New insight for simultaneous detection of methocarbamol and paracetamol in human serum samples based on poly(L-methionine)/graphene-graphene oxide nanocomposite

被引:4
作者
Ahmed, Yousef M. [1 ]
El-Zanaty, Mohamed R. [1 ]
Galal, Ahmed [1 ]
Atta, Nada F. [1 ]
机构
[1] Cairo Univ, Fac Sci, Chem Dept, Giza 12613, Egypt
关键词
Methocarbamol; Paracetamol; Poly (L-methionine); Graphene oxide; Reduced graphene oxide; Serum samples; CARBON-PASTE ELECTRODE; SKELETAL-MUSCLE RELAXANTS; HUMAN PLASMA; GRAPHENE OXIDE; SPECTROFLUOROMETRIC DETERMINATION; VOLTAMMETRIC DETERMINATION; ELECTROCHEMICAL SENSOR; ACETAMINOPHEN; TABLETS; HPLC;
D O I
10.1016/j.microc.2024.111370
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new insight is presented for simultaneous detection of methocarbamol (MET), and paracetamol (PAR), in human serum samples. The design of an electrochemical sensor is achieved by modifying the surface of a glassy carbon electrode (GCE) with two successive thin films from conductive poly(L-methionine) (PMT) and a mixture of reduced graphene oxide and graphene oxide (RGO-GO). The composite is characterized by scanning electron microscopy (SEM), atomic force microscope (AFM), and Fourier transform infrared spectroscopy (FT-IR). The sensor creates a powerful sensing platform for MET and PAR detection with kinetically enhanced oxidation process. An increase in the anodic current responses for MET and PAR is obtained with 10.1 and 5.9 times higher compared to bare GC electrode, respectively. Mixing GO with RGO increases the surface area, and the reactive sites due to presence of oxygen-functional groups at its surface. The sensor has good stability, repeatability, reproducibility, and anti-interference capability. The detection limits are 5.2 nM and 1.2 nM for the determination of MET and PAR simultaneously in serum sample in the concentration range (0.04 -> 100 mu M). Large peak potential separation between the MET and PAR about 822 mV is attained. The assay of drugs in serum samples and pharmaceutical formulations is successfully achieved with good recovery values.
引用
收藏
页数:14
相关论文
共 74 条
[1]   The Fabrication of a Highly Sensitive Nano Green Carbon Paste Electrode Modified with Yttrium Doped Manganese Oxide (Mn2O3/Y2O3) for Electrochemical Determination of Marbofloxacin and Its Residues in Bovine Meat and Milk Samples [J].
Abdel-Atty, Shimaa ;
Abdel-Raoof, Ahmed M. ;
Mohamed, Tahany F. ;
Nasr, Zeinab A. ;
Mohamed, Ghada F. ;
Elgazzar, Elsayed .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2020, 167 (10)
[2]   Boosting the charge for the selective photoelectrochemical oxidation of hydroquinone in hazardous environments using a fine-tuned heterojunction catalyst [J].
Abdelkader, Mohamed G. ;
Atta, Nada F. ;
Galal, Ahmed .
JOURNAL OF MATERIALS CHEMISTRY A, 2024, 12 (16) :9606-9626
[3]  
Adarakatti PS, 2019, SPR ELECTROCHEM, P58, DOI 10.1039/9781788013895-00058
[4]   An electrochemical sensor based on clay/graphene oxide decorated on chitosan gel for the determination of flubendiamide insecticide [J].
Aghris, S. ;
Chhaibi, B. ;
Ouatmane, F. Z. ;
Alaoui, O. Tahiri ;
Laghrib, F. ;
Farahi, A. ;
Bakasse, M. ;
Loukili, A. ;
Lahrich, S. ;
EL Mhammedi, M. A. .
MATERIALS CHEMISTRY AND PHYSICS, 2023, 296
[5]   Synthesis of clay-reduced graphene oxide composite catalysts for the electrochemical detection of flubendiamide: application in food samples [J].
Aghris, S. ;
Chhaibi, B. ;
Ettadili, F. E. ;
Alaoui, O. Tahiri ;
Laghrib, F. ;
Farahi, A. ;
Bakasse, M. ;
Loukili, A. ;
Lahrich, S. ;
EL Mhammedi, M. A. .
MATERIALS TODAY CHEMISTRY, 2022, 26
[6]  
Ahammad A.J.S., 2019, Biomedical Applications of Graphene and 2D Nanomaterials, P249, DOI [10.1016/B978-0-12-815889-0.00012-X, DOI 10.1016/B978-0-12-815889-0.00012-X]
[7]  
Ahmed NR, 2019, Chem Sci Trans, V8, P237, DOI [10.7598/cst2019.1486, DOI 10.7598/CST2019.1486]
[8]   Electrochemical sensor based on PEDOT/CNTs-graphene oxide for simultaneous determination of hazardous hydroquinone, catechol, and nitrite in real water samples [J].
Ahmed, Yousef M. ;
Eldin, Mahmoud A. ;
Galal, Ahmed ;
Atta, Nada F. .
SCIENTIFIC REPORTS, 2024, 14 (01)
[9]   Development and Validation of the UV Spectrophotometric Method for Simultaneous Determination of Paracetamol and Pseudoephedrine in Bulk and Combined Tablet Dosage Form [J].
Ali, Mohamed S. ;
Elsaman, Tilal .
PHARMACEUTICAL CHEMISTRY JOURNAL, 2021, 54 (12) :1306-1310
[10]   Paracetamol (Acetaminophen): mechanisms of action [J].
Anderson, Brian J. .
PEDIATRIC ANESTHESIA, 2008, 18 (10) :915-921