6-year follow-up on migration outcomes: a randomised clinical trial of cemented vitamin E-stabilised highly crosslinked versus standard polyethylene cup in total hip arthroplasty

被引:0
作者
Skoldenberg, Olof [1 ]
Mukka, Sebastian [2 ]
Magneli, Martin [1 ]
机构
[1] Karolinska Inst, Danderyd Hosp, Unit Orthopaed, Dept Clin Sci, Stockholm, Sweden
[2] Umea Univ, Dept Surg & Perioperat Sci Orthopaed, Umea, Sweden
关键词
Total hip replacement; Radiostereometric analysis; RSA; Randomized clinical trial; Vitamin E Highly Crosslinked Polyethylene; Highly Crosslinked Polyethylene; WEAR; RSA; RESISTANCE;
D O I
10.1177/11207000241267971
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: In a previous study we have shown that a cemented vitamin E-doped highly cross-linked polyethylene (VEPE) compared to a conventional polyethylene cup in total hip arthroplasty (THA) has a slightly higher proximal migration but significantly lower wear rates up to 2 years after surgery. In this follow-up study we investigated the same cohort at 6 years.Methods: This was a double-blinded, non-inferiority, randomised controlled trial on patients with osteoarthritis, with a mean age of 66 years. Patients were randomly assigned to receive either the conventional polyethylene cup or the VEPE cup in a 1:1 ratio. The primary endpoint was proximal implant migration of the cup measured with radiostereometric analysis (RSA). Secondary endpoints included wear rate of the cup and patient-reported outcome measurements (PROM).Results: At the 6-year follow-up, 25 patients (11 controls, 14 VEPE) were available for RSA measurements, and we found no statistically significant difference in proximal migration between the VEPE and control groups. The wear rate was significantly lower in the VEPE group compared to controls, 0.03 mm/year and 0.07 mm/year, respectively with a mean difference 0.04 mm, (95% CI, 0.02-0.06 mm). There were no cup revisions and no difference in PROM between the groups.Conclusions: Based on our 6-year results, the VEPE group exhibited no statistical or clinically relevant difference compared to the control group, and the wear rate was significantly lower in the VEPE group. The use of a cemented vitamin E-doped highly cross-linked cup is a good option in total hip arthroplasty.
引用
收藏
页码:695 / 700
页数:6
相关论文
共 19 条
  • [1] Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily: a randomized, controlled RSA trial with 5-year follow-up
    Bergvinsson, Halldor
    Zampelis, Vasilis
    Sundberg, Martin
    Tjornstrand, Jon
    Flivik, Gunnar
    [J]. ACTA ORTHOPAEDICA, 2022, 93 : 249 - 255
  • [2] Research electronic data capture (REDCap)-A metadata-driven methodology and workflow process for providing translational research informatics support
    Harris, Paul A.
    Taylor, Robert
    Thielke, Robert
    Payne, Jonathon
    Gonzalez, Nathaniel
    Conde, Jose G.
    [J]. JOURNAL OF BIOMEDICAL INFORMATICS, 2009, 42 (02) : 377 - 381
  • [3] DOES EARLY MICROMOTION OF FEMORAL STEM PROSTHESES MATTER - 4-7-YEAR STEREORADIOGRAPHIC FOLLOW-UP OF 84 CEMENTED PROSTHESES
    KARRHOLM, J
    BORSSEN, B
    LOWENHIELM, G
    SNORRASON, F
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06): : 912 - 917
  • [4] E-vitamin infused highly cross-linked polyethylene: RSA results from a randomised controlled trial using 32 mm and 36 mm ceramic heads
    Lindalen, Einar
    Nordsletten, Lars
    Hovik, Oystein
    Rohrl, Stephan M.
    [J]. HIP INTERNATIONAL, 2015, 25 (01) : 50 - 55
  • [5] The Stepwise Introduction of Innovation into Orthopedic Surgery The Next Level of Dilemmas
    Malchau, Henrik
    Bragdon, Charles R.
    Muratoglu, Orhun K.
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (06) : 825 - 831
  • [6] A vitamin E blended highly cross-linked polyethylene acetabular cup results in less wear: 6-year results of a randomized controlled trial in 199 patients
    Massier, Julie R. A.
    Van Erp, Joost H. J.
    Snijders, Thom E.
    De Gast, Arthur
    [J]. ACTA ORTHOPAEDICA, 2020, 91 (06) : 705 - 710
  • [7] Vitamin E diffused highly cross-linked polyethylene in total hip arthroplasty at five years
    Nebergall, A. K.
    Greene, M. E.
    Laursen, M. B.
    Nielsen, P. T.
    Malchau, H.
    Troelsen, A.
    [J]. BONE & JOINT JOURNAL, 2017, 99-B (05) : 577 - 584
  • [8] α-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear
    Oral, E
    Wannomae, KK
    Hawkins, N
    Harris, WH
    Muratoglu, OK
    [J]. BIOMATERIALS, 2004, 25 (24) : 5515 - 5522
  • [9] Wear resistance and mechanical properties of highly cross-linked, ultrahigh-molecular weight polyethylene doped with vitamin E
    Oral, Ebru
    Christensen, Steven D.
    Malhi, Arnaz S.
    Wannomae, Keith K.
    Muratoglu, Orhun K.
    [J]. JOURNAL OF ARTHROPLASTY, 2006, 21 (04) : 580 - 591
  • [10] Evaluation of a new cemented highly cross-linked all-polyethylene cup: a prospective and randomised study assessing wear and fixation characteristics using radiostereometric analysis
    Otten, Volker
    Wasterlund, Daniel
    Lindbjorn, John
    Mertens, Carl
    Mukka, Sebastian
    Crnalic, Sead
    Nilsson, Kjell G.
    [J]. HIP INTERNATIONAL, 2022, 32 (06) : 779 - 786