Assessing the impact of nodule features and software algorithm on pulmonary nodule measurement uncertainty for nodules sized 20 mm or less

被引:4
作者
Jirapatnakul, Artit [1 ]
Yip, Rowena [1 ]
Myers, Kyle J. [2 ]
Cai, Siyang [1 ]
Henschke, Claudia I. [1 ]
Yankelevitz, David [1 ]
机构
[1] Icahn Sch Med Mt Sinai, Dept Diagnost Mol & Intervent Radiol, 1 Gustave L Levy Pl,Box 1234, New York, NY 10029 USA
[2] Puente Solut LLC, Phoenix, AZ USA
关键词
Measurement uncertainty; measurement variability; pulmonary nodule; VOLUMETRIC MEASUREMENT; CT; VARIABILITY; LEVEL; TIME;
D O I
10.21037/qims-23-1501
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Measurements are not exact, so that if a measurement is repeated, one would get a different value each time. The spread of these values is the measurement uncertainty. Understanding measurement uncertainty of pulmonary nodules is important for proper interpretation of size and growth measurements. Larger amounts of measurement uncertainty may require longer follow-up intervals to be confident that any observed growth is due to actual growth rather than measurement uncertainty. We examined the influence of nodule features and software algorithm on measurement uncertainty of small, solid pulmonary nodules. Methods: Volumes of 107 nodules were measured on 4-6 repeated computed tomography (CT) scans (Siemens Somatom AS, 100 kVp, 120 mA, 1.0 mm slice thickness reconstruction) prospectively obtained during CT-guided fine needle aspiration biopsy between 2015-2021 at Department of Diagnostic, Molecular, and Interventional Radiology in Icahn School of Medicine at Mount Sinai, using two different automated volumetric algorithms. For each, the coefficient of variation (standard deviation divided by the mean) of nodule volume measurements was determined. The following features were considered: diameter, location, vessel and pleural attachments, nodule surface area, and extent of the nodule in the three acquisition dimensions of the scanner. Results: Median volume of 107 nodules was 515.23 and 535.53 mm(3) for algorithm #1 and #2, respectively with excellent agreement (intraclass correlation coefficient =0.98). Median coefficient of variation of nodule volume was low for the two algorithms, but significantly different (4.6% vs. 8.7%, P<0.001). Both algorithms had a trend of decreasing coefficient of variation of nodule volume with increasing nodule diameter, though only significant for algorithm #2. Coefficient of variation of nodule volume was significantly associated with nodule volume (P=0.02), attachment to blood vessels (P=0.02), and nodule surface area (P=0.001) for algorithm #2 using a multiple linear regression model. Correlation between the coefficient of variation (CoV) of nodule volume and the CoV of the x, y, z measurements for algorithm #1 were 0.29 (P=0.0021), 0.25 (P=0.009), and 0.80 (P<0.001) respectively, and for algorithm #2, 0.46 (P<0.001), 0.52 (P<0.001), and 0.58 (P<0.001), respectively. Conclusions: Even in the best-case scenario represented in this study, using the same measurement algorithm, scanner, and scanning protocol, considerable measurement uncertainty exists in nodule volume measurement for nodules sized 20 mm or less. We found that measurement uncertainty was affected by interactions between nodule volume, algorithm, and shape complexity.
引用
收藏
页码:5057 / 5071
页数:15
相关论文
共 25 条
[1]  
American Association of Physicists in Medicine, 2023, Lung Cancer Screening CT Protocols Version 6.0
[2]   Algorithm Variability in the Estimation of Lung Nodule Volume From Phantom CT Scans: Results of the QIBA 3A Public Challenge [J].
Athelogou, Maria ;
Kim, Hyun J. ;
Dima, Alden ;
Obuchowski, Nancy ;
Peskin, Adele ;
Gavrielides, Marios A. ;
Petrick, Nicholas ;
Saiprasad, Ganesh ;
Colditz, Dirk Colditz ;
Beaumont, Hubert ;
Oubel, Estanislao ;
Tan, Yongqiang ;
Zhao, Binsheng ;
Kuhnigk, Jan-Martin ;
Moltz, Jan Hendrik ;
Orieux, Guillaume ;
Gillies, Robert J. ;
Gu, Yuhua ;
Mantri, Ninad ;
Goldmacher, Gregory ;
Zhang, Luduan ;
Vega, Emilio ;
Bloom, Michael ;
Jarecha, Rudresh ;
Soza, Grzegorz ;
Tietjen, Christian ;
Takeguchi, Tomoyuki ;
Yamagata, Hitoshi ;
Peterson, Sam ;
Masoud, Osama ;
Buckler, Andrew J. .
ACADEMIC RADIOLOGY, 2016, 23 (08) :940-952
[3]   Quantitative CT: technique dependence of volume estimation on pulmonary nodules [J].
Chen, Baiyu ;
Barnhart, Huiman ;
Richard, Samuel ;
Colsher, James ;
Amurao, Maxwell ;
Samei, Ehsan .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (05) :1335-1348
[4]  
CT Volumetry Technical Committee, 2018, QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening
[5]   Information-Theoretic Approach for Analyzing Bias and Variance in Lung Nodule Size Estimation With CT: A Phantom Study [J].
Gavrielides, Marios A. ;
Zeng, Rongping ;
Kinnard, Lisa M. ;
Myers, Kyle J. ;
Petrick, Nicholas .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (10) :1795-1807
[6]   Pulmonary nodules: Interscan variability of semiautomated volume measurements with multisection CT-influence of inspiration level, nodule size, and segmentation performance [J].
Gietema, Hester A. ;
Schaefer-Prokop, Cornelia M. ;
Mali, Willem P. T. M. ;
Groenewegen, Gerard ;
Prokop, Mathias .
RADIOLOGY, 2007, 245 (03) :888-894
[7]   Pulmonary nodules detected at lung cancer screening: Interobserver variability of semiautomated volume measurements [J].
Gietema, Hester A. ;
Wang, Ying ;
Xu, Dongming ;
van Klaveren, Rob J. ;
deKoning, Harry ;
Scholten, Ernst ;
Verschakelen, Johny ;
Kohl, Gerhard ;
Oudkerk, Matthijs ;
Prokop, Mathias .
RADIOLOGY, 2006, 241 (01) :251-257
[8]   Inherent variability of CT lung nodule measurements in vivo using semiautomated volumetric measurements [J].
Goodman, LR ;
Gulsun, M ;
Washington, L ;
Nagy, PG ;
Piacsek, KL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (04) :989-994
[9]   Influence of lung nodule margin on volume- and diameter-based reader variability in CT lung cancer screening [J].
Han, Daiwei ;
Heuvelmans, Marjolein A. ;
Vliegenthart, Rozemarijn ;
Rook, Mieneke ;
Dorrius, Monique D. ;
De Jonge, Gonda J. ;
Walter, Joan E. ;
Van Ooijen, Peter M. A. ;
De Koning, Harry J. ;
Oudkerk, Matthijs .
BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1090)
[10]   Tumor volume measurement error using computed tomography imaging in a phase II clinical trial in lung cancer [J].
Henschke, Claudia I. ;
Yankelevitz, David F. ;
Yip, Rowena ;
Archer, Venice ;
Zahlmann, Gudrun ;
Krishnan, Karthik ;
Helba, Brian ;
Avila, Ricardo .
JOURNAL OF MEDICAL IMAGING, 2016, 3 (03)