High Performance Liquid chromatography - Fourier Transform Infrared Spectroscopy Coupling: A Comprehensive Review

被引:0
作者
Halko, Radoslav [1 ]
Pavelek, Denis [1 ]
Kaykhaii, Massoud [1 ,2 ]
机构
[1] Comenius Univ, Fac Nat Sci, Dept Analyt Chem, Ilkovicova 6, Bratislava 84215, Slovakia
[2] Univ Tasmania, Coll Sci & Engn, Sch Nat Sci Chem, Hobart, Australia
关键词
Fourier transform infrared spectrometry; high performance liquid chromatography; coupling; interface; flow cell; solvent elimination; SAMPLE-INTRODUCTION; SPECTROMETRY; FTIR; FRACTIONATION; COPOLYMERS; IDENTIFICATION; SEPARATION; OXIDATION; WATER; SIZE;
D O I
10.1080/10408347.2024.2391892
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This review presents a critical examination of the interface for coupling high performance liquid chromatography (HPLC) with Fourier transform infrared spectrometry (FTIR) since 2010. This coupling offers a robust analytical approach characterized by exceptional chemical specificity and the capacity to analyze complex multi-component mixtures qualitatively and quantitatively with high sensitivity, particularly in low limit of detection ranges. This coupling enables the identification of individual components of a mixture by IR after their separation by HPLC, although challenges arise from the potential distortion of infrared spectra by mobile phase components. Addressing this issue necessitates the implementation of suitable interfaces, such as flow cells or off-line indirect measurement methods like hot inert gas streams or ultrasonic nebulizers. The key parameters influencing the coupling of HPLC-FTIR include the solvent elimination methods, mode of FTIR technique, and IR background for accurate analyte identification. Moreover, the composition of the mobile phase and the utilization of buffer solutions in the HPLC mobile phase profoundly impact analyte identification by FTIR.
引用
收藏
页数:12
相关论文
共 65 条
[61]  
Taylor S. D., 2007, In-Line Combination of LC with MS, NMR, UV and IR in Drug Analysis
[62]   DIRECT AND INDIRECT DETECTION OF LIQUID-CHROMATOGRAPHY BY INFRARED THERMAL LENS SPECTROMETRY [J].
TRAN, CD ;
HUANG, GL ;
GRISHKO, VI .
ANALYTICA CHIMICA ACTA, 1995, 299 (03) :361-369
[63]   Applications of Hyphenated Liquid Chromatography Techniques for Polymer Analysis [J].
Uliyanchenko, Elena .
CHROMATOGRAPHIA, 2017, 80 (05) :731-750
[64]   Challenges in polymer analysis by liquid chromatography [J].
Uliyanchenko, Elena ;
van der Wal, Sjoerd ;
Schoenmakers, Peter J. .
POLYMER CHEMISTRY, 2012, 3 (09) :2313-2335
[65]   Separation and determination of estrogen in the water environment by high performance liquid chromatography-fourier transform infrared spectroscopy [J].
Zheng, Bei ;
Li, Wentao ;
Li, Hongyan ;
Liu, Lin ;
Lei, Pei ;
Ge, Xiaopeng ;
Yu, Zhiyong ;
Zhou, Yiqi .
SCIENTIFIC REPORTS, 2016, 6