Center of Pressure Measurement Accuracy via Insoles with a Reduced Pressure Sensor Number during Gaits

被引:2
作者
Fuchs, Philip X. [1 ,2 ]
Chen, Wei-Han [1 ,3 ]
Shiang, Tzyy-Yuang [1 ]
机构
[1] Natl Taiwan Normal Univ, Dept Athlet Performance, 88,Sect 4,Tingzhou Rd, Taipei City 116, Taiwan
[2] Natl Taiwan Normal Univ, Dept Phys Educ & Sport Sci, 162,Sect 1,Heping East Rd, Taipei City 106, Taiwan
[3] Natl Dong Hwa Univ, Dept Phys Educ & Kinesiol, 1,Sect 2,Da Hsueh Rd, Hualian City 974301, Taiwan
关键词
analysis; assessment; biomechanics; kinetics; technology; PLANTAR PRESSURE; BALANCE; FEET;
D O I
10.3390/s24154918
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The objective was to compare simplified pressure insoles integrating different sensor numbers and to identify a promising range of sensor numbers for accurate center of pressure (CoP) measurement. Twelve participants wore a 99-sensor Pedar-X insole (100 Hz) during walking, jogging, and running. Eight simplified layouts were simulated, integrating 3-17 sensors. Concordance correlation coefficients (CCC) and root mean square errors (RMSE) between the original and simplified layouts were calculated for time-series mediolateral (ML) and anteroposterior (AP) CoP. Differences between layouts and between gait types were assessed via ANOVA and Friedman test. Concordance between the original and simplified layouts varied across layouts and gaits (CCC: 0.43-0.98; chi(2)((7)) >= 34.94, p < 0.001). RMSEML and RMSEAP [mm], respectively, were smaller in jogging (5 +/- 2, 15 +/- 9) than in walking (8 +/- 2, 22 +/- 4) and running (7 +/- 4, 20 +/- 7) (eta(2)(p): 0.70-0.83, p < 0.05). Only layouts with 11+ sensors achieved CCC >= 0.80 in all tests across gaits. The 13-sensor layout achieved CCC >= 0.95 with 95% confidence, representing the most promising compromise between sensor number and CoP accuracy. Future research may refine sensor placement, suggesting the use of 11-13 sensors. For coaches, therapists, and applied sports scientists, caution is recommended when using insoles with nine or fewer sensors. Consulting task-specific validation results for the intended products is advisable.
引用
收藏
页数:12
相关论文
共 30 条
[1]  
Altman DG., 1990, PRACTICAL STAT MED R
[2]   Gait Segmentation Method Using a Plantar Pressure Measurement System with Custom-Made Capacitive Sensors [J].
Aqueveque, Pablo ;
Germany, Enrique ;
Osorio, Rodrigo ;
Pastene, Francisco .
SENSORS, 2020, 20 (03)
[3]  
Berchtold A, 2016, Methodological Innovations, V9, DOI [10.1177/2059799116672875, 10.1177/2059799116672875, DOI 10.1177/2059799116672875]
[4]   PRESSURE DISTRIBUTION UNDER SYMPTOM-FREE FEET DURING BAREFOOT STANDING [J].
CAVANAGH, PR ;
RODGERS, MM ;
IIBOSHI, A .
FOOT & ANKLE, 1987, 7 (05) :262-276
[5]   Measurement agreement between a newly developed sensing insole and traditional laboratory-based method for footstrike pattern detection in runners [J].
Cheung, Roy T. H. ;
An, Winko W. ;
Au, Ivan P. H. ;
Zhang, Janet H. ;
Chan, Zoe Y. S. ;
Man, Alfred ;
Lau, Fannie O. Y. ;
Lam, Melody K. Y. ;
Lau, K. K. ;
Leung, C. Y. ;
Tsang, N. W. ;
Sze, Louis K. Y. ;
Lam, Gilbert W. K. .
PLOS ONE, 2017, 12 (06)
[6]   The Design and Simulation of a 16-Sensors Plantar Pressure Insole Layout for Different Applications: From Sports to Clinics, a Pilot Study [J].
Ciniglio, Alfredo ;
Guiotto, Annamaria ;
Spolaor, Fabiola ;
Sawacha, Zimi .
SENSORS, 2021, 21 (04) :1-24
[7]   Discrete sensors distribution for accurate plantar pressure analyses [J].
Claverie, Laetitia ;
Ille, Anne ;
Moretto, Pierre .
MEDICAL ENGINEERING & PHYSICS, 2016, 38 (12) :1489-1494
[8]   A POWER PRIMER [J].
COHEN, J .
PSYCHOLOGICAL BULLETIN, 1992, 112 (01) :155-159
[9]   Which biomechanical models are currently used in standing posture analysis? [J].
Cretual, A. .
NEUROPHYSIOLOGIE CLINIQUE-CLINICAL NEUROPHYSIOLOGY, 2015, 45 (4-5) :285-295
[10]   Effects of an auditory biofeedback device on plantar pressure in patients with chronic ankle instability [J].
Donovan, Luke ;
Feger, Mark A. ;
Hart, Joseph M. ;
Saliba, Susan ;
Park, Joseph ;
Hertel, Jay .
GAIT & POSTURE, 2016, 44 :29-36