Evolution of dose distribution for a 125I eye plaque using MCNPX and Plaque Simulator software

被引:0
|
作者
Poorbaygi, Hosein [1 ,2 ]
Sardari, Elham [2 ]
Aghamiri, Seyed Mahmoud Reza [2 ]
Sheibani, Shahab [1 ]
Arjmand, Mojtaba [3 ]
Moradi, Somayeh [1 ]
Ghassemi, Fariba [3 ]
机构
[1] Radiat Applicat Res Sch Nucl Sci & Technol, Tehran, Iran
[2] Shahid Beheshti Univ, Dept Radiat Med, Tehran, Iran
[3] Farabi Eye Hosp, Ocular Oncol Serv, Tehran, Iran
来源
PRECISION RADIATION ONCOLOGY | 2024年 / 8卷 / 02期
关键词
I-125; seed; dosimetry calculation; Eye plaque; Monte Carlo N-Particle eXtended (MCNPX); treatment planning system; TREATMENT PLANNING SYSTEM; OCULAR MELANOMA; UVEAL MELANOMA; BRACHYTHERAPY; DOSIMETRY; PD-103; ENUCLEATION; AAPM;
D O I
10.1002/pro6.1231
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Eye plaque radiation therapy is the treatment of choice for small- and medium-sized choroidal melanomas. This study investigated the dose distribution around eye plaques containing I-125 seeds to treat ocular melanoma using Monte Carlo N-Particle eXtended (MCNPX) and Plaque Simulator (PS) software. Methods: Dosimetry evaluation and comparison of the resulting isodose curves for I-125 COMS plaques were performed using the MCNPX code. The isodose curves and dose distributions were calculated using PS treatment planning for a I-125 COMS plaque. Results: In the validation, the maximum relative difference between the results of this study and those reported in other literature was approximately 9%-10% for the COMS plaques. The dose distributions of MCNPX were lower than those of PS with a relative difference of approximately 27.7%-35.4%. The dose distribution may differ depending on the I-125 source spectra and seed design used in the two methods. In addition, the dose algorithm used in PS made a major contribution to the relative differences between the results. Conclusion: The PS did not provide accurate details of dose distribution near the surface of the plaque insert. The source parameters used in each program should be studied more carefully to determine the source of the differences in the estimated dose values.
引用
收藏
页码:70 / 76
页数:7
相关论文
共 50 条
  • [41] Clinical characteristics and postoperative complications as predictors of radiation toxicity after treatment with I125 Eye Plaque Brachytherapy for Uveal Melanomas
    Peters, Vanessa
    Gurayah, Aaron
    Jin, William
    Kwon, Deukwoo
    Zhao, Wei
    Patel, Nirav, V
    Markoe, Arnold
    Correa, Zelia
    Studenski, Matthew T.
    Harbour, J. William
    Samuels, Stuart E.
    BRACHYTHERAPY, 2022, 21 (06) : 896 - 903
  • [42] Incorporating patient-specific CT-based ophthalmic anatomy in modeling iodine-125 eye plaque brachytherapy dose distributions
    Tien, Christopher J.
    Astrahan, Melvin A.
    Kim, Jenna M.
    Materin, Miguel
    Chen, Zhe
    Nath, Ravinder
    Liu, Wu
    BRACHYTHERAPY, 2017, 16 (05) : 1057 - 1064
  • [43] Scleral necrosis in patients with posterior uveal melanomas evaluated by transcleral fine needle aspiration biopsy and treated by 125I plaque
    Correa, Zelia M.
    Huth, Bradley
    Augsburger, James J.
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2018, 81 (04) : 330 - 335
  • [44] Evaluation of material heterogeneity dosimetric effects using radiochromic film for COMS eye plaques loaded with 125I seeds (model I25.S16)
    Acar, Hilal
    Chiu-Tsao, Sou-Tung
    Ozbay, Ismail
    Kemikler, Gonul
    Tuncer, Samuray
    MEDICAL PHYSICS, 2013, 40 (01)
  • [45] Dosimetric comparison between realistic ocular model and other models for COMS plaque brachytherapy with 103Pd, 131Cs, and 125I radioisotopes
    Atiyeh Ebrahimi-Khankook
    Alireza Vejdani-Noghreiyan
    Radiation and Environmental Biophysics, 2018, 57 : 265 - 275
  • [46] Novel application of 3D printing in brachytherapy using MED610 3D printed insert for I-125 ROPES eye plaque
    L. Sim
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 863 - 870
  • [47] Dose optimization in 125I permanent prostate seed implants using the Monte Carlo method
    Reis, Juraci P.
    Menezes, Artur F.
    Souza, Edmilson M.
    Facure, Alessandro
    Medeiros, Jose A. C. C.
    Silva, Ademir X.
    COMPUTER PHYSICS COMMUNICATIONS, 2012, 183 (04) : 847 - 852
  • [48] Determination of dosimetric characteristics of a new design 125I brachytherapy source with an Ag + Al2O3 marker using the Monte Carlo code MCNPX
    Babaheidari A.E.
    Shamsaee M.
    Babaheidari, A.E., 1600, Pleiades Publishing (56): : 550 - 553
  • [49] Investigation of the 106Ru ophthalmic plaque dose distribution in a pseudotumor-contained eye-equivalent phantom by silica glass beads
    Hashemi, Samaneh
    Kahani, Mahdi
    Siavashpour, Zahra
    Karimi, Saeed
    Jaberi, Ramin
    RADIATION MEASUREMENTS, 2024, 175
  • [50] IN VIVO DOSIMETRY USING A LINEAR MOSFET-ARRAY DOSIMETER TO DETERMINE THE URETHRA DOSE IN 125I PERMANENT PROSTATE IMPLANTS
    Bloemen-van Gurp, Esther J.
    Murrer, Lars H. P.
    Haanstra, Bjork K. C.
    van Gils, Francis C. J. M.
    Dekker, Andre L. A. J.
    Munheer, Ben J.
    Lambin, Philippe
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (01): : 314 - 321