Reliability and accuracy of scoliotic parameters on using a wireless handheld 3D ultrasound for children with adolescent idiopathic scoliosis: a pilot study

被引:1
作者
Nguyen, Thi N. N. [1 ]
Le, Lawrence H. [1 ]
Emery, Derek J. [1 ]
Stampe, Kyle [1 ]
Southon, Sarah Hryniuk [2 ]
Lou, Edmond [1 ]
机构
[1] Univ Alberta, Edmonton, AB T6G 2R3, Canada
[2] Alberta Hlth Serv, Edmonton, AB T5G 0B7, Canada
关键词
Adolescent idiopathic scoliosis; 3D ultrasound; Portable; Wireless; Real-time reconstruction; Reliability; VERTEBRAL ROTATION; CORONAL CURVATURE; AID;
D O I
10.1007/s00586-024-08445-w
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
PurposeTo report the accuracy and reliability of Cobb angle (CA), axial vertebral rotation (AVR), kyphotic and lordotic angles (KA and LA) measurements on using a new 3D ultrasound (US) system.MethodsForty participants (34 F, 6 M, aged 14.0 +/- 2.3 years) were recruited. The first 20 participants were scanned by the validated US system and the new US system. The other 20 participants were scanned with the new US system only. Two raters (R1 and R2) performed the measurements: R1 has 10 years of experience in radiology but is new in ultrasound scoliosis, while R2 has 30 years of scoliosis experience. All US images were measured twice by R1, and once by R2. Forty posteroanterior and 30 lateral standing radiographs were obtained and measured once by R1. Statistical analysis consisted of mean absolute difference (MAD), intraclass correlation coefficient (ICC (2,1)), and Bland-Altman plots.ResultsR1 showed excellent intra-rater and inter-rater reliability for US measurements with ICCs(2,1) >= 0.91. The inter-method reliability was good between the two US systems for all parameters with ICCs(2,1) >= 0.85 and maximum MAD of 3.4 degrees. The new US showed good reliability and accuracy compared to radiographs for CA, AVR and KA with ICCs(2,1) >= 0.81 and maximum MAD of 5.8 degrees, but poor results for LA with ICCs(2,1) of 0.27-0.35 and MADs of 14.0 degrees-15.4 degrees.ConclusionThe new 3D US system showed good reliability and accuracy for CA, AVR and KA measurements, but a large measurement discrepancy on LA. A new measurement method for US LA may need to investigate.
引用
收藏
页码:3784 / 3791
页数:8
相关论文
共 25 条
  • [1] Chen HB, 2020, IEEE ENG MED BIO, P2105, DOI 10.1109/EMBC44109.2020.9176614
  • [2] Reliability of the axial vertebral rotation measurements of adolescent idiopathic scoliosis using the center of lamina method on ultrasound images: in vitro and in vivo study
    Chen, Wei
    Le, Lawrence H.
    Lou, Edmond H. M.
    [J]. EUROPEAN SPINE JOURNAL, 2016, 25 (10) : 3265 - 3273
  • [3] Reliability of assessing the coronal curvature of children with scoliosis by using ultrasound images
    Chen, Wei
    Lou, Edmond H. M.
    Zhang, Phoebe Q.
    Le, Lawrence H.
    Hill, Doug
    [J]. JOURNAL OF CHILDRENS ORTHOPAEDICS, 2013, 7 (06) : 521 - 529
  • [4] Cobb J, 1948, Instr Course Lect AAOS, V5, P261
  • [5] Horne JP, 2014, ADOLESCENT IDIOPATHI
  • [6] SOSORT 2012 consensus paper: reducing x-ray exposure in pediatric patients with scoliosis
    Knott, Patrick
    Pappo, Eden
    Cameron, Michelle
    deMauroy, J. C.
    Rivard, Charles
    Kotwicki, Tomasz
    Zaina, Fabio
    Wynne, James
    Stikeleather, Luke
    Bettany-Saltikov, Josette
    Grivas, Theodoros B.
    Durmala, Jacek
    Maruyama, Toru
    Negrini, Stefano
    O'Brien, Joseph P.
    Rigo, Manuel
    [J]. SCOLIOSIS AND SPINAL DISORDERS, 2014, 9
  • [7] A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research
    Koo, Terry K.
    Li, Mae Y.
    [J]. JOURNAL OF CHIROPRACTIC MEDICINE, 2016, 15 (02) : 155 - 163
  • [8] A NOVEL METHOD TO MEASURE THE SAGITTAL CURVATURE IN SPINAL DEFORMITIES: THE RELIABILITY AND FEASIBILITY OF 3-D ULTRASOUND IMAGING
    Lee, Timothy Tin-Yan
    Jiang, Wei Wei
    Cheng, Connie Lok Kan
    Lai, Kelly Ka-Lee
    To, Michael Kai Tsun
    Castelein, Rene M.
    Cheung, Jason Pui Yin
    Zheng, Yong-Ping
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2019, 45 (10) : 2725 - 2735
  • [9] COMPUTERIZED ULTRASONIC DIGITIZATION IN THE MEASUREMENT OF SPINAL CURVATURE
    LETTS, M
    QUANBURY, A
    GOUW, G
    KOLSUN, W
    LETTS, E
    [J]. SPINE, 1988, 13 (10) : 1106 - 1110
  • [10] Lou E, 2021, Stud Health Technol Inform, V280, P100, DOI 10.3233/SHTI210444