Comparability analysis of the HNT-1P Huvitz non contact tonometer for the measurement of intraocular pressure

被引:0
|
作者
Montoya, Alicia Carceles [1 ]
Lucas, Carlos Enrique Monera [1 ,2 ]
Jaimes, Claudia Patricia Tarazona [1 ,2 ]
Valero, Daniel Romero [1 ,2 ]
Martinez, Alejandro Moya [2 ,3 ]
Toldos, Jose Juan Martinez [1 ]
机构
[1] Gen Univ Hosp Elche, Ophthalmol Dept, C/Camino Almazara,11, Alicante 03203, Spain
[2] Miguel Hernandez Univ Elche, Alicante, Spain
[3] Gen Univ Hosp Elche, Biostat, FISABIO, Alicante, Spain
关键词
Glaucoma; non-contact tonometer; intraocular pressure; goldmann applanation tonometer; DYNAMIC CONTOUR TONOMETRY; REBOUND TONOMETRY; APPLANATION;
D O I
10.1177/11206721241263624
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the diagnostic validity of the HNT-1P non-contact tonometer (Huvitz) as a tool for accurately measuring intraocular pressure (IOP) in patients with healthy eyes, compared to the Goldmann applanation tonometer (GAT), which is the Gold Standard method for measurement of IOP. Methods Observational, descriptive, transversal study using 148 eyes of 74 healthy patients without a diagnosis of glaucoma or other ophthalmological diseases. Three measurements of IOP were taken in each eye, using three tonometers: HNT-1P, ICR100, and GAT. The median IOP (quartiles) and mean IOP (SD) its statistical significance were calculated, and comparisons were made between the mean and median IOP values found in three groups: GAT-HNT, GAT-ICR, and HNT-ICR. The difference in mean and median IOP was analyzed in each of the three study groups, and its statistical significance and concordance correlation coefficient (CCC) were calculated. Results The median IOP with HNT-1P was statistically significantly lower than the median IOP with GAT, (1.1 mmHg, p < 0.001). The median IOP with HNT-1P was also lower than the median IOP with ICR100. As an additional result, the median IOP with GAT was lower than the median IOP with ICR. The CCC was moderate for HNT-ICR (0.72) and low for GAT-HNT and GAT-ICR (0.43 and 0.38, respectively). Conclusions HNT-1P (Huvitz) provides statistically significantly lower IOP values than those obtained with GAT. HNT-1P could be used for screening of ocular hypertension in postoperative patients. The IOP measurement obtained with HNT-1P should be confirmed with GAT. HNT-1P yields lower IOP values than those obtained with ICR.
引用
收藏
页码:544 / 550
页数:7
相关论文
共 32 条
  • [1] New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure
    Eah, Kyu Sang
    Shin, Joong Won
    Sung, Kyung Rim
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2020, 61 (05): : 524 - 531
  • [2] A Comparative Study Between the Goldmann Applanation Tonometer and the Non-Contact Air-Puff Tonometer (Huvitz HNT 7000) in Normal Eyes
    Stock, Ricardo Alexandre
    Stroher, Carine
    Sampaio, Rodrigo Rosa
    Mergener, Rafael Andre
    Bonamigo, Elcio Luiz
    CLINICAL OPHTHALMOLOGY, 2021, 15 : 445 - 451
  • [3] Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
    Porwal, Amit C.
    Shrishrimal, Meghna
    Punamia, Ravin P.
    Mathew, Bennet Chacko
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 1927 - 1931
  • [4] Intraocular Pressure Measurement With the Noncontact Tonometer Through Soft Contact Lenses
    Liu, Yi-Chun
    Huang, Jehn-Yu
    Wang, I-Jong
    Hu, Fung-Rong
    Hou, Yu-Chih
    JOURNAL OF GLAUCOMA, 2011, 20 (03) : 179 - 182
  • [5] Measurement of intraocular pressure using the NT-4000: A new non-contact tonometer equipped with pulse synchronous measurement function
    Yaoeda, K
    Shirakashi, M
    Fukushima, A
    Funaki, S
    Funaki, H
    Ofuchi, N
    Nakatsue, T
    Abe, H
    JOURNAL OF GLAUCOMA, 2005, 14 (03) : 201 - 205
  • [6] Agreement of intraocular pressure measurement with Corvis ST, non-contact tonometer, and Goldmann applanation tonometer in children with ocular hypertension and related factors
    Li, Hou-Gang
    Chen, Yan-Hui
    Lin, Fang
    Li, Si-Yu
    Liu, Qing-Hua
    Yin, Chun-Ge
    Chen, Xi-Yue
    Zhang, Xin-Jie
    Qu, Yue
    Hui, Yan-Nian
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2023, 16 (10) : 1601 - 1607
  • [7] Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study
    Nacaroglu, Senay Asik
    Un, Emine Seker
    Ersoz, Mehmet Giray
    Tasci, Yelda
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2015, 8 (03) : 540 - 543
  • [8] COMPARISON OF INTRAOCULAR PRESSURE MEASURED WITH GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR-PUFF TONOMETER
    Mahsud, Hidayatullah
    Saleem, Muhammad Wali
    Muhammad, Rafiq
    Saleem, Muhammad
    Ayaz, Mirqad
    GOMAL JOURNAL OF MEDICAL SCIENCES, 2016, 14 (02): : 63 - 66
  • [9] Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group
    Min Chen
    Lina Zhang
    Jia Xu
    Xinyi Chen
    Yuxiang Gu
    Yuping Ren
    Kaijun Wang
    BMC Ophthalmology, 19
  • [10] Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group
    Chen, Min
    Zhang, Lina
    Xu, Jia
    Chen, Xinyi
    Gu, Yuxiang
    Ren, Yuping
    Wang, Kaijun
    BMC OPHTHALMOLOGY, 2019, 19 (01)