A Prospective, Longitudinal Study Evaluating the Efficacy of an Automated Secretion Removal Technology

被引:2
作者
Sinnige, Jante S. [1 ]
Karbing, Dan S. [2 ]
Valk, Christel M. A. [1 ]
Schultz, Marcus J. [1 ,3 ,4 ]
Rees, Stephen E. [2 ]
Paulus, Frederique [1 ,5 ]
机构
[1] Univ Amsterdam, Dept Intens Care, Amsterdam UMC, Amsterdam, Netherlands
[2] Aalborg Univ, Dept Hlth Sci & Technol, Resp & Crit Care Grp, Aalborg, Denmark
[3] Mahidol Univ, Mahidol Oxford Trop Med Res Unit, Bangkok, Thailand
[4] Univ Oxford, Nuffield Dept Med Oxford, Oxford, England
[5] Amsterdam Univ Appl Sci, Fac Hlth, Ctr Expertise, Urban Vital, Amsterdam, Netherlands
关键词
Tracheal secretions; artificial cough; endotracheal suctioning; mechanical ventilation; INTENSIVE-CARE-UNIT; RECOLLECTION; VENTILATION;
D O I
10.4187/respcare.11584
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Endotracheal suctioning causes discomfort, is associated with adverse effects, and is resource-demanding. An artificial secretion removal method, known as an automated cough, has been developed, which applies rapid, automated deflation, and inflation of the endotracheal tube cuff during the inspiratory phase of mechanical ventilation. This method has been evaluated in the hands of researchers but not when used by attending nurses. The aim of this study was to explore the efficacy of the method over the course of patient management as part of routine care. METHODS: This prospective, longitudinal, interventional study recruited 28 subjects who were intubated and mechanically ventilated. For a maximum of 7 d and on clinical need for endotracheal suctioning, the automatic cough procedure was applied. The subjects were placed in a pressure-regulated ventilation mode with elevated inspiratory pressure, and automated cuff deflation and inflation were performed 3 times, with this repeated if deemed necessary. Success was determined by resolution of the clinical need for suctioning as determined by the attending nurse. Adverse effects were recorded. RESULTS: A total of 84 procedures were performed. In 54% of the subjects, the artificial cough procedure was successful on > 70% of occasions, with 56% of all procedures considered successful. Ninety percent of all the procedures were performed in subjects who were spontaneously breathing and on pressure-support ventilation with peak inspiratory pressures of 20 cm H2O. Rates of adverse events were similar to those seen in the application of endotracheal suctioning. CONCLUSIONS: This study solely evaluated the efficacy of an automated artificial cough procedure, which illustrated the potential for reducing the need for endotracheal suctioning when applied by attending nurses in routine care.
引用
收藏
页码:931 / 936
页数:6
相关论文
共 14 条
[1]   AARC Clinical Practice Guidelines: Artificial Airway Suctioning [J].
Blakeman, Thomas C. ;
Scott, J. Brady ;
Yoder, Mark A. ;
Capellari, Emily ;
Strickland, Shawna L. .
RESPIRATORY CARE, 2022, 67 (02) :258-271
[2]   Effects of Mechanical Insufflation-Exsufflation on Airway Mucus Clearance Among Mechanically Ventilated ICU Subjects [J].
Ferreira de Camillis, Marcio Luiz ;
Savi, Augusto ;
Rosa, Regis Goulart ;
Figueiredo, Mariana ;
Wickert, Ricardo ;
Alegretti Borges, Luis Guilherme ;
Galant, Lucas ;
Teixeira, Cassiano .
RESPIRATORY CARE, 2018, 63 (12) :1471-1477
[3]   Effects of mechanical insufflation-exsufflation in preventing respiratory failure after extubation: a randomized controlled trial [J].
Goncalves, Miguel R. ;
Honrado, Teresa ;
Winck, Joao Carlos ;
Paiva, Jose Artur .
CRITICAL CARE, 2012, 16 (02)
[4]   Patients' recollections of experiences in the intensive care unit may affect their quality of life [J].
Granja, C ;
Lopes, A ;
Moreira, S ;
Dias, C ;
Costa-Pereira, A ;
Carneiro, A .
CRITICAL CARE, 2005, 9 (02) :R96-R109
[5]   Effect of open and closed endotracheal suctioning on cross-transmission with Gram-negative bacteria: A prospective crossover study [J].
Jongerden, Irene P. ;
Buiting, Anton G. ;
Leverstein-van Hall, Maurine A. ;
Speelberg, Ben ;
Zeidler, Shimriet ;
Kesecioglu, Jozef ;
Bonten, Marc J. .
CRITICAL CARE MEDICINE, 2011, 39 (06) :1313-1321
[6]   Decreasing the Adverse Effects of Endotracheal Suctioning During Mechanical Ventilation by Changing Practice [J].
Maggiore, Salvatore Maurizio ;
Lellouche, Francois ;
Pignataro, Claudia ;
Girou, Emmanuelle ;
Maitre, Bernard ;
Richard, Jean-Christophe M. ;
Lemaire, Francois ;
Brun-Buisson, Christian ;
Brochard, Laurent .
RESPIRATORY CARE, 2013, 58 (10) :1588-1597
[7]   ???????Effects of Mechanical Insufflation-Exsufflation With Different Pressure Settings on Respiratory Mucus Displacement During Invasive Ventilation br [J].
Marti, Joan-Daniel ;
Martinez-Alejos, Roberto ;
Pilar-Diaz, Xabier ;
Yang, Hua ;
Pagliara, Francesco ;
Battaglini, Denise ;
Meli, Andrea ;
Yang, Milan ;
Bobi, Joaquim ;
Rigol, Monsterrat ;
Tronstad, Oystein ;
Volpe, Marcia Souza ;
Passos Amato, Marcelo Britto ;
Li Bassi, Gianluigi ;
Torres, Antoni .
RESPIRATORY CARE, 2022, 67 (12) :1508-1516
[8]   Efficacy of an Automated Secretion Removal Technology at Different Inspiratory Pressures [J].
Nielsen, Anne H. ;
Karbing, Dan S. ;
Solling, Christoffer G. ;
Winding, Robert R. ;
Rees, Stephen E. ;
Dey, Nilanjan .
RESPIRATORY CARE, 2023, 68 (11) :1502-1509
[9]   Empirical Evidence of Study Design Biases in Randomized Trials: Systematic Review of Meta-Epidemiological Studies [J].
Page, Matthew J. ;
Higgins, Julian P. T. ;
Clayton, Gemma ;
Sterne, Jonathan A. C. ;
Hrobjartsson, Asbjorn ;
Savovic, Jelena .
PLOS ONE, 2016, 11 (07)
[10]   Preliminary experience on the safety and tolerability of mechanical “insufflation-exsufflation” in subjects with artificial airway [J].
Sánchez-García M. ;
Santos P. ;
Rodríguez-Trigo G. ;
Martínez-Sagasti F. ;
Fariña-González T. ;
del Pino-Ramírez Á. ;
Cardenal-Sánchez C. ;
Busto-González B. ;
Requesens-Solera M. ;
Nieto-Cabrera M. ;
Romero-Romero F. ;
Núñez-Reiz A. .
Intensive Care Medicine Experimental, 6 (1)