Background: Endometrial cancer is one of the most common gynecological malignancies in the world. Vaginal brachytherapy is an important postoperative adjuvant treatment for endometrial cancer. However, a common problem with existing applicators is insufficient dose at the vaginal apex. Purpose: This study describes the Hangzhou (HZ) cylinder, a novel 3D printed vaginal intracavity brachytherapy applicator, detailing its characteristics, dose distribution, and clinical applications. Methods and materials: The HZ cylinder is distinguished by its unique structure: a U-shaped channel with a 2 mm diameter, a straight central axis channel of the same diameter, and 10 parallel straight channels. For comparison, standard plans were employed, designed to ensure that a minimum of 95% of the prescribed dose reached 5 mm beneath the mucosal surface. We conducted comparative analyses of mucosal surface doses and doses at a 5 mm depth below the mucosa between the HZ cylinder and a conventional single-channel cylinder across various treatment schemes. Additionally, the study examined dose differences in target volume and organs at risk (OARs) between actual HZ cylinder plans and hypothetical single-channel plans. Results: In the standard plans, mucosal surface doses at the apex of the vagina were 209.32% and 200.61% of the prescribed dose with the HZ and single-channel cylinders, respectively. The doses on the left and right wall mucosal surfaces varied from 149.26% to 178.13% and 142.98% to 180.75% of the prescribed dose, and on the anterior and posterior wall mucosal surfaces varied from 128.87% to 138.50% and 142.98% to 180.75% of the prescribed dose. Analysis of 24 actual treatment plans revealed that when the vaginal tissue volume dose covering 98% (vaginal D98%) was comparable between the HZ cylinder and virtual single-channel plans (6.74 +/- 0.07 Gy vs. 6.69 +/- 0.10 Gy, p = 0.24), rectum doses of HZ cylinder plans were significantly lower than those of single-channel plans (D1cc, 5.96 +/- 0.56 Gy vs. 6.26 +/- 0.71 Gy, p = 0.02 and D2cc, 5.26 +/- 0.52 Gy vs. 5.56 +/- 0.62 Gy, p = 0.02). Conclusions: The HZ cylinder demonstrates a reduction in dose to the rectum and bladder while maintaining adequate target volume coverage. Its mucosal surface dose is comparable to that of the traditional single-channel cylinder. These findings suggest that the HZ cylinder is a viable and potentially safer alternative for vaginal brachytherapy, warranting further investigation with larger sample sizes.