chose to set up the region of interest (ROI) at the tracheal bifurcation and its upper and lower parts in slice images of the heart, aorta, lungs, paracranial muscles, and female breast, and the SNR of each organ was calculated. We also calculated the corresponding axial volume CT dose index (CTDIvolz) and axial size-specific dose estimate (SSDEz).Results: The correlation analysis showed that the correlation between the SNR of the slice images of most organs and SSDEz was more significant than 0.8, and that between the SNR and CTDIvol was more significant than 0.7. The simple linear regression analysis results showed that when the sex is the same, the SNR of the same organ at 100kVp was higher than 120kVp, except for the lung. In multiple regression analysis, the result indicated that the determination coefficients of the SNR and SSDEz of the four groups were 0.934, 0.971, 0.905, and 0.709, respectively.Conclusion: In chest CT examinations with TCM, the correlation between the SNR of each organ in slice images and SSDEz was better than that of CTDIvolz. And when the SSDEz was the same, the SNR at 100 kVp was better than that at 120 kVp.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Multi-slice CT examinations of adult patients at Sudanese hospitals: radiation exposure based on size-specific dose estimates (SSDE)
    Bashier, Einas H.
    Suliman, I. I.
    RADIOLOGIA MEDICA, 2018, 123 (06): : 424 - 431
  • [42] The impact of vertical off-centering on image noise and breast dose in chest CT with organ-based tube current modulation: A phantom study
    Makela, Teemu
    Kortesniemi, Mika
    Kaasalainen, Touko
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 : 153 - 163
  • [43] The impact of vertical off-centering on image noise and breast dose in chest CT with organ-based tube current modulation: A phantom study
    Makela, Teemu
    Kortesniemi, Mika
    Kaasalainen, Touko
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 : 153 - 163
  • [44] An Age-Based Size-Specific Dose Estimate for Pediatric Computed Tomography Head Examinations Performed at Songklanagarind Hospital, Thailand, from 2017 to 2019
    Poosiri, Saowapark
    Chuboonlap, Kanokkwan
    Kaewlaied, Nuttita
    APPLIED SCIENCES-BASEL, 2024, 14 (17):
  • [45] Influence of 320-detector-row volume scanning and AAPM report 111 CT dosimetry metrics on size-specific dose estimate: a Monte Carlo study
    Haba, Tomonobu
    Koyama, Shuji
    Kinomura, Yutaka
    Ida, Yoshihiro
    Kobayashi, Masanao
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (03) : 697 - 703
  • [46] Differences in behavior of tube current modulation techniques for thoracic CT examinations between male and female anthropomorphic phantoms
    Matsubara K.
    Lin P.-J.P.
    Fukuda A.
    Koshida K.
    Radiological Physics and Technology, 2014, 7 (2) : 316 - 328
  • [47] A SIMPLE METHOD FOR CALIBRATING PIXEL VALUES OF THE CT LOCALIZER RADIOGRAPH FOR CALCULATING WATER-EQUIVALENT DIAMETER AND SIZE-SPECIFIC DOSE ESTIMATE
    Anam, Choirul
    Fujibuchi, Toshioh
    Toyoda, Takatoshi
    Sato, Naoki
    Haryanto, Freddy
    Widita, Rena
    Arif, Idam
    Dougherty, Geoff
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (02) : 158 - 168
  • [48] VALIDATION OF MONTE CARLO DOSE CALCULATION FOR PAEDIATRIC CT EXAMINATIONS USING TUBE CURRENT MODULATION BASED ON IN-PHANTOM DOSIMETRY
    Fujii, K.
    Nomura, K.
    Muramatsu, Y.
    Obara, S.
    Goto, T.
    Akahane, K.
    Ota, H.
    Tsukagoshi, S.
    Kusumoto, M.
    RADIATION PROTECTION DOSIMETRY, 2018, 182 (04) : 508 - 517
  • [49] Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT? Protocols dose comparisons
    Lopez-Rendon, Xochitl
    Zhang, Guozhi
    Coudyzer, Walter
    Develter, Wim
    Bosmans, Hilde
    Zanca, Federica
    EUROPEAN RADIOLOGY, 2017, 27 (11) : 4490 - 4497
  • [50] Local diagnostic reference level based on size-specific dose estimates: case of chest-abdomen-pelvis CT examination in a Moroccan hospital
    Ouaggou, Ibrahim Ait
    Bakari, Dikra
    Zerfaoui, Mustapha
    Rrhioua, Abdeslem
    Oulhouq, Yassine
    Benmessaoud, Mounir
    Ouhadda, Hassan
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2024,