Impact of Impeller Speed Adjustment Interval on Hemolysis Performance of an Intravascular Micro-Axial Blood Pump

被引:0
|
作者
Liu, Yuan [1 ]
Zhu, Yuanfei [1 ]
Wang, Shangting [1 ]
Fu, Hualin [1 ]
Lu, Zhexin [2 ]
Yang, Ming [1 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Instrument Sci & Engn, Shanghai 200240, Peoples R China
[2] Shanghai Gen Hosp, Dept Cardiovasc Surg, Shanghai 201620, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
intravascular micro-axial blood pump; hemolysis estimation; finite element simulation; shear stress; flow rate adjustment; CONTINUOUS-FLOW; IN-VITRO; DAMAGE;
D O I
10.3390/mi15070934
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background: In recent years, intravascular micro-axial blood pumps have been increasingly used in the treatment of patients with cardiogenic shock. The flow rate of such blood pumps requires adjustment based on the patient's physiological condition. Compared to a stable flow state with fixed rotation speed, adjusting the speed of blood pump impeller to alter flow rate may lead to additional hemolysis. This study aimed at elucidating the relationship between adjusting interval of a blood pump's impeller speed and the hemolysis index. Methods: By comparing simulation results with P-Q characteristic curves of the blood pump measured by experiments, the accuracy of the blood pump flow field simulation model was confirmed. In this study, a drainage tube was employed as the device analogous to an intravascular micro-axial blood pump for achieving similar shear stress levels and residence times. The hemolysis finite element prediction method based on a power-law model was validated through hemolysis testing of porcine blood flow through the drainage tube. The validated models were subsequently utilized to investigate the impact of impeller speed adjusting intervals on hemolysis in the blood pump. Results: Compared to steady flow, the results demonstrate that the hemolysis index increased to 6.3% when changing the blood pump flow rate from 2 L/min to 2.5 L/min by adjusting the impeller speed within 0.072 s. Conclusions: An adjustment time of impeller speed longer than 0.072 s can avoid extra hemolysis when adjusting the intravascular micro-axial blood pump flow rate from 2 L/min to 2.5 L/min.
引用
收藏
页数:16
相关论文
共 5 条
  • [1] Particle Tracking Computational Prediction of Hemolysis By Blade of Micro-Axial Blood Pump
    Chen Xin
    Tan Jianping
    Yun Zhong
    MATERIALS SCIENCE AND ENGINEERING APPLICATIONS, PTS 1-3, 2011, 160-162 : 1779 - 1786
  • [2] Influence of Impeller Speed Patterns on Hemodynamic Characteristics and Hemolysis of the Blood Pump
    Wang, Yiwen
    Shen, Peng
    Zheng, Minli
    Fu, Pengqiang
    Liu, Lijia
    Wang, Jingyue
    Yuan, Lishan
    APPLIED SCIENCES-BASEL, 2019, 9 (21):
  • [3] HYDRODYNAMIC AND HEMOLYSIS ANALYSIS ON DISTANCE AND CLEARANCE BETWEEN IMPELLER AND DIFFUSER OF AXIAL BLOOD PUMP
    Yu, Jiajia
    Zhang, Xiwen
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2016, 16 (02)
  • [4] Effect of Conical Spiral Flow Channel and Impeller Parameters on Flow Field and Hemolysis Performance of an Axial Magnetic Blood Pump
    Yang, Wei-Bo
    Zhou, Jian
    Xiao, Wei-Hu
    Peng, Si-Jie
    Hu, Ye-Fa
    Li, Ming
    Wu, Hua-Chun
    PROCESSES, 2022, 10 (05)
  • [5] Numerical hemolysis performance evaluation of a rotary blood pump under different speed modulation profiles
    Huang, Feng
    Lei, Huan
    Ying, Shunv
    Fu, Yang
    Li, Qipeng
    Ruan, Xiaodong
    FRONTIERS IN PHYSIOLOGY, 2023, 14