Stem to prevent periprosthetic fracture after notching in total knee arthroplasty

被引:0
作者
Wan, Qian [1 ]
Zhang, Aobo [1 ]
Liu, Yang [1 ]
Chen, Hao [1 ]
Zhao, Xue [2 ]
Han, Qing [1 ]
Wang, Jincheng [1 ]
机构
[1] Second Hosp Jilin Univ, Dept Orthoped, Changchun 130000, Peoples R China
[2] First Hosp Jilin Univ, Dept Endocrinol & Metab, Changchun, Peoples R China
基金
中国国家自然科学基金;
关键词
aseptic loosening; finite element analysis; micromotion; notching; periprosthetic fracture; total knee arthroplasty; SUPRACONDYLAR FRACTURE; FEMUR; RISK; FAILURE; DESIGN; INTACT; LENGTH; HIP;
D O I
10.1002/cnm.3842
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Improper osteotomy during total knee arthroplasty (TKA) can lead to anterior femoral notching, which increases the risk of periprosthetic fractures due to stress concentration. One potential solution is the addition of an intramedullary stem to the femoral component. However, the optimal stem length remains unclear. In this study, we aimed to determine the optimal stem length using finite element models. Finite element models of femurs were developed with unstemmed prostheses and prostheses with stem lengths of 50, 75, and 100 mm. Under squat loading conditions, the von Mises stress at the notch and stress distribution on four transversal sections of the femur were analyzed. Additionally, micromotion of the prosthesis-bone interface was evaluated to assess initial stability. The unstemmed prosthesis exhibited a von Mises stress of 191.8 MPa at the notch, which decreased to 43.1, 8.8, and 23.5 MPa for stem lengths of 50, 75, and 100 mm, respectively. The stress reduction on four selected femoral transversal sections compared with the unstemmed prosthesis was 40.0%, 84.4%, and 67.1% for stem lengths of 50, 75, and 100 mm, respectively. Micromotion analysis showed a maximum of 118.8 mu m for the unstemmed prosthesis, which decreased significantly with the application of stems, particularly at the anterior flange. Intramedullary stems effectively reduced stress concentration at the femoral notch. The 50-mm stem length provided the optimal combination of reduced notch stress, minimized stress-shielding effect, and decreased micromotion at the anterior flange. This study used finite element models to investigate the effects of intramedullary stem on reducing risk of periprosthetic fractures after anterior femoral notching in total knee arthroplasty. Results showed that stemmed prostheses with a 50-mm stem length reduced stress concentration and micromotion at the anterior flange, making it the optimal choice.image
引用
收藏
页数:13
相关论文
共 41 条
[1]   FE analysis of the effects of simplifications in experimental testing on micromotions of uncemented femoral knee implants [J].
Berahmani, Sanaz ;
Janssen, Dennis ;
Wolfson, David ;
Malefijt, Maarten de Waal ;
Fitzpatrick, Clare K. ;
Rullkoetter, Paul J. ;
Verdonschot, Nico .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2016, 34 (05) :812-819
[2]   Experimental validation of intact and implanted distal femur finite element models [J].
Completo, A. ;
Fonseca, F. ;
Simoes, J. A. .
JOURNAL OF BIOMECHANICS, 2007, 40 (11) :2467-2476
[3]   Improved stability with intramedullary stem after anterior femoral notching in total knee arthroplasty [J].
Completo, A. ;
Fonseca, F. ;
Relvas, C. ;
Ramos, A. ;
Simoes, J. A. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2012, 20 (03) :487-494
[4]   Revision total knee arthroplasty: The influence of femoral stems in load sharing and stability [J].
Completo, A. ;
Simoes, J. A. ;
Fonseca, F. .
KNEE, 2009, 16 (04) :275-279
[5]   Optimum stem length for mitigation of periprosthetic fracture risk following primary total knee arthroplasty: a finite element study [J].
Conlisk, Noel ;
Howie, Colin R. ;
Pankaj, Pankaj .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2018, 26 (05) :1420-1428
[6]   Computational modelling of motion at the bone-implant interface after total knee arthroplasty: The role of implant design and surgical fit [J].
Conlisk, Noel ;
Howie, Colin R. ;
Pankaj, Pankaj .
KNEE, 2017, 24 (05) :994-1005
[7]   The Role of Complex Clinical Scenarios in the Failure of Modular Components Following Revision Total Knee Arthroplasty: A Finite Element Study [J].
Conlisk, Noel ;
Howie, Colin R. ;
Pankaj, Pankaj .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (08) :1134-1141
[8]  
CULP RW, 1987, CLIN ORTHOP RELAT R, P212
[9]   SKELETAL STRUCTURAL ADAPTATIONS TO MECHANICAL USAGE (SATMU) .1. REDEFINING WOLFF LAW - THE BONE MODELING PROBLEM [J].
FROST, HM .
ANATOMICAL RECORD, 1990, 226 (04) :403-413
[10]   Finite Element Model Updating Combined with Multi-Response Optimization for Hyper-Elastic Materials Characterization [J].
Iniguez-Macedo, Saul ;
Lostado-Lorza, Ruben ;
Escribano-Garcia, Ruben ;
Angeles Martinez-Calvo, Maria .
MATERIALS, 2019, 12 (07)