Reduced Intravenous Contrast Dose Portal Venous Phase Photon-Counting Computed Tomography Compared With Conventional Energy-Integrating Detector Portal Venous Phase Computed Tomography

被引:0
作者
Dane, Bari [1 ]
Mabud, Tarub [1 ]
Melamud, Kira [1 ]
Ginocchio, Luke [1 ]
Smereka, Paul [1 ]
Okyere, Mabel [1 ]
O'Donnell, Thomas [2 ]
Megibow, Alec [1 ]
机构
[1] NYU Langone Hlth, Dept Radiol, New York, NY 10016 USA
[2] Siemens Healthineers, Malvern, PA USA
关键词
photon-counting CT; intravenous contrast; oncology; portal venous phase imaging; conventional energy-integrating detector CT; CT; CONSPICUOUSNESS; AGREEMENT; NOISE; COST;
D O I
10.1097/RCT.0000000000001617
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to compare portal venous phase photon-counting CT (PCCT) using 20 cc less than weight-based contrast dosing with energy-integrating detector CT (EID-CT) using weight-based dosing by quantitative and qualitative analysis. Methods: Fifty adult patients who underwent a reduced intravenous contrast dose portal venous phase PCCT from May 1, 2023, to August 10, 2023, and a prior portal-venous EID-CT with weight-based contrast dosing were retrospectively identified. Hounsfield units (HU) and noise (SD of HU) were obtained from region-of-interest measurements on 70-keV PCCT and EID-CT in 4 hepatic segments, the main and right portal vein, and both paraspinal muscles. Signal-to-noise and contrast-to-noise ratios were computed. Three abdominal radiologists qualitatively assessed overall image quality, hepatic enhancement, and confidence for metastasis identification on 5-point Likert scales. Readers also recorded the presence/absence of hepatic metastases. Quantitative variables were compared with paired t tests, and multiple comparisons were accounted for with a Bonferroni-adjusted alpha level of .0016. Ordinal logistic regression was used to evaluate qualitative assessments. Interreader agreement for hepatic metastases was calculated using Fleiss' kappa. Results: Fifty patients (32 women; mean [SD] age, 64 [13] years) were included. There was no significant difference in hepatic HU, portal vein HU, noise, and signal-to-noise or contrast-to-noise ratio between reduced contrast dose portal venous phase PCCT versus EID-CT (all Ps > 0.0016). Image quality, degree of hepatic enhancement, and confidence for metastasis identification were not different for reduced dose PCCT 70-keV images and EID-CT (P = 0.06-0.69). kappa Value for metastasis identification was 0.86 (95% confidence interval, 0.70-1.00) with PCCT and 0.78 (95% confidence interval, 0.59-0.98) with EID-CT. Conclusion: Reduced intravenous contrast portal venous phase PCCT 70-keV images had similar attenuation and image quality as EID-CT with weight-based dosing. Metastases were identified with near-perfect agreement in reduced dose PCCT 70-keV images.
引用
收藏
页码:675 / 682
页数:8
相关论文
共 33 条
  • [21] A prospective study comparing the quality of coronary computed tomography angiography images from photon counting and energy integrating detector systems
    Skoog, Susann
    Sandborg, Michael
    Henriksson, Lilian
    Sandstedt, Marten
    Gustafsson, Hakan
    Persson, Anders
    ACTA RADIOLOGICA, 2023, 64 (11) : 2957 - 2966
  • [22] Diagnostic Accuracy of Contrast-Enhanced Thoracic Photon-Counting Computed Tomography for Opportunistic Locoregional Staging of Breast Cancer Compared With Digital Mammography: A Prospective Trial
    Neubauer, Jakob
    Wilpert, Caroline
    Gebler, Oliver
    Taran, Florin-Andrei
    Pichotka, Martin
    Stein, Thomas
    Molina-Fuentes, Moises Felipe
    Weiss, Jakob
    Juhasz-Boess, Ingolf
    Bamberg, Fabian
    Windfuhr-Blum, Marisa
    Neubauer, Claudia
    INVESTIGATIVE RADIOLOGY, 2024, 59 (07) : 489 - 494
  • [23] Complementary role of computed tomography texture analysis for differentiation of pancreatic ductal adenocarcinoma from pancreatic neuroendocrine tumors in the portal-venous enhancement phase
    Reinert, Christian Philipp
    Baumgartner, Karolin
    Hepp, Tobias
    Bitzer, Michael
    Horger, Marius
    ABDOMINAL RADIOLOGY, 2020, 45 (03) : 750 - 758
  • [24] Spectral Photon-Counting Computed Tomography (SPCCT): in-vivo single-acquisition multi-phase liver imaging with a dual contrast agent protocol
    Si-Mohamed, Salim
    Tatard-Leitman, Valerie
    Laugerette, Alexis
    Sigovan, Monica
    Pfeiffer, Daniela
    Rummeny, Ernst J.
    Coulon, Philippe
    Yagil, Yoad
    Douek, Philippe
    Boussel, Loic
    Noeel, Peter B.
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [25] Can dual-energy computed tomography improve visualization of hypoenhancing liver lesions in portal venous phase? Assessment of advanced image-based virtual monoenergetic images
    Caruso, Damiano
    De Cecco, Carlo N.
    Schoepf, U. Joseph
    Schaefer, Amanda R.
    Leland, Parker W.
    Johnson, Dustin
    Laghi, Andrea
    Hardie, Andrew D.
    CLINICAL IMAGING, 2017, 41 : 118 - 124
  • [26] Triple-energy virtual monochromatic imaging with a photon-counting detector for reducing metal artifacts in half-beam dental computed tomography
    Lee, M.
    Lee, H.
    Lee, S.
    Kim, W.
    Lim, Y.
    Choa, H.
    Cha, B.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (11):
  • [27] Contrast-enhanced computed tomography in the venous rather than the arterial phase is essential for the evaluation of the right phrenic nerve
    Tsuji, Toshihiko
    Aoyama, Daisetsu
    Ishida, Tomokazu
    Nomura, Ryohei
    Kakehashi, Shota
    Mukai, Moe
    Hasegawa, Kanae
    Uzui, Hiroyasu
    Tada, Hiroshi
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2023, 46 (12): : 1526 - 1535
  • [28] Photon-counting, energy-resolving and super-resolution phase contrast X-ray imaging using an integrating detector
    O'Connell, Dylan W.
    Morgan, Kaye S.
    Ruben, Gary
    Schaff, Florian
    Croton, Linda C. P.
    Buckley, Genevieve A.
    Paganin, David M.
    Uesugi, Kentaro
    Kitchen, Marcus J.
    OPTICS EXPRESS, 2020, 28 (05) : 7080 - 7094
  • [29] Dual-Energy Multidetector Computed Tomography With Iodine Quantification in the Evaluation of Portal Vein Thrombosis: Is It Possible to Discard the Unenhanced Phase?
    Ascenti, Giorgio
    Sofia, Carmelo
    Silipigni, Salvatore
    Vinci, Sergio
    Pergolizzi, Stefano
    Marin, Daniele
    Mileto, Achille
    Mazziotti, Silvio
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2015, 66 (04): : 348 - 355
  • [30] Coronary calcium scoring on virtual non-contrast and virtual non-iodine reconstructions compared to true non-contrast images using photon-counting computed tomography
    Sharma, Simran P.
    van der Bie, Judith
    van Straten, Marcel
    Hirsch, Alexander
    Bos, Daniel
    Dijkshoorn, Marcel L.
    Booij, Ronald
    Budde, Ricardo P. J.
    EUROPEAN RADIOLOGY, 2024, 34 (06) : 3699 - 3707