Dose profile measurement using an imaging plate: Evaluation of filters using Monte Carlo simulation of 4 MV x-rays

被引:0
|
作者
Hashimoto, Masatoshi [1 ,2 ]
Tomita, Tetsuya [3 ]
Sawada, Koichi [3 ]
Fujibuchi, Toshioh [4 ]
Nishio, Teiji [5 ]
Nakagawa, Keiichi [1 ]
机构
[1] Univ Tokyo, Grad Sch Med, Div Radiol & Biomed Engn, Bunkyo Ku, Tokyo 1138655, Japan
[2] Med Plaza Edogawa, Dept Therapeut Radiol, Edogawa Ku, Tokyo 1330052, Japan
[3] Chiba Univ Hosp, Dept Radiol, Cyuo Ku, Chiba 2608677, Japan
[4] Ibaraki Prefectural Univ, Sch Hlth Sci, Dept Radiol Sci, Ibaraki 3000394, Japan
[5] Natl Canc Ctr Hosp E, Res Ctr Innovat Oncol, Particle Therapy Div, Kashiwa, Chiba 2778577, Japan
关键词
diagnostic radiography; dosimetry; filters; lead; Monte Carlo methods; X-ray apparatus; HIGH-ENERGY PHOTON; FILM DOSIMETRY; COMPUTED RADIOGRAPHY; RADIATION-THERAPY; QUALITY-ASSURANCE; EBT FILM; BEAMS; RECONSTRUCTION; SPECTRA;
D O I
10.1063/1.3103572
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Computed radiography (CR) is gradually replacing film. The application of CR for two-dimensional profiles and off-axis ratio (OAR) measurement using an imaging plate (IP) in a CR system is currently under discussion. However, a well known problem for IPs in dosimetry is that they use high atomic number (Z) materials, such as Ba, which have an energy dependency in a photon interaction. Although there are some reports that it is possible to compensate for the energy dependency with metal filters, the appropriate thicknesses of these filters and where they should be located have not been investigated. The purpose of this study is to find the most suitable filter for use with an IP as a dosimetric tool. Monte Carlo simulation (Geant4 8.1) was used to determine the filter to minimize the measurement error in OAR measurements of 4 MV x-rays. In this simulation, the material and thickness of the filter and distance between the IP and the filter were varied to determine most suitable filter conditions that gave the best fit to the MC calculated OAR in water. With regard to changing the filter material, we found that using higher Z and higher density material increased the effectiveness of the filter. Also, increasing the distance between the filter and the IP reduced the effectiveness, whereas increasing the thickness of the filter increased the effectiveness. The result of this study showed that the most appropriate filter conditions consistent with the calculated OAR in water were the ones with the IP sandwiched between two 2 mm thick lead filters at a distance of 5 mm from the IP or the IP sandwiched directly between two 1 mm lead filters. Using these filters, we measured the OAR at 10 cm depth with 100 cm source-to-surface distance and surface 10x10 cm(2) field size. The results of this measurement represented that it is possible to achieve measurements with less than within 2.0% and 2.0% in the field and with less than 1.1% and 0.6% out of the field by using 2 and 1 mm lead filters, respectively.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Commissioning of Elekta InfinityTM 6 MV flattening filter-free using Monte Carlo simulation
    Maulana, Iswan Fadhil
    Yani, Sitti
    Sumaryada, Tony
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    RADIATION PHYSICS AND CHEMISTRY, 2023, 210
  • [32] ORGAN DOSE EVALUATIONS BASED ON MONTE CARLO SIMULATION FOR CT EXAMINATIONS USING TUBE CURRENT MODULATION
    Fujii, K.
    Nomura, K.
    Muramatsu, Y.
    Obara, S.
    Akahane, K.
    Kusumoto, M.
    RADIATION PROTECTION DOSIMETRY, 2017, 174 (03) : 387 - 394
  • [33] Evaluation of a Radiation-Induced Cell Proliferation Probability Formula Using Monte Carlo Simulation
    Watanabe, Y.
    Dahlman, E.
    MEDICAL PHYSICS, 2014, 41 (06) : 170 - 170
  • [34] Monte Carlo simulation of Novalis Classic 6 MV accelerator using phase space generation in GATE/Geant4 code
    Teixeira, M. S.
    Batista, D. V. S.
    Braz, D.
    da Rosa, L. A. R.
    PROGRESS IN NUCLEAR ENERGY, 2019, 110 : 142 - 147
  • [35] Independent checking of the delivered dose for high-energy X-rays using a hand-held PC
    Knöös, T
    Johnsson, SA
    Ceberg, CP
    Tomaszewicz, A
    Nilsson, P
    RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) : 201 - 208
  • [36] Radiation dose in breast CT imaging with monochromatic x-rays: simulation study of the influence of energy, composition and thickness
    Mittone, A.
    Bravin, A.
    Coan, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (09) : 2199 - 2217
  • [37] DOSE POINT KERNEL SIMULATION FOR MONOENERGETIC ELECTRONS AND RADIONUCLIDES USING MONTE CARLO TECHNIQUES
    Wu, J.
    Liu, Y. L.
    Chang, S. J.
    Chao, M. M.
    Tsai, S. Y.
    Huang, D. E.
    RADIATION PROTECTION DOSIMETRY, 2012, 152 (1-3) : 119 - 124
  • [38] Monte Carlo dosimetry of a cell culture irradiation model using a 6 MV X-ray beam
    Hernandez-Bojorquez, Mariana
    Trejo-Solis, Cristina
    Manuel Larraga-Gutierrez, Jose
    Martinez-Davalos, Arnulfo
    RADIATION PHYSICS AND CHEMISTRY, 2021, 180
  • [39] Predicting energy response of radiographic film in a 6 MV x-ray beam using Monte Carlo calculated fluence spectra and absorbed dose
    Palm, Å
    Kirov, AS
    LoSasso, T
    MEDICAL PHYSICS, 2004, 31 (12) : 3168 - 3178
  • [40] Intercomparison of Monte Carlo calculated dose enhancement ratios for gold nanoparticles irradiated by X-rays: Assessing the uncertainty and correct methodology for extended beams
    Rabus, H.
    Li, W. B.
    Villagrasa, C.
    Schuemann, J.
    Hepperle, P. A.
    Rosales, L. de la Fuente
    Beuve, M.
    Di Maria, S.
    Klapproth, A. P.
    Li, C. Y.
    Poignant, F.
    Rudek, B.
    Nettelbeck, H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 241 - 253