PURPOSE: To apply failure modes and effects analysis to high-dose-rate treatment planning to identify the most likely and significant sources of error in the process. METHODS: We have made a list of 25 failure modes grouped into six categories (imaging, catheter reconstruction, dwell position activity, dose points/normalization, optimization/dose, and evaluation). Each mode was rated on a one to five scale for severity, likelihood of occurrence, and probability of escaping detection. An overall ranking was formed from the product of the three scores. The authors assigned scores independently and the resulting rankings were averaged. We also analyzed 44 reported medical events related to high-dose-rate treatment planning listed on the Nuclear Regulatory Commission Web site and compared them with our own rankings. RESULTS: Failure modes associated with image sets, catheter reconstruction, indexer length, and incorrect dose points had the highest ranking in our analysis (scores higher than 20). The most often cited failure modes in the Nuclear Regulatory Commission reports examined were indexer length (20/44) and incorrect dose points (6/44). Several of our high-ranking modes are not associated with reported events. CONCLUSION: It is a useful exercise to identify failure modes locally and analyze the efficacy of the local quality assurance program. Comparison with nationally reported failures can help direct the local analysis, but the absence or small number of reports for failure modes with a high score may be owing to low detectability. Such modes obviously cannot be ignored. (C) 2013 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:382 / 386
页数:5
相关论文
共 50 条
  • [31] Evaluation of bi-objective treatment planning for high-dose-rate prostate brachytherapy-A retrospective observer study
    Maree, Stefanus C.
    Luong, Ngoc Hoang
    Kooreman, Ernst S.
    van Wieringen, Niek
    Bel, Arjan
    Hinnen, Karel A.
    Westerveld, Henrike
    Pieters, Bradley R.
    Bosman, Peter A. N.
    Alderliesten, Tanja
    BRACHYTHERAPY, 2019, 18 (03) : 396 - 403
  • [32] Salvage treatment with high-dose-rate brachytherapy for isolated vaginal endometrial cancer recurrence
    Petignat, Patrick
    Jolicoeur, Marjory
    Alobaid, Abdulaziz
    Drouin, Pierre
    Gauthier, Philippe
    Provencher, Diane
    Donath, David
    Nguyen, Thu Van
    GYNECOLOGIC ONCOLOGY, 2006, 101 (03) : 445 - 449
  • [33] Analysis of dose distribution between contemporary and standard planning in high-dose-rate endobronchial brachytherapy based on three-dimensional imaging
    Sawicki, Marcin
    Lyczek, Jaroslaw
    Szutkowski, Zbigniew
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2019, 11 (05) : 462 - 468
  • [34] A method for verifying treatment times for simple high-dose-rate endobronchial brachytherapy procedures
    Miller, AV
    Davis, MG
    Horton, JL
    MEDICAL PHYSICS, 1996, 23 (11) : 1903 - 1908
  • [35] Dosimetry modeling for focal high-dose-rate prostate brachytherapy
    Mason, Josh
    Al-Qaisieh, Bashar
    Bownes, Peter
    Thwaites, David
    Henry, Ann
    BRACHYTHERAPY, 2014, 13 (06) : 611 - 617
  • [36] Salvage high-dose-rate brachytherapy for recurrent prostate cancer
    Kaprin, A. D.
    Ivanov, S. A.
    Karyakin, O. B.
    Obuhov, A. A.
    Biryukov, V. A.
    Borysheva, N. B.
    Sanin, D. B.
    Lepilina, O. G.
    Smolkin, A. L.
    Dem'yanovich, A., V
    Minaeva, N. G.
    Mikhaylovskiy, N., V
    ONKOUROLOGIYA, 2020, 16 (04): : 112 - 119
  • [37] On the use of C-arm fluoroscopy for treatment planning in high dose rate brachytherapy
    Liu, LZ
    Bassano, DA
    Prasad, SC
    Keshler, BL
    Hahn, SS
    MEDICAL PHYSICS, 2003, 30 (09) : 2297 - 2302
  • [38] Urinary incontinence after high-dose-rate brachytherapy boost treatment for prostate cancer
    Kragelj, Borut
    Zaletel-Kragelj, Lijana
    BRACHYTHERAPY, 2016, 15 (04) : 442 - 449
  • [39] RADIOTHERAPY FOR CERVICAL-CANCER WITH HIGH-DOSE-RATE BRACHYTHERAPY - CORRELATION BETWEEN TUMOR SIZE, DOSE AND FAILURE
    ITO, H
    KUTUKI, S
    NISHIGUCHI, I
    SHIGEMATSU, N
    KURIBAYASHI, T
    UEMATSU, M
    NAKAYAMA, T
    KA, WJ
    TAKEMASA, K
    ANDO, Y
    KUBO, A
    RADIOTHERAPY AND ONCOLOGY, 1994, 31 (03) : 240 - 247
  • [40] A novel approach for Venezia ovoid commissioning with a comprehensive analysis of source positions in high-dose-rate brachytherapy
    Li, Xing
    Huang, Jessica
    Rahimi, Robabeh
    Zhao, Hui
    Kunz, Jeremy
    Suneja, Gita
    Su, Fan -Chi
    BRACHYTHERAPY, 2023, 22 (01) : 93 - 100