Assessment of the accuracy and reliability of clinical thermometers for body temperature measurements at the hospital: a laboratory study

被引:1
作者
Shinsuphan, Nikorn [1 ,3 ]
Phanthi, Aphinan [3 ]
Leeudomwong, Theera [2 ]
Sanponpute, Tassanai [1 ]
机构
[1] Natl Inst Metrol Thailand, Med Metrol Lab, Pathum Thani 12120, Thailand
[2] King Mongkuts Univ Technol North Bangkok, Dept Ind Phys & Med Instrumentat, Bangkok 10800, Thailand
[3] Chiang Mai Univ, Maharaj Nakorn Chiang Mai Hosp, Fac Med, Maintenance Sect,Med Instrument Subsect, Chiang Mai 50200, Thailand
关键词
contact and non-contact thermometers; accuracy; body temperature; uncertainty; calibration; TYMPANIC THERMOMETER; SKIN THERMOMETER; MERCURY; PERFORMANCE;
D O I
10.1088/1361-6501/ad64f7
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study aims to assess the accuracy and reliability of two different types of thermometers: contact thermometers (CT) and non-contact thermometers (NCIRT) at Maharaj Nakorn Chiang Mai Hospital. The comparison method was carried out in a hospital laboratory setting and followed a specific calibration procedure. The mean readings, errors, and measurement uncertainties were reported and categorized according to the manufacturer's labeling models. The maximum permissible errors were defined by international standards and clinical acceptances for contact and NCIRTs, respectively. The results showed that thermometers of temperature probes (PTs) provided the highest level of accuracy. When compared to a reference standard thermometer (STD), the mean errors were only 0.04 degrees C at the calibration point of 37.0 degrees C. For digital CTs, the mean error at the calibration points of 37.0 degrees C was 0.24 degrees C. Most CT models exhibited high accuracy and reliability. However, the forehead (IRFTs) and ear (IRETs) readings significantly differed from the STD, with over half of all units not meeting the ISO 80601-2-56/ASTM E1965-98 standard requirements. These results suggest that CTs and PTs are highly accurate and reliable devices for measuring body temperature. However, both IRFTs and IRETs could be useful in certain situations, such as screening many people in a short amount of time, but their accuracy should be interpreted with caution and confirmed using a more precise method if necessary.
引用
收藏
页数:15
相关论文
共 44 条
  • [21] Accuracy and reliability of measurements obtained with a noncontact tono-pachymeter for clinical use in mass screening
    Lee, Jinho
    Choi, Hyuk Jin
    SCIENTIFIC REPORTS, 2021, 11 (01) : 8900
  • [22] Contact-Less Body Temperature Monitoring by Infrared Camera: Accuracy Preliminary Assessment
    Leogrande, Elisabetta
    Botrugno, Chiara
    Natale, Teresa
    Dell'Olio, Francesco
    2023 IEEE SENSORS, 2023,
  • [23] Accuracy, agreement, and reliability of DECT-derived vBMD measurements: an initial ex vivo study
    Qin, Le
    Huang, Jingjing
    Yu, Pei
    Yan, Jing
    Ge, Yingqian
    Lu, Yong
    Yan, Fuhua
    Wang, Lei
    Du, Lianjun
    EUROPEAN RADIOLOGY, 2021, 31 (01) : 191 - 199
  • [24] Accuracy, agreement, and reliability of DECT-derived vBMD measurements: an initial ex vivo study
    Le Qin
    Jingjing Huang
    Pei Yu
    Jing Yan
    Yingqian Ge
    Yong Lu
    Fuhua Yan
    Lei Wang
    Lianjun Du
    European Radiology, 2021, 31 : 191 - 199
  • [25] The reliability of point-of-care prothrombin time testing. A comparison of CoaguChek S® and XS® INR measurements with hospital laboratory monitoring
    Ryan, F.
    O'Shea, S.
    Byrne, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2010, 32 (01) : E26 - E33
  • [26] Accuracy and usefulness in assessing proficiency of the observational clinical human reliability assessment checklist of the open inguinal hernia repair procedure: A cross-sectional study
    Nazari, T.
    Simons, M. P.
    van Merrienboer, J. J. G.
    Wiggers, T.
    INTERNATIONAL JOURNAL OF SURGERY, 2020, 82 : 156 - 161
  • [27] Long-Term Calibration and Accuracy Assessment of the FengYun-3 Microwave Temperature Sounder Radiance Measurements
    You, Ran
    Gu, Songyan
    Guo, Yang
    Wu, Xuebao
    Yang, Hu
    Chen, Wenxin
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2012, 50 (12): : 4854 - 4859
  • [28] Accuracy of recorded body temperature of critically ill patients related to measurement site: a prospective observational study
    Nonose, Y.
    Sato, Y.
    Kabayama, H.
    Arisawa, A.
    Onodera, M.
    Imanaka, H.
    Nishimura, M.
    ANAESTHESIA AND INTENSIVE CARE, 2012, 40 (05) : 820 - 824
  • [29] A Comparative Study of Non-contact Infrared and Digital Rectal Thermometer Measurements of Body Temperature in the Horse
    Ramey, David
    Bachmann, Kristin
    Lee, Martin L.
    JOURNAL OF EQUINE VETERINARY SCIENCE, 2011, 31 (04) : 191 - 193
  • [30] First and second morning spot urine protein measurements for the assessment of proteinuria: a diagnostic accuracy study in kidney transplant recipients
    Maja Mrevlje
    Manca Oblak
    Gregor Mlinšek
    Jelka Lindič
    Miha Jadranka-Buturović-Ponikvar
    BMC Nephrology, 22