Electrical Impedance Tomography-based Ventilation Patterns in Patients after Major Surgery

被引:6
|
作者
Iwata, Hirofumi [1 ]
Yoshida, Takeshi [1 ]
Hoshino, Taiki [1 ]
Aiyama, Yuki [1 ]
Maezawa, Takashi [1 ]
Hashimoto, Haruka [1 ]
Koyama, Yukiko [1 ]
Yamada, Tomomi [2 ]
Fujino, Yuji [1 ]
机构
[1] Osaka Univ, Dept Anesthesiol & Intens Care Med, Grad Sch Med, Suita, Osaka, Japan
[2] Osaka Univ Hosp, Dept Med Innovat Data Coordinating Ctr, Suita, Osaka, Japan
基金
日本学术振兴会;
关键词
electrical impedance tomography; ventilation; postoperative pulmonary complications; POSTOPERATIVE PULMONARY COMPLICATIONS; MECHANICAL VENTILATION; DRIVING PRESSURE; ATELECTASIS; STAY;
D O I
10.1164/rccm.202309-1658OC
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Rationale: General anesthesia and mechanical ventilation have negative impacts on the respiratory system, causing heterogeneous distribution of lung aeration, but little is known about the ventilation patterns of postoperative patients and their association with clinical outcomes. Objectives: To clarify the phenotypes of ventilation patterns along a gravitational direction after surgery by using electrical impedance tomography (EIT) and to evaluate their association with postoperative pulmonary complications (PPCs) and other relevant clinical outcomes. Methods: Adult postoperative patients at high risk for PPCs, receiving mechanical ventilation on ICU admission (N= 128), were prospectively enrolled between November 18, 2021 and July 18, 2022. PPCs were prospectively scored until hospital discharge, and their association with phenotypes of ventilation patterns was studied. The secondary outcomes were the times to wean from mechanical ventilation and oxygen use and the length of ICU stay. Measurements and Main Results: Three phenotypes of ventilation patterns were revealed by EIT: phenotype 1 (32% [n = 41], a predominance of ventral ventilation), phenotype 2 (41% [n = 52], homogeneous ventilation), and phenotype 3 (27% [n = 35], a predominance of dorsal ventilation). The median PPC score was higher in phenotype 1 and phenotype 3 than in phenotype 2. The median time to wean from mechanical ventilation was longer in phenotype 1 versus phenotype 2. The median duration of ICU stay was longer in phenotype 1 versus phenotype 2. The median time to wean from oxygen use was longer in phenotype 1 and phenotype 3 than in phenotype 2. Conclusions: Inhomogeneous ventilation patterns revealed by EIT on ICU admission were associated with PPCs, delayed weaning from mechanical ventilation and oxygen use, and a longer ICU stay.
引用
收藏
页码:1328 / 1337
页数:10
相关论文
共 50 条
  • [41] Electrical Impedance Tomography-guided PEEP Titration in Patients Undergoing Laparoscopic Abdominal Surgery
    He, Xingying
    Jiang, Jingjing
    Liu, Yuli
    Xu, Haitao
    Zhou, Shuangqiong
    Yang, Shibo
    Shi, Xueyin
    Yuan, Hongbin
    MEDICINE, 2016, 95 (14)
  • [42] A Coupled Double-Layer Electrical Impedance Tomography-Based Sensing Skin for Pressure and Leak Detection
    Kuusela, Petri
    Seppanen, Aku
    SENSORS, 2024, 24 (13)
  • [43] Assessment of distribution of ventilation by electrical impedance tomography in standing horses
    Ambrisko, T. D.
    Schramel, J. P.
    Adler, A.
    Kutasi, O.
    Makra, Z.
    Moens, Y. P. S.
    PHYSIOLOGICAL MEASUREMENT, 2016, 37 (02) : 175 - 186
  • [44] Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring?
    Leonhardt, Steffen
    Lachmann, Burkhard
    INTENSIVE CARE MEDICINE, 2012, 38 (12) : 1917 - 1929
  • [45] Regional Ventilation-Perfusion Matching by Electrical Impedance Tomography After Single Lung Transplant
    Scaravilli, Vittorio
    Colombo, Sebastiano Maria
    Fumagalli, Jacopo
    Mauri, Tommaso
    Zanella, Alberto
    Rosso, Lorenzo
    Grasselli, Giacomo
    ASAIO JOURNAL, 2024, 70 (03) : e49 - e52
  • [46] Monitoring of tidal ventilation by electrical impedance tomography in anaesthetised horses
    Mosing, M.
    Waldmann, A. D.
    Raisis, A.
    Boehm, S. H.
    Drynan, E.
    Wilson, K.
    EQUINE VETERINARY JOURNAL, 2019, 51 (02) : 222 - 226
  • [47] Regional ventilation in cystic fibrosis measured by electrical impedance tomography
    Zhao, Zhanqi
    Fischer, Rainald
    Frerichs, Inez
    Mueller-Lisse, Ullrich
    Moeller, Knut
    JOURNAL OF CYSTIC FIBROSIS, 2012, 11 (05) : 412 - 418
  • [48] Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring?
    Steffen Leonhardt
    Burkhard Lachmann
    Intensive Care Medicine, 2012, 38 : 1917 - 1929
  • [49] A simulation study on the ventilation inhomogeneity measured with Electrical Impedance Tomography
    Schullcke, B.
    Krueger-Ziolek, S.
    Gong, B.
    Moeller, K.
    IFAC PAPERSONLINE, 2017, 50 (01): : 8781 - 8785
  • [50] Impact of cardiac output and alveolar ventilation in estimating ventilation/perfusion mismatch in ARDS using electrical impedance tomography
    Samuel Tuffet
    Tommaso Maraffi
    Matthieu Lacheny
    François Perier
    Anne-Fleur Haudebourg
    Mohamed Ahmed Boujelben
    Glasiele Alcala
    Armand Mekontso-Dessap
    Guillaume Carteaux
    Critical Care, 27