Agar-based Phantom for Evaluating Targeting of High-intensity Focused Ultrasound Systems for Breast Ablation

被引:2
作者
Filippou, Antria [1 ]
Damianou, Christakis [1 ]
机构
[1] Cyprus Univ Technol, Dept Elect Engn Comp Engn & Informat, 30 Archbishop Kyprianou St, CY-3036 Limassol, Cyprus
关键词
breast; cancer; high-intensity focused ultrasound; phantom; RADIOFREQUENCY ABLATION; CONSERVING SURGERY; SOLID CARCINOMAS; CANCER SURGERY; THERAPY; MASTECTOMY; CELLS; TRIAL; RISK;
D O I
10.4103/jmp.jmp_52_24
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: Phantoms are often utilized for the preclinical evaluation of novel high-intensity focused ultrasound (HIFU) systems, serving as valuable tools for validating efficacy. In the present study, the feasibility of a homogeneous agar-based breast-shaped phantom as a tool for the preclinical evaluation of HIFU systems dedicated to breast cancer was assessed. Specifically, the effect of the increased phantom curvature on temperature increase was examined through sonications executed on two sides having varied curvatures.Materials and Methods: Assessment was performed utilizing a 1.1 MHz focused transducer. Sonications on the two phantom sides were executed at varied acoustical power in both a laboratory setting and inside a 1.5 T magnetic resonance imaging scanner. Sonications were independently performed on two identical phantoms for repeatability purposes.Results: Temperature changes between 7.1 degrees C-34.3 degrees C and 5.1 degrees C-21.5 degrees C were recorded within the decreased and increased curvature sides, respectively, for acoustical power of 3.75-10 W. High-power sonications created lesions which were approximately symmetrically formed around the focal point at the decreased curvature side, while they were shifted away from the focal point at the increased curvature side.Conclusions: The present findings indicate that increased curvature of the breast phantom results in deformed focal shapes and decreased temperatures induced at the focal area, thus suggesting treatment correction requirements in the form of focus control or accurate robotic movement. The developed breast-shaped phantom can be utilized as an evaluation tool of HIFU systems dedicated to breast cancer since it can visually verify the efficacy of any system.
引用
收藏
页码:343 / 355
页数:13
相关论文
共 78 条
[1]  
ABE O, 1995, NEW ENGL J MED, V333, P1444
[2]  
[Anonymous], CIRS Inc
[3]   Reproducing two-dimensional mammograms with three-dimensional printed phantoms [J].
Badal, Andreu ;
Clark, Matthew ;
Ghammraoui, Bahaa .
JOURNAL OF MEDICAL IMAGING, 2018, 5 (03)
[4]   An empirical correlation to estimate thermal properties of the breast and of the breast nodule using thermographic images and optimization techniques [J].
Bezerra, L. A. ;
Ribeiro, R. R. ;
Lyra, P. R. M. ;
Lima, R. C. F. .
INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2020, 149
[5]   SIGNAL-TO-NOISE AND CONTRAST IN FAST SPIN-ECHO (FSE) AND INVERSION RECOVERY FSE IMAGING [J].
CONSTABLE, RT ;
SMITH, RC ;
GORE, JC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (01) :41-47
[6]   Polyvinyl chloride plastisol breast phantoms for ultrasound imaging [J].
de Carvalho, Isabela Miller ;
De Matheo, Lucas Lobianco ;
Silva Costa Junior, Jose Francisco ;
Borba, Cecilia de Melo ;
von Kruger, Marco Antonio ;
Catelli Infantosi, Antonio Fernando ;
de Albuquerque Pereira, Wagner Coelho .
ULTRASONICS, 2016, 70 :98-106
[7]   PVCP-based anthropomorphic breast phantoms containing structures similar to lactiferous ducts for ultrasound imaging: A comparison with human breasts [J].
De Matheo, Lucas Lobianco ;
Geremia, Juliana ;
Gregorio Calas, Maria Julia ;
Silva Costa-Junior, Jose Francisco ;
Ferreira da Silva, Flavia Fernandes ;
von Kruger, Marco Antonio ;
de Albuquerque Pereira, Wagner Coelho .
ULTRASONICS, 2018, 90 :144-152
[8]   Experimental assessment of phase aberration correction for breast MRgFUS therapy [J].
Dillon, Christopher R. ;
Farrer, Alexis ;
McLean, Hailey ;
Almquist, Scott ;
Christensen, Douglas ;
Payne, Allison .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2018, 34 (06) :731-743
[9]   Characterization of a soft tissue-mimicking agar/wood powder material for MRgFUS applications [J].
Drakos, Theocharis ;
Giannakou, Marinos ;
Menikou, Georgios ;
Constantinides, Georgios ;
Damianou, Christakis .
ULTRASONICS, 2021, 113
[10]  
Duck FA., 1990, PHYS PROPERTIES TISS