The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study

被引:0
|
作者
Ravenscroft, L. [1 ]
Baker, L. [2 ]
机构
[1] Univ Hosp Derby & Burton NHS Fdn Trust, Derby, Staffs, England
[2] Nottingham Univ Hosp NHS Trust, Res & Dev Imaging Support Unit, Nottingham, England
关键词
Computed tomography; Miscentering; Radiation dose; Brain; Radiographic positioning; TUBE CURRENT MODULATION; IMAGE NOISE; PATIENT; MDCT; POSTEROANTERIOR; EXPOSURE; SYSTEMS; IMPACT;
D O I
10.1016/j.radi.2024.09.051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Computed Tomography (CT) chest, abdomen and pelvis research demonstrates a relationship between vertical phantom positioning and radiation dose. Moving the phantom closer or further from the x-ray source results in magnification or minimisation of the localiser. As automatic tube current modulation (ATCM) algorithms use localisers to estimate patient size and calculate required tube current, magnification or minimisation results in the incorrect provision of radiation dose. Radiation dose changes also depend on localiser orientation, changes with anteroposterior (AP) and posteroanterior (PA) localisers demonstrating an inverse relationship. However, within CT head literature often attributes radiation dose changes on impaired function of the bow-tie filter instead. The current study investigated the role of miscentering on ATCM function within CT head, paying particular attention to localiser orientation.<br /> Methods: Head scanning was performed with an anthropomorphic phantom at the isocentre, alongside ten vertically miscentered positions. This was performed three times, with an AP, PA and lateral localiser. CT dose index values at each miscentered level were compared across conditions. Results: Vertical miscentering altered radiation dose in both AP and PA conditions, radiation dose linearly increasing (up to 17.05%) when positioning the phantom closer to the x-ray source and decreasing when positioning away (up to -13.13%). Changes across AP and PA conditions demonstrated an inverse relationship. Radiation dose was unaffected in the lateral condition.<br /> Conclusions: Miscentering during CT head alters ATCM function due to magnification/minimisation of the localiser image, causing ATCM algorithms to misinterpret patient size and miscalculate required tube current.<br /> Implications for practice: Radiographers should be accurate when centering for CT head, avoiding any potential radiation dose changes. Further research into vertical miscentering and image quality during CT head is recommended. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those
引用
收藏
页码:1517 / 1523
页数:7
相关论文
共 50 条
  • [31] Influence of the adaptive iterative dose reduction 3D algorithm on the detectability of low-contrast lesions and radiation dose repeatability in abdominal computed tomography: a phantom study
    Yoon, Jeong Hee
    Lee, Jeong Min
    Hur, Bo Yun
    Baek, Jeehyun
    Shim, Hackjoon
    Han, Joon Koo
    Choi, Byung Ihn
    ABDOMINAL IMAGING, 2015, 40 (06): : 1843 - 1852
  • [32] Radiation dose evaluation in head and neck MDCT examinations with a 6-year-old child anthropomorphic phantom
    Yamauchi-Kawaura, Chiyo
    Fujii, Keisuke
    Aoyama, Takahiko
    Koyama, Shuji
    Yamauchi, Masato
    PEDIATRIC RADIOLOGY, 2010, 40 (07) : 1206 - 1214
  • [33] Assessment of radiation dose delivered and volume measurement by low- and high-dose diagnostic computed tomography: Anthropomorphic liver phantom study
    Kheruka, Subhash Chand
    Ora, Manish
    Chaudhary, Shivani
    Gambhir, Sanjay
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2020, 35 (04): : 310 - 314
  • [34] Relationship between assistant's lens exposure and dose information during computed tomography examinations
    Ito, Hajime
    Matsubara, Kosuke
    Kobayashi, Ikuo
    Shimakawa, Yurie
    Murayama, Daichi
    Sakai, Takayuki
    Isobe, Tomonori
    Yanagawa, Noriyuki
    Ochi, Shigehiro
    JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (02)
  • [35] Pediatric head computed tomography with advanced modeled iterative reconstruction: focus on image quality and reduction of radiation dose
    Cho, Hyun-Hae
    Lee, So Mi
    You, Sun Kyoung
    PEDIATRIC RADIOLOGY, 2020, 50 (02) : 242 - 251
  • [36] ESTIMATING RADIATION DOSE TO MAJOR ORGANS IN DENTAL X-RAY EXAMINATIONS: A PHANTOM STUDY
    Li, Yong
    Huang, Bingsheng
    Cao, Jun
    Fang, Tianqi
    Liu, Guoqing
    Li, Xuguang
    Wu, Jiabao
    RADIATION PROTECTION DOSIMETRY, 2020, 192 (03) : 328 - 334
  • [37] Evaluation of Radiation Dose in Coronary Computed Tomography-Angiography Examinations: Using Organ dose and Effective dose
    Shibata, Hideki
    Kondo, Yuya
    Fukada, Shinji
    Asada, Yasuki
    Kozawa, Isao
    RADIATION PHYSICS AND CHEMISTRY, 2019, 158 : 218 - 221
  • [38] Patient dose reduction for a localizer radiograph with an additional tin filter in chest-abdomen-pelvis, spine, and head computed tomography examinations
    Takemitsu, Masaki
    Takegami, Kazuki
    Kudomi, Shohei
    Kawanami, Kaito
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (02) : 160 - 167
  • [39] Pediatric head computed tomography with advanced modeled iterative reconstruction: focus on image quality and reduction of radiation dose
    Hyun-Hae Cho
    So Mi Lee
    Sun Kyoung You
    Pediatric Radiology, 2020, 50 : 242 - 251
  • [40] Regional survey of image quality and radiation dose in computed tomography examinations in Saudi Arabia
    Saeed, M. K.
    Alzoubi, A. S.
    Al-Qahtani, J. M.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 279 - 283