Severe intraprosthetic aortic insufficiency immediately after implantation of a SAPIEN 3 Ultra RESILIA valve with spontaneous resolution: case report

被引:0
|
作者
Diaz, Juan Armando [1 ,2 ]
Kobayashi, Ken [1 ]
Hachinohe, Daisuke [1 ]
Horita, Ryo [1 ]
Shitan, Hidemasa [1 ]
机构
[1] Sapporo Cardio Vasc Clin, Sapporo Heart Ctr, Cardiovasc Med, Asia Med Grp, North 49,East 16,8-1 Higashi Ward, Sapporo 0070849, Japan
[2] Makati Med Ctr, Makati, Philippines
关键词
Aortic valve stenosis; Complication; Intravalvular aortic insufficiency; Intraprosthetic aortic insufficiency; Transcatheter aortic valve replacement; Case report; REGURGITATION; REPLACEMENT; LEAFLET;
D O I
10.1093/ehjcr/ytae402
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Aortic insufficiency (AI) is often encountered in transcatheter aortic valve replacement (TAVR) but is only rarely haemodynamically significant. Even more uncommon is the occurrence of intraprosthetic AI that often leads to haemodynamic compromise requiring additional therapeutic intervention.Case summary An 85-year-old female with severe aortic stenosis underwent elective TAVR with a size 23 mm SAPIEN 3 Ultra RESILIA (S3UR) valve. After implantation, the patient developed hypotension. Transthoracic as well as transoesophageal echocardiogram identified significant transvalvular AI. Persistence of AI led the team to consider a TAV-in-TAV strategy. Before a second valve system could be inserted, the patient's blood pressure improved and AI resolved spontaneously. The patient was discharged and recovered with no AI on follow-up imaging studies.Discussion This is the first reported case of significant intraprosthetic AI with the new S3UR in our literature search as of writing. The new S3UR valve has several improvements from previous generations designed to increase valve area and durability as well as decreasing the incidence of paravalvular and intraprosthetic leaks. A stuck leaflet was inferred as the cause of the intraprosthetic AI, and the improved design with excessive expansion may have led to this. Unlike previous case reports, intraprosthetic AI resolved without further intervention likely due to the turbulence of flow releasing the stuck leaflet. Cautious observation prevented the need for a TAV-in-TAV for a rare and possibly catastrophic complication.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Acute leaflet malfunction of Navitor valve with severe intraprosthetic aortic insufficiency immediately after implantation
    Horita, Ryo
    Hachinohe, Daisuke
    Otake, Ryo
    Diaz, Juan Armando
    Shitan, Hidemasa
    Fujita, Tsutomu
    CARDIOVASCULAR INTERVENTION AND THERAPEUTICS, 2024, 39 (03) : 331 - 332
  • [2] Transcatheter aortic valve implantation for structural valve deterioration of homograft surgical aortic valve using SAPIEN3 Ultra RESILIA: a case report
    Mizutani, Kazuki
    Ueno, Masafumi
    Sakaguchi, Genichi
    Nakazawa, Gaku
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2024, 8 (03)
  • [3] A stuck leaflet after balloon post-dilatation in transcatheter aortic valve implantation with a SAPIEN-3 ultra RESILIA valve: a case report
    Yamazoe, Shinji
    Ogawa, Yasuhiro
    Kano, Naoaki
    Mamiya, Keita
    Kawaguchi, Katsuhiro
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2025, 9 (01)
  • [4] Case report describing balloon dilatation as a treatment for acute intraprosthetic regurgitation immediately following implantation of a SAPIEN-3 transcatheter aortic valve
    Pal, Nikhil
    Chew, Derek
    Sinhal, Ajay
    CLINICAL CASE REPORTS, 2021, 9 (06):
  • [5] Severe intraprosthetic regurgitation and valve embolization after transcatheter aortic valve implantation
    Ben-Dor, Itsik
    Satler, Lowell F.
    Pichard, Augusto D.
    Waksman, Ron
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (02) : 392 - 396
  • [6] Acute Severe Mitral Stenosis Immediately After Transcatheter Aortic Valve Implantation
    Auer, Johann
    Grund, Michael
    Puschmann, Rudolf
    Berent, Robert
    JOURNAL OF INVASIVE CARDIOLOGY, 2017, 29 (10) : E154 - E154
  • [7] The feasibility of transcatheter aortic valve implantation using the Edwards SAPIEN 3 for patients with severe bicuspid aortic stenosis
    Arai, Takahide
    Lefevre, Thierry
    Hovasse, Thomas
    Morice, Marie-Claude
    Romano, Mauro
    Benamer, Hakim
    Garot, Philippe
    Hayashida, Kentaro
    Bouvier, Erik
    Chevalier, Bernard
    JOURNAL OF CARDIOLOGY, 2017, 70 (3-4) : 220 - 224
  • [8] Conduction disorders after transcatheter aortic valve implantation: A comparison between SAPIEN 3 and SAPIEN 3 Ultra balloon-expandable valves
    Monizzi, Giovanni
    Olivares, Paolo
    Makmur, Giulio
    Fabbiocchi, Franco
    Grancini, Luca
    Mastrangelo, Angelo
    Ferrari, Cristina
    Galli, Stefano
    Montorsi, Piero
    Bartorelli, Antonio L. L.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [9] Transfemoral aortic valve implantation is more successful with the Edwards Sapien 3 compared with the Edwards XT for the treatment of symptomatic severe aortic stenosis
    Gonska, Birgid
    Seeger, Julia
    Junker, Alexander
    Rodewald, Christoph
    Trepte, Ulrike
    Scharnbeck, Dominik
    Rottbauer, Wolfgang
    Woehrle, Jochen
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2018, 111 (8-9) : 470 - 479
  • [10] Changes in the Pacemaker Rate After Transition From Edwards SAPIEN XT to SAPIEN 3 Transcatheter Aortic Valve Implantation The Critical Role of Valve Implantation Height
    De Torres-Alba, Fernando
    Kaleschke, Gerrit
    Diller, Gerhard Paul
    Vormbrock, Julia
    Orwat, Stefan
    Radke, Robert
    Reinke, Florian
    Fischer, Dieter
    Reinecke, Holger
    Baumgartner, Helmut
    JACC-CARDIOVASCULAR INTERVENTIONS, 2016, 9 (08) : 805 - 813