A feasibility study to predict 3D dose delivery accuracy for IMRT using DenseNet with log files

被引:0
|
作者
Huang, Ying [1 ,2 ,3 ]
Cai, Ruxin [3 ]
Pi, Yifei [4 ]
Ma, Kui [5 ]
Kong, Qing [1 ]
Zhuo, Weihai [2 ]
Kong, Yan [6 ]
机构
[1] Fudan Univ, Inst Modern Phys, Shanghai, Peoples R China
[2] Fudan Univ, Inst Radiat Med, Shanghai 200032, Peoples R China
[3] Shanghai Jiao Tong Univ, Shanghai Chest Hosp, Sch Med, Shanghai, Peoples R China
[4] hengzhou Univ, Affiliated Hosp 1, Dept Radiat Oncol, Zhengzhou, Henan, Peoples R China
[5] Varian Med Syst, Beijing, Peoples R China
[6] Jiangnan Univ, Dept Radiat Oncol, Affiliated Hosp, Wuxi 214122, Jiangsu, Peoples R China
基金
中国国家自然科学基金;
关键词
Three-dimensional gamma prediction; log files; patient-specific quality assurance; MLC POSITIONAL ERRORS; CLINICAL IMPLEMENTATION; QUALITY-ASSURANCE; SLIDING WINDOW; RADIOTHERAPY; VERIFICATION; EXPERIENCE;
D O I
10.3233/XST-230412
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
OBJECTIVE: This study aims to explore the feasibility of DenseNet in the establishment of a three-dimensional (3D) gamma prediction model of IMRT based on the actual parameters recorded in the log files during delivery. METHODS: A total of 55 IMRT plans (including 367 fields) were randomly selected. The gamma analysis was performed using gamma criteria of 3% /3 mm (Dose Difference/Distance to Agreement), 3% /2 mm, 2% /3 mm, and 2% /2 mm with a 10% dose threshold. In addition, the log files that recorded the gantry angle, monitor units (MU), multi-leaf collimator (MLC), and jaws position during delivery were collected. These log files were then converted to MU-weighted fluence maps as the input of DenseNet, gamma passing rates (GPRs) under four different gamma criteria as the output, and mean square errors (MSEs) as the loss function of this model. RESULTS: Under different gamma criteria, the accuracy of a 3D GPR prediction model decreased with the implementation of stricter gamma criteria. In the test set, the mean absolute error (MAE) of the prediction model under the gamma criteria of 3% /3 mm, 2% /3 mm, 3% /2 mm, and 2% /2 mm was 1.41, 1.44, 3.29, and 3.54, respectively; the root mean square error (RMSE) was 1.91, 1.85, 4.27, and 4.40, respectively; the Sr was 0.487, 0.554, 0.573, and 0.506, respectively. There was a correlation between predicted and measured GPRs (P < 0.01). Additionally, there was no significant difference in the accuracy between the validation set and the test set. The accuracy in the high GPR group was high, and the MAE in the high GPR group was smaller than that in the low GPR group under four different gamma criteria. CONCLUSIONS: In this study, a 3D GPR prediction model of patient-specific QA using DenseNet was established based on log files. As an auxiliary tool for 3D dose verification in IMRT, this model is expected to improve the accuracy and efficiency of dose validation.
引用
收藏
页码:1199 / 1208
页数:10
相关论文
共 50 条
  • [31] Feasibility study of skin dosimetry with TLD sheets for measuring the effect of 3D printed bolus in radiotherapy
    Miyasaka, Yuya
    Ichikawa, Mayumi
    Akira, Takagi
    Yamazawa, Yoshifumi
    Chai, Hongbo
    Souda, Hikaru
    Ishizawa, Miyu
    Sato, Hiraku
    Iwai, Takeo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [32] The influence of respiratory motion on dose distribution of 3D-CRT and IMRT- A simulation study
    Chang-li, R.
    Yu-xin, C.
    Lu-zhou, W.
    Wu-bing
    Qi-bin, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2015, 13 (01): : 39 - 43
  • [33] Feasibility and accuracy of 3D printed patient-specific skull contoured brain biopsy guides
    Shinn, Richard
    Park, Clair
    DeBose, Kyrille
    Hsu, Fang-Chi
    Cecere, Thomas
    Rossmeisl, John
    VETERINARY SURGERY, 2021, 50 (05) : 933 - 943
  • [34] Error detection during VMAT delivery using EPID-based 3D transit dosimetry
    Mijnheer, B.
    Jomehzadeh, A.
    Gonzalez, P.
    Olaciregui-Ruiz, I
    Rozendaal, R.
    Shokrani, P.
    Spreeuw, H.
    Tielenburg, R.
    Mans, A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 54 : 137 - 145
  • [35] A dosimetric phantom study of dose accuracy and build-up effects using IMRT and RapidArc in stereotactic irradiation of lung tumours
    Seppala, Jan
    Suilamo, Sami
    Kulmala, Jarmo
    Mali, Pekka
    Minn, Heikki
    RADIATION ONCOLOGY, 2012, 7
  • [36] Dosiomics: Extracting 3D Spatial Features From Dose Distribution to Predict Incidence of Radiation Pneumonitis
    Liang, Bin
    Yan, Hui
    Tian, Yuan
    Chen, Xinyuan
    Yan, Lingling
    Zhang, Tao
    Zhou, Zongmei
    Wang, Lvhua
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2019, 9
  • [37] A novel mathematical model to generate semi-automated optimal IMRT treatment plan based on predicted 3D dose distribution and prescribed dose
    Yousefi, Ali
    Ketabi, Saeedeh
    Abedi, Iraj
    MEDICAL PHYSICS, 2023, 50 (05) : 3148 - 3158
  • [38] Sigmoid dose using 3D imaging in cervical-cancer brachytherapy
    Holloway, Caroline L.
    Racine, Marie-Lynn
    Cormack, Robert A.
    O'Farrell, Desmond A.
    Viswanathan, Akila N.
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) : 307 - 310
  • [39] Feasibility study of the Vitom 3D exoscope in ENT surgery using an animal model: relevance to coronavirus disease 2019
    Yeung, W.
    Vakharia, N.
    Johnatty, C.
    Achar, P.
    Kasbekar, A., V
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2021, 135 (08) : 710 - 717
  • [40] Evaluation of an Electromagnetic 3D Navigation System to Facilitate Endovascular Tasks: A Feasibility Study
    Sidhu, R.
    Weir-McCall, J.
    Cochennec, F.
    Riga, C.
    DiMarco, A.
    Bicknell, C. D.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2012, 43 (01) : 22 - 29