A multianalyte LC-MS/MS method for accurate quantification of Nitrosamines in Olmesartan tablets

被引:2
|
作者
Pawar, Namdev [1 ]
Bhardwaj, Aparna [1 ]
Vora, Amisha [2 ]
Sharma, Sanjay [3 ]
机构
[1] Chauhan Inst Sci & Amrutben Jivanlal Coll Commerce, Mithibai Coll Arts, Dept Chem, Vile Parle West, Mumbai 400056, India
[2] NMIMS, Shobhaben Pratapbhai Patel Sch Pharm & Technol Man, Dept Chem, Vile Parle West, Mumbai 400056, India
[3] NMIMS, Shobhaben Pratapbhai Patel Sch Pharm & Technol Man, Dept Qual Assurance, Vile Parle West, Mumbai 400056, India
关键词
Olmesartan; Nitrosamines; LC-MS/MS; Analysis; QUANTITATIVE-DETERMINATION; PHARMACEUTICAL PRODUCTS; HUMAN PLASMA; RP-HPLC; HYDROCHLOROTHIAZIDE; MEDOXOMIL; AMLODIPINE; URINE; NDMA;
D O I
10.1016/j.chroma.2024.465176
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This research summaries the development, optimization and validation of liquid chromatography tandem mass spectrometric (LC-MS/MS) method for concurrent measurement of seven nitrosamines viz; NDMA, NDEA, NDIPA, NDPA, NEIPA, NMPA & NMBA in Olmesartan tablet. Controlling these nitrosamines at trace levels is imperative for ensuring the safety of drug substances and products for consumption. Various regulatory authorities stress the significance of utilizing highly sensitive analytical methods to precisely measure nitrosamines at trace levels. The method applied effective chromatographic separation and optimized parameters for mass spectrometric detection. Detection was carried out using APCI positive ion mode. Chromatographic separation was achieved using a Thermo Accucore PFP column (150 mm x 4.6 mm, 2.6 mu), with a simple gradient elution of mobile phase consisting of 0.1 % formic acid in water (mobile phase A) and methanol (mobile phase B). The total run time was 20 min, with a flow rate of 0.800 mL/min. The method was validated according to the International Council on Harmonisation (ICH Q2 (R2)) guidelines. The established method demonstrated excellent linearity (R-2> 0.99) and sensitivity for all the nitrosamines. Detection and quantification limits were sufficiently low for trace nitrosamine levels having good S/N ratio. The method showed good accuracy in Olmesartan tablet samples, with recoveries ranges between 80 % to 120 %. The new analytical approach has exceptional repeatability and reliability, making it possible to precisely quantify the levels of seven nitrosamines in Olmesartan medoxomil tablets in a single analytical run.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] Development and application of LC-MS/MS method for the quantification of hydrogen sulfide in the eye
    Okolie, Anthonia
    Nigro, Maria Rincon
    Polk, Sharhazad
    Stubbs, Keyona
    Chelliah, Selvam
    Ohia, Sunny E.
    Liang, Dong
    Mbye, Ya Fatou Njie
    ANALYTICAL BIOCHEMISTRY, 2024, 687
  • [32] Development of extraction method suitable for LC-MS/MS quantification of antibodies in tissues
    Fuchigami, Hirobumi
    Koganemaru, Shigehiro
    Yasunaga, Masahiro
    CANCER SCIENCE, 2025, 116 : 1560 - 1560
  • [33] LC-MS/MS method for the simultaneous quantification of intestinal CYP and UGT activity
    Busch, Diana
    Fritz, Anja
    Partecke, Lars Ivo
    Heidecke, Claus-Dieter
    Oswald, Stefan
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 155 : 194 - 201
  • [34] LC-MS/MS bioanalytical method for quantification of binimetinib and venetoclax, and their pharmacokinetic interaction
    Reddy, Anuradha
    Jadav, Tarang
    Sahu, Amit Kumar
    Sengupta, Pinaki
    BIOANALYSIS, 2021, 14 (02) : 75 - 86
  • [35] A rapid LC-MS/MS method for lutein quantification in spinach (Spinacia oleracea)
    Miura, Masashi
    Sakai, Miho
    Nogami, Mamiko
    Sato, Miwa
    Yatsushiro, Takeshi
    MICROCHEMICAL JOURNAL, 2020, 153
  • [36] A validated LC-MS/MS method for the quantification of climbazole metabolites in human urine
    Schmidtkunz, Christoph
    Kuepper, Katja
    Gries, Wolfgang
    Leng, Gabriele
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2021, 1173
  • [37] A simple LC-MS/MS method for the simultaneous quantification of endocannabinoids in biological samples
    Bobrich, Maria
    Schwarz, Rico
    Ramer, Robert
    Borchert, Philipp
    Hinz, Burkhard
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1161
  • [38] Response Factor based Quantification in LC-MS/MS
    Pauwels, S.
    Peersman, N.
    Gerits, M.
    Desmet, K.
    Vermeersch, P.
    ACTA CLINICA BELGICA, 2014, 69 : 6 - 7
  • [39] Standardized LC-MS/MS steroid hormone quantification
    Rauh, M.
    Koal, T.
    CLINICAL BIOCHEMISTRY, 2011, 44 (07) : 542 - 542
  • [40] Drugs of abuse quantification using LC-MS/MS
    Gauthier, Jeannette
    Petryayeva, Eleonora
    Nichols, Matthew
    Macri, Joseph
    Potter, Murray
    CLINICAL BIOCHEMISTRY, 2020, 86 : 80 - 80