Capacitive measurement of infusion fluid volume

被引:0
|
作者
Dziadak, Bogdan [1 ]
Graffstein, Piotr [2 ]
机构
[1] Warsaw Univ Technol, Inst Theory Elect Engn Measurement & Informat Syst, Warsaw, Poland
[2] Inst Elect Engn, Lukasiewicz Res Network, Ul Mieczyslawa Pozaryskiego 28, PL-04703 Warsaw, Poland
来源
PRZEGLAD ELEKTROTECHNICZNY | 2024年 / 100卷 / 05期
关键词
infusion fluid; capacity measurement; volume measurement; NE555;
D O I
10.15199/48.2024.05.49
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
.The article presents a system for monitoring fluid during intravenous infusion. The system is based on a capacitive method in which the change in electrical capacitance of the infusion container is converted into the current volume of the fluid. The capacitor pads of the measurement system are integrated into the housing of the container, facilitating easy installation in hospital conditions without significantly altering the dimensions of the infusion set. The system has been tested with target infusion fluids, and the volume measurement error within the entire measurement range is less than 6%. The current fluid volume is displayed on an LED strip to quickly inform the personnel, and the system also has the capability to provide an alarm for low fluid levels.
引用
收藏
页码:258 / 262
页数:5
相关论文
共 50 条
  • [1] Application of the Balloon Snake in the Volume Measurement of Subretinal Fluid in Central Serous Chorioretinopathy
    Chien, Hung-Jen
    Chang, Chia-Jen
    SEMINARS IN OPHTHALMOLOGY, 2019, 34 (06) : 403 - 408
  • [2] Non-invasive measurement of normal skin impedance for determining the volume of the transdermally extracted interstitial fluid
    Li, Dachao
    Pu, Zhihua
    Liang, Wenshuai
    Liu, Tongkun
    Wang, Ridong
    Yu, Haixia
    Xu, Kexin
    MEASUREMENT, 2015, 62 : 215 - 221
  • [3] Magnetofluidic sensor for volume measurement
    Stanci, A
    Iusan, V
    Buioca, CD
    SENSORS AND ACTUATORS A-PHYSICAL, 2000, 84 (03) : 246 - 249
  • [4] Measurement of volume and flow in gases
    Mandal, Nanda Gopal
    ANAESTHESIA AND INTENSIVE CARE MEDICINE, 2009, 10 (01) : 52 - 56
  • [5] Tumor volume measurement and volume measurement comparison plug-ins for VolView using ITK
    Popa, Teo
    Ibanez, Luis
    Levy, Elliot
    White, Amy
    Bruno, Jill
    Cleary, Kevin
    MEDICAL IMAGING 2006: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2006, 6141
  • [6] Optical System for Drop Volume Measurement
    Norgia, Michele
    Magnani, Alessandro
    Melchionni, Dario
    Pesatori, Alessandro
    2014 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC) PROCEEDINGS, 2014, : 1322 - 1325
  • [7] Influence of Bubbles in Diesel Volume Measurement
    Guoyu Chen
    Tiange Xiao
    Yong Wan
    Houlin Hu
    Huanhuan Lin
    Liling Chen
    Guixiong Liu
    Yanhua Peng
    Chemistry and Technology of Fuels and Oils, 2022, 57 : 1005 - 1012
  • [8] Measurement of tumor volume in acoustic neurinomas
    Wolf, SR
    Triebswetter, F
    Wigand, ME
    ACOUSTIC NEUROMA: CONSENSUS ON SYSTEMS FOR REPORTING RESULTS, 2003, 10 : 49 - 53
  • [9] High precision measurement of cartridge volume
    Guo, Hao
    Feng, Jin-Liang
    Zhang, Yao-Yu
    Cai, Cun-Liang
    Li, Si-Qi
    Binggong Xuebao/Acta Armamentarii, 2015, 36 (04): : 758 - 762
  • [10] Nasopharyngeal carcinoma tumor volume measurement
    Chong, VFH
    Zhou, JY
    Khoo, JBK
    Huang, J
    Lim, TK
    RADIOLOGY, 2004, 231 (03) : 914 - 921