Functional Acetabular Component Positioning During Direct Anterior Approach Hip Arthroplasty Using a Novel Three-Dimensional Virtual Mesh Imaging System With Fluoroscopy

被引:0
作者
Coxe, Francesca R. [1 ]
Jordan, L. Andrew [1 ]
Wong, Zachary P. [1 ]
Spaan, Jonathan C. [1 ]
Ren, Renee [2 ]
Su, Edwin P. [1 ]
机构
[1] Hosp Special Surg, Div Adult Reconstruct & Joint Replacement, 535 E 70th St, New York, NY 10021 USA
[2] Icahn Sch Med Mt Sinai, New York, NY 10029 USA
关键词
total hip arthroplasty; navigation; technology; radiographic outcomes; component positioning; PELVIC TILT; SAFE ZONE; RADIOGRAPHIC CALCULATION; CUP; PLACEMENT; ORIENTATION; DISLOCATION; THA; REPLACEMENT; NAVIGATION;
D O I
10.1016/j.arth.2024.03.042
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Optimal position of total hip arthroplasty (THA) components is critical for joint mechanics and stability. Acetabular component positioning during supine surgery in direct anterior approach (DAA) THA may be different in the standing position postoperatively, which traditional fluoroscopy is unable to predict. A novel 3-dimensional (3D) image analysis technology (IAT) that uses artificial intelligence to measure the tilt and rotation of the pelvis has enabled prediction of component positioning from supine to standing. The purpose of this study was to compare intraoperative fluoroscopy, non-3D-IAT, and 3D-IAT with postoperative standing radiographs to assess the accuracy of component positioning. Methods: From 2022 to 2023, 30 consecutive patients (86.6% women, mean age 59 [range, 55 to 67]) undergoing primary DAA THA with the use of the 3D-IAT were identified. A separate cohort of 148 patients from 2020 to 2021 (85% women, mean age 65 [range, 55 to 69]) who underwent DAA THA with non-3D-IAT was used for comparison. Leg length discrepancy (LLD), cup anteversion, and inclination were manually measured on intraoperative fluoroscopic images and digitally measured using IAT. Follow-up evaluation occurred at 1 month with standing pelvis radiographs measured using Ein Bild Rontgen Analyze-Cup software. Measurements were compared via Wilcoxon signed rank tests where P <= .05 indicates significantly different measurements. Results: Median LLD, inclination, and anteversion measurements via non-3D-IAT and fluoroscopy were significantly different compared to postoperative standing radiographs (P < .001). The 3D-IAT more accurately predicted LLD, abduction, and anteversion, with values not significantly different from post-operative standing measurements (P = .23, P = .93, and P = .36, respectively). Conclusions: The use of the 3D-IAT during DAA THA allowed for the more accurate prediction of acetabular component position in the standing position postoperatively.
引用
收藏
页码:S88 / S96
页数:9
相关论文
共 69 条
  • [1] What Safe Zone? The Vast Majority of Dislocated THAs Are Within the Lewinnek Safe Zone for Acetabular Component Position
    Abdel, Matthew P.
    von Roth, Philipp
    Jennings, Matthew T.
    Hanssen, Arlen D.
    Pagnano, Mark W.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (02) : 386 - 391
  • [2] The Use of Computer Navigation in Total Hip Arthroplasty Is Associated with a Reduced Rate of Revision for Dislocation A Study of 6,912 Navigated THA Procedures from the Australian Orthopaedic Association National Joint Replacement Registry
    Agarwal, Sujit
    Eckhard, Lukas
    Walter, William L.
    Peng, Andrea
    Hatton, Alesha
    Donnelly, Bill
    de Steiger, Richard
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2021, 103 (20) : 1900 - 1905
  • [3] Current Concepts in Acetabular Positioning in Total Hip Arthroplasty
    Bhaskar, Deepu
    Rajpura, Asim
    Board, Tim
    [J]. INDIAN JOURNAL OF ORTHOPAEDICS, 2017, 51 (04) : 386 - 396
  • [4] Biedermann R, 2005, J BONE JOINT SURG BR, V87B, P762, DOI 10.1302/0301-620X.87B6
  • [5] Estimation of an inter-rater intra-class correlation coefficient that overcomes common assumption violations in the assessment of health measurement scales
    Bobak, Carly A.
    Barr, Paul J.
    O'Malley, A. James
    [J]. BMC MEDICAL RESEARCH METHODOLOGY, 2018, 18
  • [6] Comparative Epidemiology of Revision Arthroplasty: Failed THA Poses Greater Clinical and Economic Burdens Than Failed TKA
    Bozic, Kevin J.
    Kamath, Atul F.
    Ong, Kevin
    Lau, Edmund
    Kurtz, Steve
    Chan, Vanessa
    Vail, Thomas P.
    Rubash, Harry
    Berry, Daniel J.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (06) : 2131 - 2138
  • [7] The Use of Computerized Tomography Scans in Elective Knee and Hip ArthroplastydWhat Do They Tell Us and at What Risk?
    'Amore, Taylor
    Klein, Gregg
    Lonner, Jess
    [J]. ARTHROPLASTY TODAY, 2022, 15 : 132 - 138
  • [8] D'Lima D.D., 2000, J BONE JOINT SURG A, V82, P315
  • [9] The influence of pelvic tilt on acetabular orientation and cover: a three-dimensional computerised tomography analysis
    Dandachli, Wael
    Ul Islam, Saif
    Richards, Robin
    Hall-Craggs, Margaret
    Witt, Johan
    [J]. HIP INTERNATIONAL, 2013, 23 (01) : 87 - 92
  • [10] 2D measurements of cup orientation are less reliable than 3D measurements A retrospective study of 87 metal-on-metal hips
    Davda, Kinner
    Smyth, Niall
    Cobb, Justin P.
    Hart, Alister J.