Application of low-dose CT in image-guided radiotherapy based on CT-linac

被引:1
|
作者
Yang, Jingru [1 ]
Liang, Yongguang [1 ]
Yang, Bo [1 ]
Zhang, Wenjun [1 ]
Leng, Lingxuan [1 ]
Wang, Jiaxin [1 ]
Zhu, Heling [1 ]
He, Shumeng [2 ]
Liao, Can [3 ]
Qiu, Jie [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Peking Union Med Coll Hosp, Dept Radiat Oncol, Beijing, Peoples R China
[2] United Imaging Res Inst Intelligent Imaging, Beijing 100094, Peoples R China
[3] Shanghai United Imaging Healthcare Co Ltd, Shanghai 201807, Peoples R China
关键词
Fan-beam CT; Image-guided radiotherapy; Low-dose CT; Image quality; ITERATIVE RECONSTRUCTION;
D O I
10.1016/j.jrras.2024.101034
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Image-guided radiation therapy (IGRT) involves online medical imaging systems. CT-guided imaging systems not only have high image quality but also allow for direct dose calculation for radiation treatment planning, making it an ideal image guidance method. However, the imaging dose from CT images may pose additional risks to patients. Therefore, the aim of this study was to evaluate the substitutability of using low-dose CT images for IGRT. Materials and methods: The CIRS-062 phantom, Catphan 604 phantom, CT dose index phantom were used to obtain the conversion curves of the CT values to the relative electron density (RED), image quality, and imaging dose of the low-dose CT, respectively. Catphan 604 phantom and clinical cases were used to analyze the feasibility of low-dose CT for dose calculation. Results: No significant differences were observed in the conversion curves of CT values to RED between low-dose and standard-dose CT scans. In terms of image quality, the geometric distortion of low-dose CT was 0.1 mm, which was the same as that of standard-dose CT. The spatial resolution of the images was 0.42-0.45 lp/mm, which was slightly better than that of standard-dose CT (0.40-0.44 lp/mm). The image uniformity was 98.4%- 99.0%, slightly worse than that of standard-dose CT (99.4%-99.5%), but still met the clinical requirements. The imaging dose was 2.96-4.62 mGy, significantly lower than that of standard-dose CT (9.71-26.68 mGy), especially for head CT protocol. In terms of dose calculation, the gamma passing rates were not lower than 98% for either the Catphan 604 phantom or the clinical cases. Conclusion: Low-dose CT can significantly reduce the imaging dose without a significant loss of image quality and can be directly used for dose calculation in radiation treatment planning, providing a safer and more effective image guidance method for IGRT.
引用
收藏
页数:7
相关论文
共 50 条
  • [11] Statistical image reconstruction for low-dose CT using nonlocal means-based regularization
    Zhang, Hao
    Ma, Jianhua
    Wang, Jing
    Liu, Yan
    Lu, Hongbing
    Liang, Zhengrong
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2014, 38 (06) : 423 - 435
  • [12] Low-dose CT image reconstruction using gain intervention-based dictionary learning
    Pathak, Yadunath
    Arya, K., V
    Tiwari, Shailendra
    MODERN PHYSICS LETTERS B, 2018, 32 (14):
  • [13] Low-dose CT image restoration based on noise prior regression network
    Jin, Yan
    Jiang, Zhiwei
    Huang, Mengjia
    Xue, Zhizhong
    VISUAL COMPUTER, 2023, 39 (01) : 459 - 471
  • [14] Low-dose CT image restoration based on noise prior regression network
    Yan Jin
    Zhiwei Jiang
    Mengjia Huang
    Zhizhong Xue
    The Visual Computer, 2023, 39 : 459 - 471
  • [15] GDAFormer: Gradient-guided Dual Attention Transformer for Low-Dose CT image denoising
    Jiang, Guowei
    Luo, Ting
    Xu, Haiyong
    Nie, Sheng
    Song, Yang
    He, Zhouyan
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 94
  • [16] Iterative reconstruction of low-dose CT based on differential sparse
    Lu, Siyu
    Yang, Bo
    Xiao, Ye
    Liu, Shan
    Liu, Mingzhe
    Yin, Lirong
    Zheng, Wenfeng
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2023, 79
  • [17] Database-assisted low-dose CT image restoration
    Xu, Wei
    Ha, Sungsoo
    Mueller, Klaus
    MEDICAL PHYSICS, 2013, 40 (03)
  • [18] Perceptual CT Loss: Implementing CT Image Specific Perceptual Loss for CNN-Based Low-Dose CT Denoiser
    Han, Minah
    Shim, Hyunjung
    Baek, Jongduk
    IEEE ACCESS, 2022, 10 : 62412 - 62422
  • [19] Low-Dose CT Image Reconstruction With a Deep Learning Prior
    Park, Hyoung Suk
    Kim, Kyungsang
    Jeon, Kiwan
    IEEE ACCESS, 2020, 8 : 158647 - 158655
  • [20] Low-dose CT: technique, reading methods and image interpretation
    Rampinelli, Cristiano
    Origgi, Daniela
    Bellomi, Massimo
    CANCER IMAGING, 2013, 12 (03): : 548 - 556