Relationships between Size-specific Dose Estimate and Signal to Noise Ratio under Chest CT Examinations with Tube Current Modulation

被引:0
作者
Qin, Tian [1 ]
Wang, Jing [1 ]
Wang, Mengting [1 ]
Gu, Ye [1 ]
Xie, Zongyu [2 ]
Liang, Baohui [1 ]
机构
[1] Bengbu Med Coll, Sch Med Imaging, Bengbu, Peoples R China
[2] Bengbu Med Coll, Affiliated Hosp 1, 287 Changhuai Rd, Bengbu 233004, Anhui, Peoples R China
关键词
Chest computed tomography (CT); Size-specific dose estimate; Volume CT dose index; Automatic tube current modulation; Signal-to-noise ratio; ROI; CANCER-RISK ESTIMATION; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; SPECTRUM; HEAD;
D O I
10.2174/0115734056267251231121064259
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Exploring the relationship between the signal-to-noise ratio (SNR) of organs and size-specific dose estimate (SSDE) in tube current modulation (TCM) chest CT examination.Methods: Forty patients who received TCM chest CT scanning were retrospectively collected and divided into four groups according to the tube voltage and sexes. We chose to set up the region of interest (ROI) at the tracheal bifurcation and its upper and lower parts in slice images of the heart, aorta, lungs, paracranial muscles, and female breast, and the SNR of each organ was calculated. We also calculated the corresponding axial volume CT dose index (CTDIvolz) and axial size-specific dose estimate (SSDEz).Results: The correlation analysis showed that the correlation between the SNR of the slice images of most organs and SSDEz was more significant than 0.8, and that between the SNR and CTDIvol was more significant than 0.7. The simple linear regression analysis results showed that when the sex is the same, the SNR of the same organ at 100kVp was higher than 120kVp, except for the lung. In multiple regression analysis, the result indicated that the determination coefficients of the SNR and SSDEz of the four groups were 0.934, 0.971, 0.905, and 0.709, respectively.Conclusion: In chest CT examinations with TCM, the correlation between the SNR of each organ in slice images and SSDEz was better than that of CTDIvolz. And when the SSDEz was the same, the SNR at 100 kVp was better than that at 120 kVp.
引用
收藏
页数:10
相关论文
共 28 条
[1]   Patient size matters: Effect of tube current modulation on size-specific dose estimates (SSDE) and image quality in low-dose lung cancer screening CT [J].
Barreto, Izabella ;
Verma, Nupur ;
Quails, Nathan ;
Olguin, Catherine ;
Correa, Nathalie ;
Mohammed, Tan-Lucien .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (04) :87-94
[2]  
Boone J., 2011, SIZE SPECIFIC DOSE E, DOI [10.37206/143, DOI 10.37206/143]
[3]   Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility [J].
Braun, Franziska M. ;
Johnson, Thorsten R. C. ;
Sommer, Wieland H. ;
Thierfelder, Kolja M. ;
Meinel, Felix G. .
EUROPEAN RADIOLOGY, 2015, 25 (06) :1598-1606
[4]  
Burton C., 2017, J. Appl. Clin. Med. Phys, P19
[5]   Gender based lung cancer risks for symptomatic coronary artery disease patients undergone cardiac CT [J].
Dalah, Entesar Zawam ;
Obaideen, Abdulmunhem ;
Anam, Sabaa ;
Alzimami, Khalid ;
Jambi, Layal Khalid ;
Bradley, David A. .
PLOS ONE, 2022, 17 (04)
[6]   CT Dosimetry What Has Been Achieved and What Remains to Be Done [J].
Damilakis, John .
INVESTIGATIVE RADIOLOGY, 2021, 56 (01) :62-68
[7]   The role of Size-Specific Dose Estimate (SSDE) in patient-specific organ dose and cancer risk estimation in paediatric chest and abdominopelvic CT examinations [J].
Franck, Caro ;
Vandevoorde, Charlot ;
Goethals, Ingeborg ;
Smeets, Peter ;
Achten, Eric ;
Verstraete, Koenraad ;
Thierens, Hubert ;
Bacher, Klaus .
EUROPEAN RADIOLOGY, 2016, 26 (08) :2646-2655
[8]   Automatic exposure control at MDCT based on the contrast-to-noise ratio: Theoretical background and phantom study [J].
Funama, Yoshinori ;
Sugaya, Yoshiaki ;
Miyazaki, Osamu ;
Utsunomiya, Daisuke ;
Yamashita, Yasuyuki ;
Awai, Kazuo .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (01) :39-47
[9]   Pulmonary blood volume imaging with dual-energy computed tomography: Spectrum of findings [J].
Hagspiel, K. D. ;
Flors, L. ;
Housseini, A. M. ;
Phull, A. ;
Ahmad, E. Ali ;
Bozlar, U. ;
Norton, P. T. ;
Bonatti, H. J. R. .
CLINICAL RADIOLOGY, 2012, 67 (01) :69-77
[10]   Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size [J].
Kawashima, Hiroki ;
Ichikawa, Katsuhiro ;
Hanaoka, Shinsuke ;
Matsubara, Kosuke ;
Takata, Tadanori .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (04) :246-251