Comparing a Head-Mounted Smartphone Visual Field Analyzer to Standard Automated Perimetry in Glaucoma: A Prospective Study

被引:0
作者
Wang, Sean K. [1 ]
Tran, Elaine M. [1 ]
Yan, William [2 ]
Kosaraju, Reshma [3 ]
Sun, Yang [1 ]
Chang, Robert T. [1 ]
机构
[1] Stanford Univ, Byers Eye Inst, Dept Ophthalmol, Palo Alto, CA USA
[2] Royal Victorian Eye & Ear Hosp, Dept Ophthalmol, East Melbourne, Vic, Australia
[3] Harker Sch, San Jose, CA USA
关键词
glaucoma; visual field; perimetry; smartphone; head-mounted;
D O I
10.1097/IJG.0000000000002452
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Precis: Wang et al compare an FDA-registered head-mounted smartphone device (PalmScan VF2000) with standard automated perimetry (SAP) in glaucoma patients and find that the head-mounted device may not fully recapitulate SAP testing. Purpose: This study prospectively compared visual field testing using the PalmScan VF2000 Visual Field Analyzer, a head-mounted smartphone device, with standard automated perimetry (SAP). Methods: Patients with glaucoma undergoing Humphrey Field Analyzer SAP testing were asked to complete in-office PalmScan testing using a Samsung S5 smartphone in a virtual reality-style headset. Glaucoma severity was defined as SAP mean deviation (MD) >-6 dB for mild, between -6 and -12 dB for moderate, and <-12 dB for severe. Global parameters MD and pattern SD from PalmScan and SAP were compared using t-tests and Bland-Altman analyses. Bland-Altmann analyses of PalmScan and SAP MD were conducted for the superonasal, superotemporal, inferonasal, and inferotemporal visual field quadrants. The repeatability of PalmScan was assessed using Spearman's correlations and intraclass correlation coefficients (ICCs). Results: Fifty-one patients (51 eyes) completed both SAP and PalmScan testing and met the criteria for analysis. Compared with SAP, global MD and pattern SD measurements from PalmScan differed by an average of +0.62 +/- 0.26 dB (range: -3.25 to +4.60 dB) and -1.00 +/- 0.24 dB (range: -6.03 to +2.77 dB), respectively, while MD scores from individual visual field quadrants differed by as much as -6.58 to +11.43 dB. The agreement between PalmScan and SAP in classifying glaucoma severity was 86.3% across all eyes. PalmScan and SAP identified the same quadrant as having the worst visual field defect in 66.7% of eyes. Conclusions: Despite advantages in cost and accessibility, the PalmScan head-mounted perimetry device may not be able to fully recapitulate SAP testing.
引用
收藏
页码:742 / 747
页数:6
相关论文
共 50 条
  • [21] The Groningen Longitudinal Glaucoma Study. II. A prospective comparison of frequency doubling perimetry, the GDx nerve fibre analyser and standard automated perimetry in glaucoma suspect patients
    Jansonius, Nomdo M.
    Heeg, Govert P.
    ACTA OPHTHALMOLOGICA, 2009, 87 (04) : 429 - 432
  • [22] Study of anxiety in patients with glaucoma undergoing standard automated perimetry and optical coherence tomography - A prospective comparative study
    Kaliaperumal, Subashini
    Janani, V. S.
    Menon, Vikas
    Sarkar, Sandip
    Behera, Geeta
    Kattamani, Shivanand
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2022, 70 (08) : 2883 - 2887
  • [23] Standard automated perimetry for glaucoma and diseases of the retina and visual pathways: Current and future perspectives
    Phu, Jack
    Khuu, Sieu K.
    Nivison-Smith, Lisa
    Kalloniatis, Michael
    PROGRESS IN RETINAL AND EYE RESEARCH, 2025, 104
  • [24] COMPARISON OF THE OCULOKINETIC PERIMETRY GLAUCOMA SCREENER WITH 2 TYPES OF VISUAL-FIELD ANALYZER
    GREVE, M
    CHISHOLM, IA
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 1993, 28 (05): : 201 - 206
  • [25] Sensitivity of size I stimulus in achromatic automated perimetry for detection of glaucomatous visual field defects: a comparative analysis with short wavelength automated perimetry and standard automated perimetry (SITA)
    Lemos Mattos, Tessa de Cerqueira
    Kasahara, Niro
    Della Paolera, Mauricio
    Cohen, Ralph
    Mandia Junior, Carmo
    de Almeida, Geraldo Vicente
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2008, 71 (02) : 142 - 148
  • [26] Low-cost, smartphone-based frequency doubling technology visual field testing using a head-mounted display
    Alawa, Karam AlRahman
    Nolan, Ryan P.
    Han, Elaine
    Arboleda, Alejandro
    Durkee, Heather
    Sayed, Mohamed S.
    Aguilar, Mariela C.
    Lee, Richard K.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2021, 105 (03) : 440 - 444
  • [27] Comparison of Visual Field Test Measurements With a Novel Approach on a Wearable Headset to Standard Automated Perimetry
    Johnson, Catherine
    Sayed, Ahmed
    McSoley, John
    Durbin, Mary
    Kashem, Rashed
    Nicklin, Alexandra
    Lopez, Valeria
    Mijares, Georgeana
    Chen, Michael
    Shaheen, Abdulla
    Segarra, Steven
    Rady, Nadine
    Duque, Christian Andres
    Opoku-Baah, Collins
    Abou Shousha, Mohamed
    JOURNAL OF GLAUCOMA, 2023, 32 (08) : 647 - 657
  • [28] The psychometric function and reaction times of automated perimetry in normal and abnormal areas of the visual field in patients with glaucoma
    Wall, M
    Maw, RJ
    Stanek, KE
    Chauhan, BC
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (05) : 878 - 885
  • [29] Prediction of visual field defects on standard automated perimetry by screening C-20-1 frequency doubling technology perimetry
    Kamantigue, MEG
    Joson, PJ
    Chen, PP
    JOURNAL OF GLAUCOMA, 2006, 15 (01) : 35 - 39
  • [30] COURSE OF VISUAL-FIELD LOSS ON AUTOMATED PERIMETRY IN CHRONIC GLAUCOMA STABILIZED BY TRABECULECTOMY
    BOISSONNOT, M
    HIRSCH, AL
    ROVIRA, JC
    RISSE, JF
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 1992, 15 (12): : 645 - 650