Assessment of the accuracy and reliability of clinical thermometers for body temperature measurements at the hospital: a laboratory study

被引:1
作者
Shinsuphan, Nikorn [1 ,3 ]
Phanthi, Aphinan [3 ]
Leeudomwong, Theera [2 ]
Sanponpute, Tassanai [1 ]
机构
[1] Natl Inst Metrol Thailand, Med Metrol Lab, Pathum Thani 12120, Thailand
[2] King Mongkuts Univ Technol North Bangkok, Dept Ind Phys & Med Instrumentat, Bangkok 10800, Thailand
[3] Chiang Mai Univ, Maharaj Nakorn Chiang Mai Hosp, Fac Med, Maintenance Sect,Med Instrument Subsect, Chiang Mai 50200, Thailand
关键词
contact and non-contact thermometers; accuracy; body temperature; uncertainty; calibration; TYMPANIC THERMOMETER; SKIN THERMOMETER; MERCURY; PERFORMANCE;
D O I
10.1088/1361-6501/ad64f7
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study aims to assess the accuracy and reliability of two different types of thermometers: contact thermometers (CT) and non-contact thermometers (NCIRT) at Maharaj Nakorn Chiang Mai Hospital. The comparison method was carried out in a hospital laboratory setting and followed a specific calibration procedure. The mean readings, errors, and measurement uncertainties were reported and categorized according to the manufacturer's labeling models. The maximum permissible errors were defined by international standards and clinical acceptances for contact and NCIRTs, respectively. The results showed that thermometers of temperature probes (PTs) provided the highest level of accuracy. When compared to a reference standard thermometer (STD), the mean errors were only 0.04 degrees C at the calibration point of 37.0 degrees C. For digital CTs, the mean error at the calibration points of 37.0 degrees C was 0.24 degrees C. Most CT models exhibited high accuracy and reliability. However, the forehead (IRFTs) and ear (IRETs) readings significantly differed from the STD, with over half of all units not meeting the ISO 80601-2-56/ASTM E1965-98 standard requirements. These results suggest that CTs and PTs are highly accurate and reliable devices for measuring body temperature. However, both IRFTs and IRETs could be useful in certain situations, such as screening many people in a short amount of time, but their accuracy should be interpreted with caution and confirmed using a more precise method if necessary.
引用
收藏
页数:15
相关论文
共 44 条
[41]   Validity, reliability and feasibility of a new observation rating tool and a post encounter rating tool for the assessment of clinical reasoning skills of medical students during their internal medicine clerkship: a pilot study [J].
Haring, Catharina M. ;
Klaarwater, Claudia C. R. ;
Bouwmans, Geert A. ;
Cools, Bernadette M. ;
van Gurp, Petra J. M. ;
van der Meer, Jos W. M. ;
Postma, Cornelis T. .
BMC MEDICAL EDUCATION, 2020, 20 (01)
[42]   The Reliability of Intraoperative Assessment on Predicting Tumor Size, Myometrial Invasion, and Cervical Involvement in Patients With a Preoperative Diagnosis of Complex Atypical Hyperplasia or (Clinical Stage I) Endometrial Cancer A Prospective Cohort Study [J].
Smith, Brentley Q. ;
Boone, Jonathan D. ;
Thomas, Eric D. ;
Turner, Taylor B. ;
McGwin, Gerald ;
Stisher, Amanda M. ;
Leath, Charles A. ;
Novak, Lea ;
Huh, Warner K. .
AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2020, 43 (02) :122-127
[43]   Radiographic and clinical assessment on the accuracy and complications of C1 anterior lateral mass and C2 anterior pedicle screw placement in the TARP-III procedure: a study of 106 patients [J].
Xueshi Li ;
Fuzhi Ai ;
Hong Xia ;
Zenghui Wu ;
Xiangyang Ma ;
Qingshui Yin .
European Spine Journal, 2014, 23 :1712-1719
[44]   Radiographic and clinical assessment on the accuracy and complications of C1 anterior lateral mass and C2 anterior pedicle screw placement in the TARP-III procedure: a study of 106 patients [J].
Li, Xueshi ;
Ai, Fuzhi ;
Xia, Hong ;
Wu, Zenghui ;
Ma, Xiangyang ;
Yin, Qingshui .
EUROPEAN SPINE JOURNAL, 2014, 23 (08) :1712-1719