Perovskite computed tomography imager and three-dimensional reconstruction

被引:49
作者
He, Yuhong [1 ]
Song, Jinmei [1 ]
Li, Mingbian [1 ]
Sakhatskyi, Kostiantyn [2 ,3 ]
Li, Weijun [1 ]
Feng, Xiaopeng [1 ]
Yang, Bai [1 ,4 ]
Kovalenko, Maksym [2 ,3 ]
Wei, Haotong [1 ,4 ]
机构
[1] Jilin Univ, Coll Chem, State Key Lab Supramol Struct & Mat, Changchun, Peoples R China
[2] Swiss Fed Inst Technol, Dept Chem & Appl Biosci, Lab Inorgan Chem, Zurich, Switzerland
[3] Empa Swiss Fed Labs Mat Sci & Technol, Dubendorf, Switzerland
[4] First Hosp Jilin Univ, Opt Funct Theragnost Joint Lab Med & Chem, Changchun, Peoples R China
基金
中国国家自然科学基金;
关键词
RADIATION; CT; DEFECTS; NOISE;
D O I
10.1038/s41566-024-01506-y
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Indirect scintillator computed tomography (CT) imaging suffers from reduced image contrast and high-dose X-ray exposure due to inevitable light losses and multiple energy conversion steps. Here we report a direct lead-halide perovskite CT imager through low-cost spray-coating processes. Detector arrays with 980 mu m absorber thickness and <10 nm surface roughness yield uniform X-ray response with detection quantum efficiency of 80% and noise-equivalent dose of 153 pGy(air). The perovskite CT imager affords the three-dimensional reconstruction of a tooth under a low effective dose of 5.5 mu Sv, about two orders of magnitude smaller than dental cone-beam CT, and low-contrast detectability by resolving a 5 Hounsfield unit difference within a 5 mm region of interest.
引用
收藏
页码:1052 / 1058
页数:9
相关论文
共 41 条
[1]   Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition [J].
Achenbach, Stephan ;
Marwan, Mohamed ;
Ropers, Dieter ;
Schepis, Tiziano ;
Pflederer, Tobias ;
Anders, Katharina ;
Kuettner, Axel ;
Daniel, Werner G. ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN HEART JOURNAL, 2010, 31 (03) :340-346
[2]   Computed tomography imaging with the Adaptive Statistical Iterative Reconstruction (ASIR) algorithm: dependence of image quality on the blending level of reconstruction [J].
Barca, Patrizio ;
Giannelli, Marco ;
Fantacci, Maria Evelina ;
Caramella, Davide .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2018, 41 (02) :463-473
[3]   CHARACTERISTICS AND CONTROL OF CONTRAST IN CT [J].
BARNES, JE .
RADIOGRAPHICS, 1992, 12 (04) :825-837
[4]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[5]   Integrated electroanatomic mapping with three-dimensional computed tomographic images for real-time guided ablations [J].
Dong, J ;
Calkins, H ;
Solomon, SB ;
Lai, SH ;
Dalal, D ;
Lardo, A ;
Brem, E ;
Preiss, A ;
Berger, RD ;
Halperin, H ;
Dickfeld, T .
CIRCULATION, 2006, 113 (02) :186-194
[6]   Effect of scattered radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N ;
Yoshida, K .
MEDICAL PHYSICS, 2001, 28 (04) :469-474
[7]   Exposure to Low-Dose Ionizing Radiation from Medical Imaging Procedures. [J].
Fazel, Reza ;
Krumholz, Harlan M. ;
Wang, Yongfei ;
Ross, Joseph S. ;
Chen, Jersey ;
Ting, Henry H. ;
Shah, Nilay D. ;
Nasir, Khurram ;
Einstein, Andrew J. ;
Nallamothu, Brahmajee K. .
NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (09) :849-857
[8]   Spray-coated perovskite hemispherical photodetector featuring narrow-band and wide-angle imaging [J].
Feng, Xiaopeng ;
He, Yuhong ;
Qu, Wei ;
Song, Jinmei ;
Pan, Wanting ;
Tan, Mingrui ;
Yang, Bai ;
Wei, Haotong .
NATURE COMMUNICATIONS, 2022, 13 (01)
[9]   Dual-Energy CT: New Horizon in Medical Imaging [J].
Goo, Hyun Woo ;
Goo, Jin Mo .
KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (04) :555-569
[10]   Ceramic scintillators [J].
Greskovich, C ;
Duclos, S .
ANNUAL REVIEW OF MATERIALS SCIENCE, 1997, 27 :69-88