Validation of oscillometric ratio and maximum gradient methods for non-invasive blood pressure measurement with intra-arterial blood pressure measurements as reference

被引:0
|
作者
Celler, Branko G. [1 ]
Yong, Andy [2 ,3 ]
Rubenis, Imants [2 ]
Butlin, Mark [3 ]
Argha, Ahmadreza [4 ]
Rehan, Rajan [2 ]
Avolio, Alberto [3 ]
机构
[1] Univ New South Wales, Biomed Syst Res Lab, Kensington, Australia
[2] Univ Sydney, Concord Repatriat Hosp, Cardiol, Camperdown, Australia
[3] Macquarie Univ, Fac Med Hlth & Human Sci, Macquarie Med Sch, Sydney, NSW, Australia
[4] Univ New South Wales, Grad Sch Biomed Engn, Sydney, NSW, Australia
关键词
auscultatory methods; blood pressure; Korotkoff sounds; linear models; non-invasive blood pressure measurement; sphygmomanometry; SBP and DBP estimation; ALGORITHM; HYPERTENSION; PULSE;
D O I
10.1097/HJH.0000000000003698
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Most non-invasive blood pressure (BP) measurements are carried out using instruments which implement either the Ratio or the Maximum Gradient oscillometric method, mostly during cuff deflation, but more rarely during cuff inflation. Yet, there is little published literature on the relative advantages and accuracy of these two methods. In this study of 40 lightly sedated individuals aged 64.1 +/- 9.6 years, we evaluate and compare the performance of the oscillometric ratio (K) and gradient (Grad) methods for the non-invasive estimation of mean pressure, SBP and DBP with reference to invasive intra-arterial values. There was no significant difference between intra-arterial estimates of mean pressure made via Korotkoff sounds (MP-OWE) or the gradient method (MP-Grad). However, 17.7% of MP-OWE and 15% of MP-Grad were in error by more than 10 mmHg. SBP-K and SBP-Grad underestimated SBP by 14 and 18 mmHg, whilst accurately estimating DBP with mean errors of 0.4 +/- 5.0 and 1.7 +/- 6.1 mmHg, respectively. Relative to the reference standard SBP-K, SBP-Grad and DBP-Grad were estimated with a mean error of -4.5 +/- 6.6 and 1.4 +/- 5.6 mmHg, respectively, noting that using the full range of recommended ratios introduces errors of 12 and 7 mmHg in SBP and DBP, respectively. We also show that it is possible to find ratios which minimize the root mean square error (RMSE) and the mean error for any particular individual cohort. We developed linear models for estimating SBP and SBP-K from a range of demographic and non-invasive OWE variables with resulting mean errors of 0.15 +/- 5.6 and 0.3 +/- 5.7 mmHg, acceptable according to the Universal standard.
引用
收藏
页码:1075 / 1085
页数:11
相关论文
共 50 条
  • [31] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, JN
    Geake, WB
    Scott, DHT
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (02) : 131 - 134
  • [32] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, J.N.
    Geake, W.B.
    Scott, D.H.T.
    Medical and Biological Engineering and Computing, 1997, 35 (02): : 131 - 134
  • [33] Oscillometric non-invasive blood pressure measurements: the influence of the make of instrument on readings?
    J. N. Amoore
    W. B. Geake
    D. H. T. Scott
    Medical and Biological Engineering and Computing, 1997, 35 : 131 - 134
  • [34] A comparison of invasive and non-invasive arterial blood pressure measurements in the neonate
    Shea, J. E. O'
    Dempsey, E. M.
    ACTA PAEDIATRICA, 2007, 96 : 190 - 191
  • [35] ANALYSIS OF AGREEMENT BETWEEN NON-INVASIVE AND INTRA-ARTERIAL BLOOD PRESSURE IN PEDIATRIC BURNED PATIENTS
    Rodriguez, N. A.
    Branski, L. K.
    Finnerty, C.
    Jennings, K.
    Mlcak, R.
    Herndon, D. N.
    SHOCK, 2011, 35 : 1 - 1
  • [36] A COMPARISON OF INVASIVE AND NON-INVASIVE ARTERIAL BLOOD PRESSURE MEASUREMENTS IN THE NEONATE
    O'Shea, J.
    Dempsey, E.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2010, 179 : S174 - S174
  • [37] Blood Pressure Measurement Compared Between Oscillometric Blood Pressure Monitoring and Gold Standard Intra-Arterial Blood Pressure Monitoring in Adult Shock Resuscitation Patients
    Tongyoo, Surat
    Visuthisakchai, Suharit
    Permpikul, Chairat
    JOURNAL OF CRITICAL & INTENSIVE CARE, 2022, 13 (03): : 102 - 109
  • [38] ERRORS IN INTRA-ARTERIAL BLOOD-PRESSURE MEASUREMENTS
    ZINGG, W
    BAINES, WD
    MATHUR, AN
    CANADIAN JOURNAL OF SURGERY, 1968, 11 (01) : 93 - &
  • [39] A new indicator for non-invasive arterial blood pressure measurement
    Regueiro-Gomez, A
    Pallas-Areny, R
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 117 - 118
  • [40] A comparative study of invasive and oscillometric methods of arterial blood pressure measurement in the anesthetized rabbit
    Ypsilantis, P
    Didilis, VN
    Politou, M
    Bougioukas, I
    Bougioukas, G
    Simopoulos, C
    RESEARCH IN VETERINARY SCIENCE, 2005, 78 (03) : 269 - 275