Assessment of the accuracy and reliability of clinical thermometers for body temperature measurements at the hospital: a laboratory study

被引:1
|
作者
Shinsuphan, Nikorn [1 ,3 ]
Phanthi, Aphinan [3 ]
Leeudomwong, Theera [2 ]
Sanponpute, Tassanai [1 ]
机构
[1] Natl Inst Metrol Thailand, Med Metrol Lab, Pathum Thani 12120, Thailand
[2] King Mongkuts Univ Technol North Bangkok, Dept Ind Phys & Med Instrumentat, Bangkok 10800, Thailand
[3] Chiang Mai Univ, Maharaj Nakorn Chiang Mai Hosp, Fac Med, Maintenance Sect,Med Instrument Subsect, Chiang Mai 50200, Thailand
关键词
contact and non-contact thermometers; accuracy; body temperature; uncertainty; calibration; TYMPANIC THERMOMETER; SKIN THERMOMETER; MERCURY; PERFORMANCE;
D O I
10.1088/1361-6501/ad64f7
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study aims to assess the accuracy and reliability of two different types of thermometers: contact thermometers (CT) and non-contact thermometers (NCIRT) at Maharaj Nakorn Chiang Mai Hospital. The comparison method was carried out in a hospital laboratory setting and followed a specific calibration procedure. The mean readings, errors, and measurement uncertainties were reported and categorized according to the manufacturer's labeling models. The maximum permissible errors were defined by international standards and clinical acceptances for contact and NCIRTs, respectively. The results showed that thermometers of temperature probes (PTs) provided the highest level of accuracy. When compared to a reference standard thermometer (STD), the mean errors were only 0.04 degrees C at the calibration point of 37.0 degrees C. For digital CTs, the mean error at the calibration points of 37.0 degrees C was 0.24 degrees C. Most CT models exhibited high accuracy and reliability. However, the forehead (IRFTs) and ear (IRETs) readings significantly differed from the STD, with over half of all units not meeting the ISO 80601-2-56/ASTM E1965-98 standard requirements. These results suggest that CTs and PTs are highly accurate and reliable devices for measuring body temperature. However, both IRFTs and IRETs could be useful in certain situations, such as screening many people in a short amount of time, but their accuracy should be interpreted with caution and confirmed using a more precise method if necessary.
引用
收藏
页数:15
相关论文
共 44 条
  • [1] Evaluation of clinical thermometers for accuracy and reliability
    Latman, N.S.
    Hans, P.
    Nicholson, L.
    Delee Zint, S.
    Lewis, K.
    Shirey, A.
    Biomedical Instrumentation and Technology, 2001, 35 (04) : 259 - 265
  • [2] AN OBSERVATIONAL STUDY OF THE RELIABILITY OF BODY TEMPERATURE MEASUREMENTS USING DIFFERENT THERMOMETERS BY NPC TEST METHOD
    Giancristofaro, Rosa Arboretti
    Corain, Livio
    Di Teodoro, Linda
    Franceschini, Samuela
    Spinosi, Pasqualino
    JP JOURNAL OF BIOSTATISTICS, 2009, 3 (01) : 41 - 57
  • [3] An assessment of infrared tympanic thermometers for body temperature measurement
    Betta, V
    Cascetta, F
    Sepe, D
    PHYSIOLOGICAL MEASUREMENT, 1997, 18 (03) : 215 - 225
  • [4] Reliability of Body Temperature Measurements Obtained with Contactless Infrared Point Thermometers Commonly Used during the COVID-19 Pandemic
    Piccinini, Filippo
    Martinelli, Giovanni
    Carbonaro, Antonella
    SENSORS, 2021, 21 (11)
  • [5] Identifying Critical Design Parameters for Improved Body Temperature Measurements: A Clinical Study Comparing Transient and Predicted Temperature Measurements
    Vesnovsky, Oleg
    Zhu, Liang
    Grossman, Laurence W.
    Casamento, Jon P.
    Chamani, Alireza
    Wijekoon, Nadeesri
    Topoleski, L. D. Timmie
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2019, 13 (01):
  • [6] LABORATORY ASSESSMENT OF IMPRESSION ACCURACY BY CLINICAL SIMULATION
    WASSELL, RW
    ABUASI, HA
    JOURNAL OF DENTISTRY, 1992, 20 (02) : 108 - 114
  • [7] The assessment of electronic balances for accuracy of mass measurements in the analytical laboratory
    Gustavo Gonzalez, A.
    Angeles Herrador, M.
    ACCREDITATION AND QUALITY ASSURANCE, 2007, 12 (01) : 21 - 29
  • [8] Accuracy and Reliability of Peer Assessment of Athletic Training Psychomotor Laboratory Skills
    Marty, Melissa C.
    Henning, Jolene M.
    Willse, John T.
    JOURNAL OF ATHLETIC TRAINING, 2010, 45 (06) : 609 - 614
  • [9] Measurement of body temperature in adult patients: Comparative study of accuracy, reliability and validity of different devices
    Rubia-Rubia, J.
    Arias, A.
    Sierra, A.
    Aguirre-Jaime, A.
    INTERNATIONAL JOURNAL OF NURSING STUDIES, 2011, 48 (07) : 872 - 880
  • [10] The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis
    Pecoraro, Valentina
    Petri, Davide
    Costantino, Giorgio
    Squizzato, Alessandro
    Moja, Lorenzo
    Virgili, Gianni
    Lucenteforte, Ersilia
    INTERNAL AND EMERGENCY MEDICINE, 2021, 16 (04) : 1071 - 1083