Glycemic Outcomes and Nurse Perceptions of Continuous Glucose Monitoring for Hospitalized Patients

被引:0
|
作者
Thullen, Alexandra [1 ]
Gerber, Rebecca [1 ]
Keen, Alyson [1 ]
机构
[1] Indiana Univ Hlth, Adult Acad Hlth Ctr, Nursing Qual, 550 Univ Blvd, Indianapolis, IN 46202 USA
关键词
blood glucose; continuous glucose monitoring; costs and cost analysis; hypoglycemia; nurses; CRITICALLY-ILL PATIENTS;
D O I
10.1097/NCQ.0000000000000791
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Background: Continuous glucose monitoring (CGM) can decrease hypoglycemic events and health care costs; however, barriers and facilitators that influence CGM use are unknown. Purpose: The purpose of this study was to evaluate hypoglycemic events and cost outcomes after CGM implementation and describe associated barriers and facilitators. Methods: A mixed-methods study design was used to evaluate CGM implementation on 2 pulmonary units within an academic health center. Hypoglycemic events were evaluated before and after CGM implementation, and nurses were interviewed about facilitators and barriers that influence CGM use. Results: Hypoglycemic events decreased from a rate of 0.0906 per 1000 patient days to 0.0503 postimplementation, P < .0001. A $105 766 cost avoidance was recognized. Barriers and facilitators to CGM use are described. Conclusions: Findings support CGM implementation, while uniquely contributing financial impact and device use barriers and facilitators. Hospitals may consider CGM use to improve timely identification and treatment of hypoglycemia.
引用
收藏
页码:310 / 316
页数:7
相关论文
共 50 条
  • [31] Mean amplitude of glycemic excursions in septic patients and its association with outcomes: A prospective observational study using continuous glucose monitoring
    Furushima, Nana
    Egi, Moritoki
    Obata, Norihiko
    Sato, Hitoaki
    Mizobuchi, Satoshi
    JOURNAL OF CRITICAL CARE, 2021, 63 : 218 - 222
  • [32] Factors associated with optimal glycemic control in patients with insulin pump therapy and continuous glucose monitoring in real time
    Maria Gomez, Ana
    Maria Grizales, Ana
    Veloza, Angelica
    Marin, Alejandro
    Mauricio Munoz, Oscar
    Alonso Rondon, Martin
    AVANCES EN DIABETOLOGIA, 2013, 29 (03): : 74 - 80
  • [33] Novel Glycemic Index Based on Continuous Glucose Monitoring to Predict Poor Clinical Outcomes in Critically Ill Patients: A Pilot Study
    Ha, Eun Yeong
    Chung, Seung Min
    Park, Il Rae
    Lee, Yin Young
    Choi, Eun Young
    Moon, Jun Sung
    FRONTIERS IN ENDOCRINOLOGY, 2022, 13
  • [34] Continuous Glucose Monitoring vs. Capillary Blood Glucose in Hospitalized Type 2 Diabetes Patients
    Verissimo, David
    Vinhais, Joana
    Ivo, Catarina
    Martins, Ana Claudia
    Nunes e Silva, Joao
    Passos, Dolores
    Castro, Joao Jacome
    Lopes, Luis
    Marcelino, Mafalda
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (08)
  • [35] Continuous Glucose Monitoring in Patients Undergoing Extracorporeal Ventricular Assist Therapy
    Gottschalk, Antje
    Welp, Henryk A.
    Leser, Laura
    Lanckohr, Christian
    Wempe, Carola
    Ellger, Bjoern
    PLOS ONE, 2016, 11 (03):
  • [36] Determination of Postprandial Glycemic Responses by Continuous Glucose Monitoring in a Real-World Setting
    Roehling, Martin
    Martin, Tobias
    Wonnemann, Meinolf
    Kragl, Martin
    Klein, Horst Harald
    Heinemann, Lutz
    Martin, Stephan
    Kempf, Kerstin
    NUTRIENTS, 2019, 11 (10)
  • [37] Continuous Glucose Monitoring for the Detection of Hypoglycemia in Patients With Diabetes of the Exocrine Pancreas
    Shivaprasad, Channabasappa
    Gautham, Kolla
    Shah, Kejal
    Gupta, Soumya
    Palani, Preethika
    Anupam, Biswas
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2021, 15 (06): : 1313 - 1319
  • [38] Evaluating Rate of Change as an Index of Glycemic Variability, Using Continuous Glucose Monitoring Data
    Whitelaw, Benjamin C.
    Choudhary, Pratik
    Hopkins, David
    DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (06) : 631 - 636
  • [39] Glycemic Profile of Intravenous Dexamethasone-Induced Hyperglycemia Using Continuous Glucose Monitoring
    Zhang, Fan
    Karam, Jocelyne G.
    AMERICAN JOURNAL OF CASE REPORTS, 2021, 22
  • [40] Value of continuous glucose monitoring for minimizing severe hypoglycemia during tight glycemic control
    Steil, Garry M.
    Langer, Monica
    Jaeger, Karen
    Alexander, Jamin
    Gaies, Michael
    Agus, Michael S. D.
    PEDIATRIC CRITICAL CARE MEDICINE, 2011, 12 (06) : 643 - 648