Intraocular Pressure Measurements in Standing, Sitting, Prone, and Supine Positions

被引:2
作者
De Bernardo, Maddalena [1 ]
Cione, Ferdinando [1 ]
De Pascale, Ilaria [1 ]
Pagliarulo, Sergio [1 ]
Rosa, Nicola [1 ]
机构
[1] Univ Salerno, Scuola Med Salernitana, Dept Med Surg & Dent, Eye Unit, I-84081 Salerno, Italy
来源
JOURNAL OF PERSONALIZED MEDICINE | 2024年 / 14卷 / 08期
关键词
intraocular pressure; hand-held tonometer; standing position; supine position; sitting position; normal-tension glaucoma; BODY POSITION; GLAUCOMA PROGRESSION; RISK-FACTOR; FLUCTUATION; PNEUMATONOMETER; TONOMETER; HEAD;
D O I
10.3390/jpm14080826
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
In this study, intraocular pressure (IOP) was measured in sitting, supine, prone, and standing (ST) positions and again five minutes after standing (ST-5) utilizing a Tono-Pen AVIA in 124 eyes of 62 healthy subjects with ages ranging from 21 to 59 years (mean 30 +/- 10 years). In each subject, the average IOP of both eyes was used for the statistical evaluation. The mean IOP difference between the ST and sitting positions was -0.13 +/- 1.63 mmHg (p = 0.548); between ST-5 and sitting, it was 0.53 +/- 1.24 mmHg (p = 0.001); between supine and sitting, it was 1.30 +/- 1.48 mmHg (p < 0.001); between ST and supine, it was -1.43 +/- 1.74 mmHg (p < 0.001); between ST-5 and supine, it was -0.77 +/- 1.59 mmHg (p < 0.001); between prone and supine, it was 2.24 +/- 1.92 mmHg (p < 0.001); between ST and ST-5, it was -0.67 +/- 1.84 mmHg (range: -7.5 to 5 mmHg) (p = 0.007); between prone and ST, it was 3.46 +/- 2.01 mmHg (p < 0.001); between ST-5 and prone, it was -2.46 +/- 1.67 mmHg (p < 0.001); and between sitting and prone, it was -3.22 +/- 1.56 mmHg (p < 0.001). The results show a significant IOP increase in the ST-5 position, suggesting that such measurements need to be performed in an attempt to explain the progression of glaucoma in apparently normal-tension patients.
引用
收藏
页数:13
相关论文
共 38 条
[11]  
Cowley AW., 1996, Primer on the Autonomic Nervous System, V1, P49
[12]   Comparison between dynamic contour tonometry and Goldmann applanation tonometry correcting equations [J].
De Bernardo, Maddalena ;
Casaburi, Claudia ;
De Pascale, Ilaria ;
Capasso, Luigi ;
Cione, Ferdinando ;
Rosa, Nicola .
SCIENTIFIC REPORTS, 2022, 12 (01)
[13]   Intraocular Pressure Measurements in Standing, Sitting, and Supine Position: Comparison between Tono-Pen Avia and Icare Pro Tonometers [J].
De Bernardo, Maddalena ;
Abbinante, Giulia ;
Borrelli, Maria ;
Di Stasi, Margherita ;
Cione, Ferdinando ;
Rosa, Nicola .
JOURNAL OF CLINICAL MEDICINE, 2022, 11 (21)
[14]   Comparison between two devices in the detection of corneal thickness changes after cataract surgery [J].
De Bernardo, Maddalena ;
Vitiello, Livio ;
Abbinante, Giulia ;
De Pascale, Ilaria ;
Capasso, Luigi ;
Marotta, Giuseppe ;
Rosa, Nicola .
SCIENTIFIC REPORTS, 2021, 11 (01)
[15]   Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer [J].
De Bernardo, Maddalena ;
Borrelli, Maria ;
Cembalo, Giovanni ;
Rosa, Nicola .
MEDICINA-LITHUANIA, 2019, 55 (10)
[16]   Intraocular Pressure Evaluation after Myopic Refractive Surgery: A Comparison of Methods in 121 Eyes [J].
De Bernardo, Maddalena ;
Capasso, Luigi ;
Caliendo, Luisa ;
Vosa, Ylenia ;
Rosa, Nicola .
SEMINARS IN OPHTHALMOLOGY, 2016, 31 (03) :233-242
[17]   The Circadian Curve of Intraocular Pressure: Can We Estimate Its Characteristics during Office Hours? [J].
Fogagnolo, Paolo ;
Orzalesi, Nicola ;
Ferreras, Antonio ;
Rossetti, Luca .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (05) :2209-2215
[18]   Reduction of intraocular pressure and glaucoma progression - Results from the early manifest glaucoma trial [J].
Heijl, A ;
Leske, MC ;
Bengtsson, B ;
Hyman, L ;
Bengtsson, B ;
Hussein, M .
ARCHIVES OF OPHTHALMOLOGY, 2002, 120 (10) :1268-1279
[19]   Four Measures of Intraocular Pressure Fluctuation: Which Correlates Most Optimally With Actual Office-hour Readings? [J].
Huang, Rui ;
Ge, Jian ;
Chen, Geng ;
Gao, Xinbo ;
Laties, Alan M. ;
Zhang, Xiulan .
JOURNAL OF GLAUCOMA, 2015, 24 (07) :550-555
[20]   RAPID PNEUMATIC AND MACKAY-MARG APPLANATION TONOMETRY TO EVALUATE POSTURAL EFFECT ON INTRAOCULAR-PRESSURE [J].
JAIN, MR ;
MARMION, VJ .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1976, 60 (10) :687-693