Digital analysis of fabrication accuracy and fit in additively and subtractively manufactured implant-supported fixed complete dentures

被引:1
作者
Maldonado, Paola [1 ]
Donmez, Mustafa Borga [1 ,2 ]
Guven, Mehmet Esad [3 ]
Schimmel, Martin [1 ,4 ]
Revilla-Leon, Marta [5 ,6 ,7 ]
Cakmak, Gulce [1 ]
Yilmaz, Burak [1 ,8 ,9 ]
机构
[1] Univ Bern, Sch Dent Med, Dept Reconstruct Dent & Gerodontol, Freiburgstr, CH-73007 Bern, Switzerland
[2] Biruni Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkiye
[3] Necmettin Erbakan Univ, Fac Dent, Dept Prosthodont, Konya, Turkiye
[4] Univ Clin Geneva, Div Gerodontol & Removable Prosthodont, Geneva, Switzerland
[5] Tufts Univ, Sch Dent Med, Dept Prosthodont, Boston, MA USA
[6] Univ Washington, Sch Dent, Dept Restorat Dent, Seattle, WA USA
[7] Kois Ctr, Res & Digital Dent, Seattle, WA USA
[8] Univ Bern, Sch Dent Med, Dept Restorat Prevent & Pediat Dent, Bern, Switzerland
[9] Ohio State Univ, Coll Dent, Div Restorat & Prosthet Dent, Columbus, OH USA
关键词
Additive manufacturing; Fixed complete denture; Fit; Resin; Strength gradient zirconia; Trueness; PROSTHESES; TITANIUM;
D O I
10.1016/j.jdent.2024.105332
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To digitally evaluate the trueness and fit of additively and subtractively manufactured fixed complete dentures in materials intended for definitive use. Methods: An edentulous maxillary model with implants at the left first molar, left canine, right canine, and right first molar site was digitized and a fixed complete denture was designed. This design was used to fabricate fixed dentures in an additively manufactured resin for definitive use (AM), a high-impact polymer composite (SM-CR), and a strength gradient zirconia (SM-ZR) (n = 10). Each fixed denture was digitized and the surface (overall, occlusal, except occlusal, and abutments), linear, and interimplant distance deviations were analyzed. The fit was assessed with the triple-scan protocol. Data were analyzed with Welch analysis of variance and Games-Howell tests (alpha = 0.05). Results: SM-ZR led to lower overall deviations than AM, which had the highest occlusal and the lowest abutments deviations (P <= 0.007). SM-ZR had the lowest occlusal and SM-CR had the highest except occlusal deviations (P <= 0.002). AM mostly had higher linear and SM-CR mostly had higher interimplant distance deviations (P <= 0.043). AM led to the highest marginal gap at the left canine site, while SM-CR had the highest and SM-ZR had the lowest gaps at the right canine site (P <= 0.022). Conclusions: SM-ZR dentures mostly had trueness and marginal fit similar to or better than the other groups. Tested fixed complete dentures were mostly smaller than the design file in terms of interimplant distances.
引用
收藏
页数:7
相关论文
共 36 条
  • [1] Magnitude of misfit threshold in implant-supported restorations: A systematic review
    Abdelrehim, Aly
    Etajuri, Enas Abdalla
    Sulaiman, Eshamsul
    Sofian, Hazrina
    Salleh, Nosizana Mohd
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2024, 132 (03) : 528 - 535
  • [2] Abduo J, 2010, INT J ORAL MAX IMPL, V25, P506
  • [3] Trueness and marginal fit of implant-supported complete-arch fixed prosthesis frameworks made of high-performance polymers and titanium: An explorative in-vitro study
    Abou-Ayash, Samir
    Schimmel, Martin
    Ozcan, Mutlu
    Ozcelik, Burak
    Bragger, Urs
    Yilmaz, Burak
    [J]. JOURNAL OF DENTISTRY, 2021, 113
  • [4] In vitro fit of CAD-CAM complete arch screw-retained titanium and zirconia implant prostheses fabricated on 4 implants
    AL-Meraikhi, Hadi
    Yilmaz, Burak
    McGlumphy, Edwin
    Brantley, William
    Johnston, William M.
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (03) : 409 - 416
  • [5] AL-Meraikhi H, 2018, J PROSTHET DENT, V119, P116, DOI 10.1016/j.prosdent.2017.02.003
  • [6] The effect of framework fabrication technique on the fit accuracy of full arch screw retained implant supported prostheses
    AlRasheed, Fellwa
    AlWazzan, Khalid
    [J]. SAUDI DENTAL JOURNAL, 2022, 34 (04) : 288 - 297
  • [7] The influence of prosthetic material on implant and prosthetic survival of implant-supported fixed complete dentures: a systematic review and meta-analysis
    Bagegni, Aimen
    Abou-Ayash, Samir
    Ruecker, Gerta
    Algarny, Ahmad
    Att, Wael
    [J]. JOURNAL OF PROSTHODONTIC RESEARCH, 2019, 63 (03) : 251 - 265
  • [8] Biomechanical behavior of CAD/CAM cobalt-chromium and zirconia full-arch fixed prostheses
    Barbin, Thais
    Silva, Leticia Del Rio
    Veloso, Daniele Valente
    Borges, Guilherme Almeida
    Camacho Presotto, Anna Gabriella
    Ricardo Barao, Valentim Adelino
    Groppo, Francisco Carlos
    Mesquita, Marcelo Ferraz
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2020, 12 (06) : 329 - 337
  • [9] Load to Failure of High-Density Polymers for Implant-Supported Fixed Cantilevered Prostheses with Titanium Bases
    Batak, Burcu
    Cakmak, Guelce
    Seidt, Jeremy
    Yilmaz, Burak
    [J]. INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2021, 34 (05) : 608 - 614
  • [10] Trueness of crowns fabricated by using additively and subtractively manufactured resin-based CAD-CAM materials
    Cakmak, Guelce
    Rusa, Ana Maria
    Donmez, Mustafa Borga
    Akay, Canan
    Kahveci, Cigdem
    Schimmel, Martin
    Yilmaz, Burak
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (05) : 951 - 958