The Effect of Increasing Thread Depth on the Initial Stability of Dental Implants: An In Vitro Study

被引:1
|
作者
Cucinelli, Chiara [1 ]
Pereira, Miguel Silva [1 ,2 ]
Borges, Tiago [1 ,2 ]
Figueiredo, Rui [3 ,4 ]
Leitao-Almeida, Bruno [1 ,2 ]
机构
[1] Univ Catolica Portuguesa, Fac Dent, P-3510 Viseu, Portugal
[2] Univ Catolica Portuguesa, Ctr Interdisciplinary Res Hlth CIIS, P-3510 Viseu, Portugal
[3] Univ Barcelona, Fac Med & Hlth Sci, Barcelona 08907, Spain
[4] Inst Invest Biomed Bellvitge IDIBELL, Barcelona 08908, Spain
来源
SURGERIES | 2024年 / 5卷 / 03期
关键词
dental implants; osseointegration; torque; resonance frequency analysis; RESONANCE FREQUENCY-ANALYSIS; INSERTION TORQUE; BONE-DENSITY; GEOMETRY; QUALITY;
D O I
10.3390/surgeries5030065
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: The long-term success of dental implants largely depends on achieving primary stability, previously described as crucial to obtaining osseointegration and immediate loading protocol requirements. Implant thread depths seem to be one of the key factors influencing primary stability, particularly in low-density bone. Insertion torque (IT) and resonance frequency analysis (RFA) are considered the most reliable tests to assess primary stability. The aim of this work was to evaluate how different thread depths of commercially available dental implants affect primary stability in low-density D3 bone. Materials and Methods: An in vitro study was carried out between February 2024 and March 2024. Twenty-four dental implants were divided into four groups (six implants each) according to their thread depths (Group A: 4 mm, Group B: 4.5 mm, Group C: 5 mm, Group D: 5.5 mm) and were inserted in D3-type artificial bone blocks. The main outcome variables were the IT and the Implant Stability Quotient (ISQ) measured in four different areas of the implant (buccal, lingual, mesial, and distal) with an Osstel (R) ISQ reader. Descriptive and inferential analyses of the data were performed, and the significance value was set at 5%. Results: A total of 24 implants were analyzed. The highest IT values were obtained in Group D, with a mean of 54.03 Ncm (standard deviation (SD) = 8.99), while the lowest measurements were observed in Group A (mean = 25.12; SD: 2.96 N.cm). The mean ISQ values were consistently higher in Group D for each analyzed area, with a mean of 70.13 N.cm (SD = 1.12). Conclusions: Taking into consideration the limitations of this in vitro study, greater thread depths seem to increase the primary stability of dental implants placed in soft bone. Furthermore, a positive correlation was observed between all IT and ISQ values, regardless of the thread depth.
引用
收藏
页码:817 / 825
页数:9
相关论文
共 50 条
  • [1] Primary Stability Optimization by Using Fixtures with Different Thread Depth According To Bone Density: A Clinical Prospective Study on Early Loaded Implants
    Makary, Christian
    Menhall, Abdallah
    Zammarie, Carole
    Lombardi, Teresa
    Lee, Seung Yeup
    Stacchi, Claudio
    Park, Kwang Bum
    MATERIALS, 2019, 12 (15)
  • [2] In vitro analysis of the influence of surface treatment of dental implants on primary stability
    Mazzo, Carla Rodrigues
    dos Reis, Andrea Candido
    Shimano, Antonio Carlos
    da Costa Valente, Mariana Lima
    BRAZILIAN ORAL RESEARCH, 2012, 26 (04) : 313 - 317
  • [3] Primary stability of narrow-diameter dental implants with a multiple condensing thread design placed in bone with and without osteotomes: An in vitro study
    Romanos, Georgios E.
    Lau, Jessica
    Delgado-Ruiz, Rafael
    Javed, Fawad
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2020, 22 (03) : 409 - 414
  • [4] Effect of Location on Primary Stability and Healing of Dental Implants
    Monje, Alberto
    Suarez, Fernando
    Garaicoa, Carlos Andres
    Monje, Florencio
    Galindo-Moreno, Pablo
    Garcia-Nogales, Agustin
    Wang, Hom-Lay
    IMPLANT DENTISTRY, 2014, 23 (01) : 69 - 73
  • [5] The Effect of V-Thread and Square Thread Dental Implants on Bone
    AL-Sanea, Abdulrahman
    Mutlu, Ibrahim
    Kisioglu, Yasin
    Mohamed, Eltahir
    JOURNAL OF BIOMIMETICS BIOMATERIALS AND BIOMEDICAL ENGINEERING, 2023, 60 : 83 - 96
  • [6] Initial stability and bone strain evaluation of the immediately loaded dental implant: an in vitro model study
    Huang, Heng-Li
    Chang, Yin-Yu
    Lin, Dan-Jae
    Li, Yu-Fen
    Chen, Kuan-Ting
    Hsu, Jui-Ting
    CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (07) : 691 - 698
  • [7] Optimizing the primary stability of dental implants in type IV bone: in-vitro comparison of machine-driven and ratcheting insertion protocols
    Taysi, Nuri Mert
    Taysi, Aysegul Erten
    Ercal, Pinar
    Sismanoglu, Soner
    EUROPEAN ORAL RESEARCH, 2025, 59 (01) : 46 - 52
  • [8] The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants
    Wu, Shu-Wei
    Lee, Chia-Ching
    Fu, Ping-Yuen
    Lin, Shang-Chih
    MEDICAL ENGINEERING & PHYSICS, 2012, 34 (07) : 797 - 805
  • [9] The effect of the thread depth on the mechanical properties of the dental implant
    Lee, Sun-Young
    Kim, Sung-Jun
    An, Hyun-Wook
    Kim, Hyun-Seung
    Ha, Dong-Guk
    Ryo, Kyung-Ho
    Park, Kwang-Bum
    JOURNAL OF ADVANCED PROSTHODONTICS, 2015, 7 (02) : 115 - 121
  • [10] Primary Stability/Mobility of 1-Stage and 2-Stage Dental Implants: A Comparative In Vitro Study
    Gulay, Gulsah
    Asar, Neset Volkan
    Tulunoglu, Ibrahim
    Turkyilmaz, Ilser
    Wang, Hom-Lay
    Tozum, Tolga F.
    IMPLANT DENTISTRY, 2012, 21 (06) : 461 - 466