Operating Room Air May Harbor Pathogens: The Role of an Ultraviolet Air Filtration Unit

被引:0
作者
Fernandez-Rodriguez, Diana [1 ,2 ]
Tarabichi, Saad [3 ]
Golankiewicz, Krystal [1 ]
Zappley, Nicolina [1 ]
Parvizi, Javad [4 ]
机构
[1] Thomas Jefferson Univ, Rothman Inst, Dept Orthoped Surg, Philadelphia, PA USA
[2] Univ Nacl Autonoma Mexico, Fac Med, Mexico City, Mexico
[3] Mayo Clin Arizona, Dept Orthopaed Surg, Phoenix, AZ USA
[4] Acibadem Univ Hosp, Int Joint Ctr, Istanbul, Turkiye
关键词
surgical site infection; arthroplasty; operating room; sterilization; contamination; SURGICAL SITE INFECTIONS; CONTAMINATION; JOINT; SURGERY; HIP; VENTILATION; REPLACEMENT; RADIATION; QUALITY; CARE;
D O I
10.1016/j.arth.2024.05.072
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: One important factor for the prevention of surgical site infections is ultraclean air in the operating room (OR). Still, the direct sterilization potential of most technologies, especially in a dynamic clinical setting, is not well understood. We aimed to determine and compare the microbial presence from the inlet and outlet flow of a filtration unit with crystalline ultraviolet-C (C-UVC) light. Methods: A prospective study was conducted at a single institution, where primary total joint arthroplasty and spine surgeries were performed. The OR was fitted with a positive ventilation system. In addition, a filtration unit with a C-UVC sterilizing light was placed in the OR. The inlet and outlet air flows were swabbed simultaneously and compared. Swabs were processed for culture and next-generation sequencing. Results: The mean length of the surgical procedures sampled was 68 +/- 13 minutes. Overall, 19 out of 200 (9.5%) swabs isolated microorganisms. Inlet air swabs were positive at a higher rate (16 versus 3%; P < .01) compared to the outlet air swabs. A wide variety of Gram-positive, Gram-negative, and anaerobic bacteria were isolated, but fungi were only recovered from inlet air swabs. The detection of microorganisms was also higher when more door openings were performed (32.5 +/- 7.1 versus 27.9 +/- 5.6; P < .01). Conclusions: Air swabs mainly isolated microorganisms from the inlet flow to the filtration unit with a CUVC light. The sterilizing unit counteracted factors affecting the air quality in the OR, namely door openings, surgical personnel, and tissue combustion. (c) 2024 Elsevier Inc. All rights are reserved, including those for text and data mining, AI training, and similar technologies.
引用
收藏
页码:2857 / 2862
页数:6
相关论文
共 50 条
  • [41] Time-dependent bacterial air contamination of sterile fields in a controlled operating room environment: an experimental intervention study
    Wistrand, C.
    Soderquist, B.
    Sundqvist, A-S
    [J]. JOURNAL OF HOSPITAL INFECTION, 2021, 110 : 97 - 102
  • [42] A stochastic model for infective events in operating room caused by air contamination
    Abundo, Paolo
    Rosato, Nicola
    Abundo, Mario
    [J]. COLLECTIVE DYNAMICS: TOPICS ON COMPETITION AND COOPERATION IN THE BIOSCIENCES, 2008, 1028 : 198 - +
  • [43] Effect of Urban Heat Islands on the Air-Conditioning of an Operating Room
    Ascionel, Fabrizio
    De Masi, Rosa Francesca
    Tariello, Francesco
    Vanoli, Giuseppe Peter
    [J]. 6TH INTERNATIONAL CONFERENCE ON COUNTERMEASURES TO URBAN HEAT ISLANDS, UHI 2023, 2023, : 1067 - 1076
  • [44] Numerical simulation of the dual-loop different velocity air curtain system in the operating room
    Cao, Weixue
    Zhang, Yixuan
    Zhang, Xudong
    Shao, Tianqi
    Fan, Man
    Shi, Feng
    Yang, Bin
    [J]. JOURNAL OF BUILDING ENGINEERING, 2024, 98
  • [45] Improving infectious bioaerosols removal in the ophthalmic negative-pressure operating room: The important role of air cleaners
    Bao, Haifeng
    Lin, Hai
    Huang, Junzhang
    Li, Zongkun
    Li, Yi
    [J]. INDOOR AND BUILT ENVIRONMENT, 2025,
  • [46] Three-dimensional air distribution analysis of different outflow typed operating rooms at different inlet velocities and room temperatures
    Ufat, Hande
    Kaynakli, Omer
    Yamankaradeniz, Nurettin
    Yamankaradeniz, Recep
    [J]. ADVANCES IN MECHANICAL ENGINEERING, 2017, 9 (07): : 1 - 12
  • [47] Catalytic Surgical Smoke Filtration Unit Reduces Formaldehyde Levels in a Simulated Operating Room Environment
    Carroll, Gregory T.
    Kirschman, David L.
    [J]. ACS CHEMICAL HEALTH & SAFETY, 2023, 30 (01) : 21 - 28
  • [48] Effect of air exhaust location on surgical site particle distribution in an operating room
    Agirman, Aleyna
    Cetin, Yunus Emre
    Avci, Mete
    Aydin, Orhan
    [J]. BUILDING SIMULATION, 2020, 13 (05) : 979 - 988
  • [49] Comparison of three distinct clean air suits to decrease the bacterial load in the operating room: An observational study
    Kasina P.
    Tammelin A.
    Blomfeldt A.-M.
    Ljungqvist B.
    Reinmüller B.
    Ottosson C.
    [J]. Patient Safety in Surgery, 10 (1)
  • [50] A Fuzzy Inference System for the Assessment of Indoor Air Quality in an Operating Room to Prevent Surgical Site Infection
    Colella, Ylenia
    Valente, Antonio Saverio
    Rossano, Lucia
    Trunfio, Teresa Angela
    Fiorillo, Antonella
    Improta, Giovanni
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2022, 19 (06)