Relationships between Size-specific Dose Estimate and Signal to Noise Ratio under Chest CT Examinations with Tube Current Modulation

被引:0
作者
Qin, Tian [1 ]
Wang, Jing [1 ]
Wang, Mengting [1 ]
Gu, Ye [1 ]
Xie, Zongyu [2 ]
Liang, Baohui [1 ]
机构
[1] Bengbu Med Coll, Sch Med Imaging, Bengbu, Peoples R China
[2] Bengbu Med Coll, Affiliated Hosp 1, 287 Changhuai Rd, Bengbu 233004, Anhui, Peoples R China
关键词
Chest computed tomography (CT); Size-specific dose estimate; Volume CT dose index; Automatic tube current modulation; Signal-to-noise ratio; ROI; CANCER-RISK ESTIMATION; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; SPECTRUM; HEAD;
D O I
10.2174/0115734056267251231121064259
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Exploring the relationship between the signal-to-noise ratio (SNR) of organs and size-specific dose estimate (SSDE) in tube current modulation (TCM) chest CT examination.Methods: Forty patients who received TCM chest CT scanning were retrospectively collected and divided into four groups according to the tube voltage and sexes. We chose to set up the region of interest (ROI) at the tracheal bifurcation and its upper and lower parts in slice images of the heart, aorta, lungs, paracranial muscles, and female breast, and the SNR of each organ was calculated. We also calculated the corresponding axial volume CT dose index (CTDIvolz) and axial size-specific dose estimate (SSDEz).Results: The correlation analysis showed that the correlation between the SNR of the slice images of most organs and SSDEz was more significant than 0.8, and that between the SNR and CTDIvol was more significant than 0.7. The simple linear regression analysis results showed that when the sex is the same, the SNR of the same organ at 100kVp was higher than 120kVp, except for the lung. In multiple regression analysis, the result indicated that the determination coefficients of the SNR and SSDEz of the four groups were 0.934, 0.971, 0.905, and 0.709, respectively.Conclusion: In chest CT examinations with TCM, the correlation between the SNR of each organ in slice images and SSDEz was better than that of CTDIvolz. And when the SSDEz was the same, the SNR at 100 kVp was better than that at 120 kVp.
引用
收藏
页数:10
相关论文
共 28 条
[21]   A METHOD FOR DESCRIBING THE DOSES DELIVERED BY TRANSMISSION-X-RAY COMPUTED-TOMOGRAPHY [J].
SHOPE, TB ;
GAGNE, RM ;
JOHNSON, GC .
MEDICAL PHYSICS, 1981, 8 (04) :488-495
[22]  
Shu Hongmin., 2019, Acta Universitatis Medicinalis Anhui, V54, P1489
[23]   RADIATION DOSE AND IMAGE QUALITY IN PEDIATRIC HEAD CT [J].
Spampinato, Maria Vittoria ;
Stalcup, Seth ;
Matheus, Maria Gisele ;
Byington, Kathleen ;
Tyler, Michael ;
Bickley, Stetson ;
Tipnis, Sameer .
RADIATION PROTECTION DOSIMETRY, 2018, 182 (03) :310-316
[24]   CAN THE SIZE-SPECIFIC DOSE ESTIMATE BE DERIVED FROM THE BODY MASS INDEX? A FEASIBILITY STUDY [J].
Steiniger, Beatrice ;
Klippel, Chris ;
Teichgraeber, Ulf ;
Reichenbach, Juergen R. ;
Fiebich, Martin .
RADIATION PROTECTION DOSIMETRY, 2022, 198 (06) :325-333
[25]  
Strauss K., 2014, Size Specific Dose Estimates in Pediatric and Adult Body CT Examinations, V2011
[26]   Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study [J].
Tang, Kun ;
Wang, Ling ;
Li, Rui ;
Lin, Jie ;
Zheng, Xiangwu ;
Cao, Guoquan .
JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2012,
[27]  
[袁子龙 Yuan Zilong], 2016, [中华放射医学与防护杂志, Chinese Journal of Radiological Medicine and Protection], V36, P852
[28]   Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography [J].
Yurt, Aysegul ;
Ozsoykal, Ismail ;
Obuz, Funda .
MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2019, 28 (03) :96-103