Evaluation of forces applied to tissues during robotic-assisted surgical tasks using a novel force feedback technology

被引:2
作者
Awad, Michael M. [1 ]
Raynor, Mathew C. [2 ]
Padmanabhan-Kabana, Mika [3 ]
Schumacher, Lana Y. [4 ]
Blatnik, Jeffrey A. [1 ]
机构
[1] Washington Univ, Sch Med, Dept Surg, St Louis, MO 63110 USA
[2] Univ N Carolina, Sch Med, Dept Urol, Chapel Hill, NC USA
[3] Intuit Surg, Sunnyvale, CA USA
[4] Tufts Med Sch, Div Thorac Surg, Boston, MA USA
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2024年 / 38卷 / 10期
关键词
Force feedback; Haptic technology; Robotic-assisted surgery; Inanimate model; Ex-vivo tissue; HAPTIC FEEDBACK; RECTAL-CANCER; SURGERY; MANIPULATION;
D O I
10.1007/s00464-024-11131-z
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The absence of force feedback (FFB) is considered a technical limitation in robotic-assisted surgery (RAS). This pre-clinical study aims to evaluate the forces applied to tissues using a novel integrated FFB technology, which allows surgeons to sense forces exerted at the instrument tips. Methods: Twenty-eight surgeons with varying experience levels employed FFB instruments to perform three robotic-assisted surgical tasks, including retraction, dissection, and suturing, on inanimate or ex-vivo models, while the instrument sensors recorded and conveyed the applied forces to the surgeon hand controllers of the robotic system. Generalized Estimating Equations (GEE) models were used to analyze the mean and maximal forces applied during each task with the FFB sensor at the "Off" setting compared to the "High" sensitivity setting for retraction and to the "Low", "Medium", and "High" sensitivity settings for dissection and suturing. Sub-analysis was also performed on surgeon experience levels. Results: The use of FFB at any of the sensitivity settings resulted in a significant reduction in both the mean and maximal forces exerted on tissue during all three robotic-assisted surgical tasks (p < 0.0001). The maximal force exerted, potentially associated with tissue damage, was decreased by 36%, 41%, and 55% with the use of FFB at the "High" sensitivity setting while performing retraction, dissection, and interrupted suturing tasks, respectively. Further, the use of FFB resulted in substantial reductions in force variance during the performance of all three types of tasks. In general, reductions in mean and maximal forces were observed among surgeons at all experience levels. The degree of force reduction depends on the sensitivity setting selected and the types of surgical tasks evaluated. Conclusions: Our findings demonstrate that the utilization of FFB technology integrated in the robotic surgical system significantly reduced the forces exerted on tissue during the performance of surgical tasks at all surgeon experience levels. The reduction in the force applied and a consistency of force application achieved with FFB use, could result in decreases in tissue trauma and blood loss, potentially leading to better clinical outcomes in patients undergoing RAS. Future studies will be important to determine the impact of FFB instruments in a live clinical environment.
引用
收藏
页码:6193 / 6202
页数:10
相关论文
共 33 条
  • [31] The first year is the hardest: a comparison of early versus late experience after the introduction of robotic hiatal hernia repair
    Washington, Kimberly
    Watkins, Jeffrey R.
    Jeyarajah, D. Rohan
    [J]. JOURNAL OF ROBOTIC SURGERY, 2020, 14 (01) : 205 - 210
  • [32] Effects of a force feedback function in a surgical robot on the suturing procedure
    Yamasaki, Yoshimi
    Tokunaga, Masanori
    Sakai, Yoshihiro
    Kayasuga, Hiroki
    Nishihara, Teruyuki
    Tadano, Kotaro
    Kawashima, Kenji
    Haruki, Shigeo
    Kinugasa, Yusuke
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2024, 38 (03): : 1222 - 1229
  • [33] Comparative clinical outcomes of robot-assisted liver resection versus laparoscopic liver resection: A meta-analysis
    Zhang, Lilong
    Yuan, Qihang
    Xu, Yao
    Wang, Weixing
    [J]. PLOS ONE, 2020, 15 (10):