Comparability analysis of the HNT-1P Huvitz non contact tonometer for the measurement of intraocular pressure

被引:0
作者
Montoya, Alicia Carceles [1 ]
Lucas, Carlos Enrique Monera [1 ,2 ]
Jaimes, Claudia Patricia Tarazona [1 ,2 ]
Valero, Daniel Romero [1 ,2 ]
Martinez, Alejandro Moya [2 ,3 ]
Toldos, Jose Juan Martinez [1 ]
机构
[1] Gen Univ Hosp Elche, Ophthalmol Dept, C/Camino Almazara,11, Alicante 03203, Spain
[2] Miguel Hernandez Univ Elche, Alicante, Spain
[3] Gen Univ Hosp Elche, Biostat, FISABIO, Alicante, Spain
关键词
Glaucoma; non-contact tonometer; intraocular pressure; goldmann applanation tonometer; DYNAMIC CONTOUR TONOMETRY; REBOUND TONOMETRY; APPLANATION;
D O I
10.1177/11206721241263624
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the diagnostic validity of the HNT-1P non-contact tonometer (Huvitz) as a tool for accurately measuring intraocular pressure (IOP) in patients with healthy eyes, compared to the Goldmann applanation tonometer (GAT), which is the Gold Standard method for measurement of IOP. Methods Observational, descriptive, transversal study using 148 eyes of 74 healthy patients without a diagnosis of glaucoma or other ophthalmological diseases. Three measurements of IOP were taken in each eye, using three tonometers: HNT-1P, ICR100, and GAT. The median IOP (quartiles) and mean IOP (SD) its statistical significance were calculated, and comparisons were made between the mean and median IOP values found in three groups: GAT-HNT, GAT-ICR, and HNT-ICR. The difference in mean and median IOP was analyzed in each of the three study groups, and its statistical significance and concordance correlation coefficient (CCC) were calculated. Results The median IOP with HNT-1P was statistically significantly lower than the median IOP with GAT, (1.1 mmHg, p < 0.001). The median IOP with HNT-1P was also lower than the median IOP with ICR100. As an additional result, the median IOP with GAT was lower than the median IOP with ICR. The CCC was moderate for HNT-ICR (0.72) and low for GAT-HNT and GAT-ICR (0.43 and 0.38, respectively). Conclusions HNT-1P (Huvitz) provides statistically significantly lower IOP values than those obtained with GAT. HNT-1P could be used for screening of ocular hypertension in postoperative patients. The IOP measurement obtained with HNT-1P should be confirmed with GAT. HNT-1P yields lower IOP values than those obtained with ICR.
引用
收藏
页码:544 / 550
页数:7
相关论文
共 32 条
  • [21] Comparison of Icare HOME and non-contact tonometer in intraocular pressure measurement in the early stage after ICL V4c implantation
    Ye, Yuhao
    Yang, Fangwen
    Ding, Lan
    Xia, Fei
    Niu, Lingling
    Wang, Xiaoying
    Zhao, Jing
    Zhou, Xingtao
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2022, 32 (06) : 3303 - 3311
  • [22] The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
    Jethani, Jitendra
    Dave, Paaraj
    Jethani, Monica
    Desai, Yogesh
    Patel, Purvi
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2016, 30 (01) : 25 - 28
  • [23] Influence of Corneal Biomechanical Properties on Intraocular Pressure Differences Between a non-contact ORA Tonometer and the Goldmann Applanation Tonometer in primary congenital glacomatous children
    Zareei, Athar
    Razeghinejad, Mohamad Reza
    Salouti, Ramin
    [J]. JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2018, 6 (03): : 109 - 113
  • [24] Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry
    Luebke, Jan
    Bryniok, L.
    Neuburger, M.
    Jordan, J. F.
    Boehringer, D.
    Reinhard, T.
    Wecker, T.
    Anton, A.
    [J]. INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (11) : 2517 - 2521
  • [25] Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry
    Jan Luebke
    L. Bryniok
    M. Neuburger
    J. F. Jordan
    D. Boehringer
    T. Reinhard
    T. Wecker
    A. Anton
    [J]. International Ophthalmology, 2019, 39 : 2517 - 2521
  • [26] Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices
    Michele Lanza
    Michele Rinaldi
    Ugo Antonello Gironi Carnevale
    Silvio di Staso
    Mario Bifani Sconocchia
    Ciro Costagliola
    [J]. BMC Ophthalmology, 18
  • [27] The impact of silicone hydrogel contact lenses on the measurement of intraocular pressure using non-contact tonometry
    Pesic, Snezana
    Jovanovic, Svetlana
    Mitrasevic, Milos
    Vuletic, Biljana
    Jovanovic, Milena
    Jovanovic, Zorica
    [J]. VOJNOSANITETSKI PREGLED, 2017, 74 (08) : 763 - 766
  • [28] A Piezoresistive-Based Contact Lens for Non-invasive Intraocular Pressure Measurement
    Kumuda, S.
    Gandhi, Uma
    Mangalanathan, Umapathy
    Rajanna, K.
    Pandi, N. Veera
    [J]. SENSING AND IMAGING, 2024, 25 (01):
  • [29] Non-Contact Intraocular Pressure Measurement Method using Relation between Deformed Cornea and Reflected Pneumatic Pressure
    Kim, Hyung Jin
    Seo, Young Ho
    Kim, Byeong Hee
    [J]. INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2018, 19 (05) : 737 - 744
  • [30] Non-Contact Intraocular Pressure Measurement Method using Relation between Deformed Cornea and Reflected Pneumatic Pressure
    Hyung Jin Kim
    Young Ho Seo
    Byeong Hee Kim
    [J]. International Journal of Precision Engineering and Manufacturing, 2018, 19 : 737 - 744