Performance of a Novel Continuous Glucose Monitoring Device in People With Diabetes

被引:3
|
作者
Mader, Julia K. [1 ]
Waldenmaier, Delia [2 ]
Mueller-Hoffmann, Wiebke [3 ]
Mueller, Katrin [3 ]
Angstmann, Michael [3 ]
Vogt, Gerhard [3 ]
Rieger, Cosima C. [3 ]
Eichenlaub, Manuel [2 ]
Forst, Thomas [4 ]
Freckmann, Guido [2 ]
机构
[1] Med Univ Graz, Dept Internal Med, Div Endocrinol & Diabetol, Graz, Austria
[2] Univ Ulm, Inst Diabet Technol Forsch & Entwicklungsgesell mb, Lise Meitner Str 8-2, D-89081 Ulm, Germany
[3] Roche Diabet Care GmbH, Mannheim, Germany
[4] CRS Clin Res Serv Mannheim GmbH, Mannheim, Germany
来源
关键词
accuracy; continuous glucose monitoring; MARD; performance;
D O I
10.1177/19322968241267774
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: In this multicenter study, performance of a novel continuous glucose monitoring (CGM) system was evaluated. Methods: Adult participants with diabetes were included in the study. They each wore three sensors of the CGM system on the upper arms for up to 14 days. During four in-clinic visits, frequent comparison measurements with capillary blood glucose (BG) samples were performed. The primary endpoint was the 20/20 agreement rate (AR): the percentage of CGM readings within +/- 20 mg/dL (at BG values <100 mg/dL) or +/- 20% (at BG values >= 100 mg/dL) of the comparator. Further evaluations included mean absolute relative difference (MARD) and 20/20 AR in different BG ranges and across the wear time. Results: Data from 48 participants and 139 sensors were analyzed. During in-clinic sessions the 20/20 AR was 90.5% and the MARD was 9.2%. For BG ranges <70, 70-180, and >180 mg/dL, 20/20 AR was 94.3%, 89.0%, and 92.5%, respectively. At the beginning, middle, and end of sensor wear time, 20/20 AR was 92.8%, 91.5%, and 85.9%, respectively. The 14-day survival probability was 82.4%. Pain and bleeding after sensor insertion were within the expected range. Based on the study outcome, the use of the device is regarded as safe. Conclusions: The system showed a good performance compared to capillary BG measurements. This level of accuracy could be shown over the entire measurement range, especially in the low glycemic range, and the whole wear time of the sensors. The results of this study are supporting a non-adjunctive use of the device.
引用
收藏
页码:1044 / 1051
页数:8
相关论文
共 50 条
  • [1] Pregnancy outcomes in people with diabetes using continuous glucose monitoring
    Bitar, Ghamar
    Cornthwaite, John
    Cornthwaite, Joy A. Ashby
    Nazeer, Sarah
    Ghafir, Danna
    Sadek, Sandra
    Ghorayeb, Tala
    Daye, Nahla
    Chauhan, Suneet P.
    Sibai, Baha M.
    Bartal, Michal Fishel
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2023, 228 (01) : S312 - S313
  • [2] Performance of a non-invasive glucose monitoring device for people with prediabetes and type 2 diabetes
    Bahartan, K.
    Gal, A.
    Naidis, E.
    Rozner, A.
    Drexler, A.
    Horman, K.
    Lin, T.
    DIABETOLOGIA, 2017, 60 : S321 - S321
  • [3] Performance of Sensors for Continuous Glucose Monitoring (CGM) under Extreme Sports Conditions in People with Diabetes
    Delgado, Elias
    Fokkert, Marion M. J.
    Van Dijk, P. R.
    Edens, Mireille A.
    Diez, Alberto, Sr.
    Slingerland, R. J.
    Bilo, Henk
    DIABETES, 2019, 68
  • [4] Clinical performance and reliability of a microdialysis continuous glucose monitoring device
    Freckmann, Guido
    Jendrike, Nina
    Heinemann, Lutz
    Christiansen, Jens Sandahl
    Nielsen, Jannik Kruse
    Amann-Zalan, Ildiko
    Ocvirk, Gregor
    DIABETES, 2008, 57 : A119 - A119
  • [5] Continuous glucose monitoring in people with diabetes: the randomized controlled Glucose Level Awareness in Diabetes Study (GLADIS)
    New, J. P.
    Ajjan, R.
    Pfeiffer, A. F. H.
    Freckmann, G.
    DIABETIC MEDICINE, 2015, 32 (05) : 609 - 617
  • [6] Putting Continuous Glucose Monitoring to Work for People With Type 1 Diabetes
    Hood, Korey K.
    DiMeglio, Linda A.
    Riddle, Matthew C.
    DIABETES CARE, 2020, 43 (01) : 19 - 21
  • [7] Meal responses in people without diabetes using continuous glucose monitoring
    Waldenmaier, D.
    Buck, S.
    Pleus, S.
    Beltzer, A.
    Schauer, S.
    Link, M.
    Zschornack, E.
    Jendrike, N.
    Haug, C.
    Freckmann, G.
    DIABETES STOFFWECHSEL UND HERZ, 2023, 32 (02): : 61 - 68
  • [8] Continuous glucose monitoring profiles in people living with HIV without diabetes
    Daultrey, H.
    Levett, T.
    Wright, J.
    Davies, K.
    Oliver, N.
    Chakera, A. J.
    DIABETIC MEDICINE, 2020, 37 : 56 - 56
  • [9] COST EFFECTIVENESS OF CONTINUOUS GLUCOSE MONITORING IN PEOPLE WITH TYPE 1 DIABETES
    Ljaljic, A.
    Yoels, B.
    Villa, Zapata L.
    VALUE IN HEALTH, 2023, 26 (06) : S101 - S101
  • [10] Continuous glucose monitoring in diabetes
    Moe, Samantha S.
    Kolber, Michael R.
    Falk, Jamie
    CANADIAN FAMILY PHYSICIAN, 2023, 69 (06) : 406 - 406