The influence of miscentering on radiation dose during computed tomography head examinations and the role of localiser orientation: A phantom study

被引:0
|
作者
Ravenscroft, L. [1 ]
Baker, L. [2 ]
机构
[1] Univ Hosp Derby & Burton NHS Fdn Trust, Derby, Staffs, England
[2] Nottingham Univ Hosp NHS Trust, Res & Dev Imaging Support Unit, Nottingham, England
关键词
Computed tomography; Miscentering; Radiation dose; Brain; Radiographic positioning; TUBE CURRENT MODULATION; IMAGE NOISE; PATIENT; MDCT; POSTEROANTERIOR; EXPOSURE; SYSTEMS; IMPACT;
D O I
10.1016/j.radi.2024.09.051
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Computed Tomography (CT) chest, abdomen and pelvis research demonstrates a relationship between vertical phantom positioning and radiation dose. Moving the phantom closer or further from the x-ray source results in magnification or minimisation of the localiser. As automatic tube current modulation (ATCM) algorithms use localisers to estimate patient size and calculate required tube current, magnification or minimisation results in the incorrect provision of radiation dose. Radiation dose changes also depend on localiser orientation, changes with anteroposterior (AP) and posteroanterior (PA) localisers demonstrating an inverse relationship. However, within CT head literature often attributes radiation dose changes on impaired function of the bow-tie filter instead. The current study investigated the role of miscentering on ATCM function within CT head, paying particular attention to localiser orientation.<br /> Methods: Head scanning was performed with an anthropomorphic phantom at the isocentre, alongside ten vertically miscentered positions. This was performed three times, with an AP, PA and lateral localiser. CT dose index values at each miscentered level were compared across conditions. Results: Vertical miscentering altered radiation dose in both AP and PA conditions, radiation dose linearly increasing (up to 17.05%) when positioning the phantom closer to the x-ray source and decreasing when positioning away (up to -13.13%). Changes across AP and PA conditions demonstrated an inverse relationship. Radiation dose was unaffected in the lateral condition.<br /> Conclusions: Miscentering during CT head alters ATCM function due to magnification/minimisation of the localiser image, causing ATCM algorithms to misinterpret patient size and miscalculate required tube current.<br /> Implications for practice: Radiographers should be accurate when centering for CT head, avoiding any potential radiation dose changes. Further research into vertical miscentering and image quality during CT head is recommended. (c) 2024 The College of Radiographers. Published by Elsevier Ltd. All rights are reserved, including those
引用
收藏
页码:1517 / 1523
页数:7
相关论文
共 50 条
  • [21] Feasibility of using the computed tomography dose indices to estimate radiation dose to partially and fully irradiated brains in pediatric neuroradiology examinations
    Januzis, Natalie
    Giao Nguyen
    Frush, Donald P.
    Hoang, Jenny K.
    Lowry, Carolyn
    Yoshizumi, Terry T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (14) : 5699 - 5710
  • [22] Determination of radiation dose during computed tomography examinations in southern Saudi Arabian hospitals using size-specific dose estimates
    Saeed, Mohammed K.
    Asiri, Abdullah Ali M.
    MAEJO INTERNATIONAL JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 15 (02) : 147 - 158
  • [23] Assessing Radiation Dosage in Pediatric Head and Neck Computed Tomography Examinations During COVID-19 in a Tertiary Hospital in Saudi Arabia, Jeddah
    Khafaji, Mawya
    AlNajjar, Abdulrahman Y.
    Matbouli, Sultan
    Alnahdi, Yasser A.
    Meriky, Lama H.
    Hagi, Sara
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (01)
  • [24] A survey of the pediatric radiation doses during multiphase abdominal computed tomography examinations
    Sulieman, A.
    Adam, H.
    Tamam, N.
    Alkhorayef, M.
    Alhailiy, A.
    Alghamdi, S.
    Elnour, A.
    Alomair, O.
    Alashban, Y.
    Khandaker, M. U.
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 188
  • [25] Automated Low-Kilovoltage Selection in Pediatric Computed Tomography Angiography Phantom Study Evaluating Effects on Radiation Dose and Image Quality
    Siegel, Marilyn Joy
    Ramirez-Giraldo, Juan Carlos
    Hildebolt, Charles
    Bradley, David
    Schmidt, Bernhard
    INVESTIGATIVE RADIOLOGY, 2013, 48 (08) : 584 - 589
  • [26] Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans
    Buchhave, E. O.
    Rasmussen, C. D.
    Lentz, R.
    Frederiksen, E. H.
    Kipp, J. O.
    Leftes, M. O.
    RADIOGRAPHY, 2024, 30 (04) : 1035 - 1040
  • [27] Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates
    Mohamed M. Abuzaid
    Wiam Elshami
    Huseyin Ozan Tekin
    Hatem Ghonim
    Mona Shawki
    Dina H. Salama
    Insights into Imaging, 11
  • [28] Influence of Arm Positioning on Radiation Dose for Whole Body Computed Tomography in Trauma Patients
    Bayer, Joerg
    Pache, Gregor
    Strohm, Peter C.
    Zwingmann, Joern
    Blanke, Philipp
    Baumann, Tobias
    Suedkamp, Norbert P.
    Hammer, Thorsten
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 70 (04): : 900 - 905
  • [29] Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study
    Lukas Lehmkuhl
    Dieter Gosch
    H. D. Nagel
    Patrick Stumpp
    Thomas Kahn
    Matthias Gutberlet
    European Radiology, 2010, 20 : 2116 - 2125
  • [30] The effective and collective dose to patients undergoing abdominopelvic and trunk computed tomography examinations: A Belgian multicentre study
    De Roo, Bieke
    Stichelbaut, Nele
    Hoste, Pieter
    Verstraete, Koenraad
    Bacher, Klaus
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 111 : 81 - 87