Quantification accuracy in photon-counting detector CT for coronary artery calcium score: a pilot study

被引:2
作者
Wang, Mengzhen [1 ]
Zhang, Xu [1 ]
Li, Jiqang [1 ]
Dong, Haipeng [1 ]
Xu, Zhihan [2 ]
Yan, Fuhua [1 ]
Yang, Wenjie [1 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Dept Radiol, 197 Ruijin Er Rd, Shanghai 200025, Peoples R China
[2] Siemens Healthineers CT Collaborat, Shanghai, Peoples R China
关键词
Photon-counting detector CT; Coronary artery calcification score; Agatston score; Radiation dose; COMPUTED-TOMOGRAPHY; SCANS; RISK;
D O I
10.1007/s10554-024-03209-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To validate the accuracy of coronary artery calcium score (CACS) using photon-counting detector (PCD) CT under various scanning settings and explore the optimized scanning settings considering both the accuracy and the radiation dose. A CACS phantom containing six hollow cylindrical hydroxyapatite calcifications of two sizes with three densities and 12 patients underwent CACS scans. For PCD-CT, two scanning modes (sequence and flash [high-pitch spiral mode]) and five tube voltages (90kV, 120kV, 140kV, Sn100kV, and Sn140kV) at different image quality (IQ) levels were set for phantom, and patients were scanned with 120kV at IQ19 using flash mode. All acquisitions from PCD-CT were reconstructed at 70keV. Acquisitions in sequence mode at 120kV on an energy-integrating detector CT (EID-CT) was used as the reference. Agatston, mass, and volume scores were calculated. The CACS from PCD-CT exhibited excellent agreements with the reference (all intraclass correlation coefficient [ICC] > 0.99). The root mean square error (RMSE) between the Agatston score acquired from PCD-CT and the reference (5.4-11.5) was small. A radiation dose reduction (16-75%) from PCD-CT compared with the reference was obtained in all protocols using flash mode, albeit with IQ20 only at sequence mode (22-44%). For the patients, ICC ( all ICC > 0.98) and Bland-Altman analysis of CACS all showed high agreements between PCD-CT and the reference, without reclassifying CACS categories(P = 0.317). PCD-CT yields repeatable and accurate CACS across diverse scanning protocols according to our pilot study. Sn100kV, 90kV, and 120kV using flash mode at IQ20 are recommended for clinical applications considering both accuracy and radiation dose.
引用
收藏
页码:2181 / 2191
页数:11
相关论文
共 31 条
[1]   Low-dose coronary artery calcium scoring compared to the standard protocol [J].
Allio, Ileana Rosely ;
Caobelli, Federico ;
Popescu, Cristina Elena ;
Haaf, Philip ;
Alberts, Ian ;
Frey, Simon M. ;
Zellweger, Michael J. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2023, 30 (03) :1191-1198
[2]   Scoring of coronary artery calcium scans: History, assumptions, current limitations, and future directions [J].
Alluri, Krishna ;
Joshi, Parag H. ;
Henry, Travis S. ;
Blumenthal, Roger S. ;
Nasir, Khurram ;
Blaha, Michael J. .
ATHEROSCLEROSIS, 2015, 239 (01) :109-117
[3]  
Boone JM, 2011, REPORT NO 204 SIZE S, DOI [10.37206/143, DOI 10.37206/143]
[4]   Coronary artery calcium score, risk factors, and incident coronary heart disease events [J].
Church, Timothy S. ;
Levine, Benjamin D. ;
McGuire, Darren K. ;
LaMonte, Michael J. ;
FitzGerald, Shannon J. ;
Cheng, Yiling J. ;
Kimball, Thomas E. ;
Blair, Steven N. ;
Gibbons, Larry W. ;
Nichaman, Milton Z. .
ATHEROSCLEROSIS, 2007, 190 (01) :224-231
[5]   Radiation dose optimization for photon-counting CT coronary artery calcium scoring for different patient sizes: a dynamic phantom study [J].
Dobrolinska, Magdalena M. M. ;
van der Werf, Niels R. R. ;
van der Bie, Judith ;
de Groen, Joel ;
Dijkshoorn, Marcel ;
Booij, Ronald ;
Budde, Ricardo P. J. ;
Greuter, Marcel J. W. ;
van Straten, Marcel .
EUROPEAN RADIOLOGY, 2023, 33 (07) :4668-4675
[6]   Coronary Calcium Scoring with First Generation Dual-Source Photon-Counting CT-First Evidence from Phantom and In-Vivo Scans [J].
Eberhard, Matthias ;
Mergen, Victor ;
Higashigaito, Kai ;
Allmendinger, Thomas ;
Manka, Robert ;
Flohr, Thomas ;
Schmidt, Bernhard ;
Euler, Andre ;
Alkadhi, Hatem .
DIAGNOSTICS, 2021, 11 (09)
[7]   High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose [J].
Euler, Andre ;
Higashigaito, Kai ;
Mergen, Victor ;
Sartoretti, Thomas ;
Zanini, Bettina ;
Schmidt, Bernhard ;
Flohr, Thomas G. ;
Ulzheimer, Stefan ;
Eberhard, Matthias ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2022, 57 (02) :115-121
[8]   Lifetime Cardiovascular Disease Risk by Coronary Artery Calcium Score in Individuals With and Without Diabetes: An Analysis From the Multi-Ethnic Study of Atherosclerosis [J].
Ferket, Bart S. ;
Hunink, M. G. Myriam ;
Masharani, Umesh ;
Max, Wendy ;
Yeboah, Joseph ;
Burke, Gregory L. ;
Fleischmann, Kirsten E. .
DIABETES CARE, 2022, 45 (04) :975-982
[9]   Photon-counting CT review [J].
Flohr, Thomas ;
Petersilka, Martin ;
Henning, Andre ;
Ulzheimer, Stefan ;
Ferda, Jiri ;
Schmidt, Bernhard .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 :126-136
[10]   Coronary Artery Calcium Scoring on Different 64-detector Scanners Using a Low-tube Voltage (80 kVp) [J].
Fujioka, Chikako ;
Funama, Yoshinori ;
Kiguchi, Masao ;
Ishifuro, Minoru ;
Kihara, Yasuki ;
Nagata, Yasushi ;
Awai, Kazuo .
ACADEMIC RADIOLOGY, 2012, 19 (11) :1402-1407