Method Development and Validation for the Quantification of Abametapir in Biological Matrices by LC-ESI-MS/MS

被引:0
|
作者
Krishnan, Vinayaga Sundaram
Bhikshapathi, Darna [1 ,2 ,3 ]
Cheruku, Shankar [2 ]
机构
[1] Bir Tikandrajit Univ, Dept Chem, Imphal, Manipur, India
[2] Bir Tikandrajit Univ, Dept Pharm, Imphal, Manipur, India
[3] TRR Coll Pharm, Dept Pharmaceut Anal, Hyderabad, Telangana, India
关键词
LC-MS/MS; Abametapir; Metalloproteinase Inhibitor; Validation; Accuracy; Linearity; HUMAN PLASMA; LICE;
D O I
10.5530/ijper.58.3s.102
中图分类号
G40 [教育学];
学科分类号
040101 ; 120403 ;
摘要
Aim: A precise, linear and specific liquid chromatography-tandem mass spectrometry procedure was executed and subjected for validation for the quantification of plasma. Materials and Methods: The chromatography elution was carried out using a C18-Discovery column of 15 cm in length and 2.1 mm in internal diameter, packed with stationary phase particles of 5 mu m size, at a flow rate of 0.80 mL/min. An isocratic elution process was conducted with a mobile phase solution consisting of methanol and 0.10% V/V HCOOH in a ratio of 90:10. The separation of Abametapir and Metformin was achieved by a liquid-liquid extraction process using ethyl acetate as the solvent. Results: A triple quadrupole mass spectrometer was used for the measurement of ions. Electrospray ionization is a technology that ionizes positively, and it was used in Multiple Reaction Monitoring (MRM) with parent/product ionic transitions m/z 185.1-106.06 for Abametapir and 130.1-60.20 for the Metformin internal standard. A calibration graph was constructed with values ranging from 2.05 to 82.00 ng/mL, resulting in the equation y=0.0149x+0.00221, with a r2 value above 0.99. The recovery values of Abametapir exceeded 94.27%, with its accuracy assessed in terms of relative error ranging from -4.04% to 6.52%. Conclusion: The accuracy, recovery and sensitivity values of Abametapir in the plasma sample, as shown by the established approach, highlight its significance in pharmacokinetic and bioequivalence research.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] LC-ESI-MS/MS Method for the Fosamprenavir Quantification Bioanalytical Method Development and Validation for the Quantification Fosamprenavir in Human Plasma by LC-ESI-MS/MS
    Susmitha, Kondam
    Menaka, Muthukkarupan
    INDIAN JOURNAL OF PHARMACEUTICAL EDUCATION AND RESEARCH, 2022, 56 (03) : S599 - S604
  • [2] Development and Validation of LC-ESI-MS Method for the Quantification of Dapoxetine in Rat Plasma
    Wang, Zhiyi
    Wang, Zhibin
    Chen, Chan
    Wang, Jinjin
    Luo, Shunbin
    Ma, Jianshe
    LATIN AMERICAN JOURNAL OF PHARMACY, 2012, 31 (01): : 5 - 9
  • [3] An LC-MS/MS Method Development and Validation for the Quantification of Abametapir in Plasma Samples
    Sankar, Palakollu D. S.
    Panigrahi, Naresh
    SEPARATION SCIENCE PLUS, 2025, 8 (01):
  • [4] Development and validation of a rapid LC-ESI-MS/MS method for quantification of fluoxetine and its application to MS binding assays
    Marielle Hess
    Georg Höfner
    Klaus Theodor Wanner
    Analytical and Bioanalytical Chemistry, 2011, 400 : 3505 - 3515
  • [5] Development and validation of a rapid LC-ESI-MS/MS method for quantification of fluoxetine and its application to MS binding assays
    Hess, Marielle
    Hoefner, Georg
    Wanner, Klaus Theodor
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (10) : 3505 - 3515
  • [6] Development and Validation of LC-ESI-MS/MS Method for the Determination of Alfuzosin in Human Plasma
    Rania M. Moustapha E. Moustapha
    Mehnaz El Gamal
    Journal of Analytical Chemistry, 2021, 76 : 1327 - 1335
  • [7] Development and Validation of LC-ESI-MS/MS Method for the Determination of Alfuzosin in Human Plasma
    Moustapha, Moustapha E.
    El Gamal, Rania M.
    Kamal, Mehnaz
    JOURNAL OF ANALYTICAL CHEMISTRY, 2021, 76 (11) : 1327 - 1335
  • [8] LC-ESI-MS/MS method for the simultaneous determination of hydroxysterols and bile acids in different biological matrices
    Rohn, Sascha
    Wegner, Katrin
    Werner, Philipp
    John, Clara
    Worthmann, Anna
    Scheja, Ludger
    Heeren, Joerg
    Fischer, Markus
    ANNALS OF NUTRITION AND METABOLISM, 2015, 67 : 230 - 230
  • [9] Development and validation of an LC-ESI-MS/MS method for the triple reuptake inhibitor indatraline enabling its quantification in MS Binding Assays
    Grimm, Stefanie H.
    Hoefner, Georg
    Wanner, Klaus T.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (02) : 471 - 485
  • [10] Development and validation of an LC-ESI-MS/MS method for the triple reuptake inhibitor indatraline enabling its quantification in MS Binding Assays
    Stefanie H. Grimm
    Georg Höfner
    Klaus T. Wanner
    Analytical and Bioanalytical Chemistry, 2015, 407 : 471 - 485