A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans

被引:0
作者
Abdulkadir, Muhammad Kabir [1 ,3 ]
Osman, Noor Diyana [1 ]
Achuthan, Anusha [2 ]
Nasirudin, Radin A. [1 ]
Ahmad, Muhammad Zabidi [1 ]
Zain, Noor Hasyima Mat [1 ]
Shuaib, Ibrahim Lutfi [1 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Dept Biomed Imaging, George Town, Malaysia
[2] Univ Sains Malaysia, Sch Comp Sci, Dept Comp Sci, George Town, Malaysia
[3] Univ Ilorin, Fac Basic Clin Sci, Dept Radiog, Ilorin 240213, Nigeria
关键词
Algorithm; automation; body size; computed tomography scan; image segmentation; pediatric; radiation dosage; size-specific dose estimate; CT; FEASIBILITY; VARIABILITY; DIAMETER; CHEST;
D O I
10.4103/jmp.jmp_26_24
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background and Purpose: Size-specific dose estimates (SSDE) have been introduced into computed tomography (CT) dosimetry to tailor patients' unique sizes to facilitate accurate CT radiation dose quantification and optimization. The purpose of this study was to develop and validate an automated algorithm for the determination of patient size (effective diameter) and SSDE. Materials and Methods: A MATLAB platform was used to develop software of algorithms based on image segmentation techniques to automate the calculation of patient size and SSDE. The algorithm was used to automatically estimate the individual size and SSDE of four CT dose index phantoms and 80 CT images of pediatric patients comprising head, thorax, and abdomen scans. For validation, the American Association of Physicists in Medicine (AAPM) manual methods were used to determine the patient's size and SSDE for the same subjects. The accuracy of the proposed algorithm in size and SSDE calculation was evaluated for agreement with the AAPM's estimations (manual) using Bland-Altman's agreement and Pearson's correlation coefficient. The normalized error, system bias, and limits of agreement (LOA) between methods were derived. Results: The results demonstrated good agreement and accuracy between the automated and AAPM's patient size estimations with an error rate of 1.9% and 0.27% on the patient and phantoms study, respectively. A 1% percentage difference was found between the automated and manual (AAPM) SSDE estimates. A strong degree of correlation was seen with a narrow LOA between methods for clinical study (r > 0.9771) and phantom study (r > 0.9999). Conclusion: The proposed automated algorithm provides an accurate estimation of patient size and SSDE with negligible error after validation.
引用
收藏
页码:456 / 463
页数:8
相关论文
共 26 条
[1]   Comparison between Deff and Dw approaches for estimation of size-specific dose in paediatric CT imaging [J].
Abdulkadir, Muhammad Kabir ;
Shuaib, Ibrahim Lutfi ;
Nasirudin, R. A. ;
Samsudin, Ahmad Hadif Zaidin ;
Osman, Noor Diyana .
RADIATION PHYSICS AND CHEMISTRY, 2022, 198
[2]  
Almohiy H, 2014, WORLD J RADIOL, V6, P1, DOI 10.4329/wjr.v6.i1.1
[3]   Automated Calculation of Water-equivalent Diameter (DW) Based on AAPM Task Group 220 [J].
Anam, Choirul ;
Haryanto, Freddy ;
Widita, Rena ;
Arif, Idam ;
Dougherty, Geoff .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04) :320-333
[4]  
[Anonymous], NCRP Report 160
[5]  
Boone JM, 2019, AAPM Report No.293
[6]  
Boone JM., 2011, Report No. 204: Size-Specific Dose Estimates in Pediatric and Adult Body CT Examinations
[7]   Attenuation-based size metric for estimating organ dose to patients undergoing tube current modulated CT exams [J].
Bostani, Maryam ;
McMillan, Kyle ;
Lu, Peiyun ;
Kim, Hyun J. ;
Cagnon, Chris H. ;
DeMarco, John J. ;
McNitt-Gray, Michael F. .
MEDICAL PHYSICS, 2015, 42 (02) :958-968
[8]   Automated Estimation of Abdominal Effective Diameter for Body Size Normalization of CT Dose [J].
Cheng, Phillip M. .
JOURNAL OF DIGITAL IMAGING, 2013, 26 (03) :406-411
[9]   Evaluation of radiation dose accuracy calculated using IndoseCT software with direct measurement on polyester-resin phantoms [J].
Dio, Putryani ;
Anam, Choirul ;
Hidayanto, Eko ;
Sutanto, Heri ;
Asiah, Rin H. ;
Fatichah, Lintang N. ;
Rukmana, Dito A. ;
Dougherty, Geoff .
RADIATION PHYSICS AND CHEMISTRY, 2022, 201
[10]   Estimating the Patient-specific Dose to the Thyroid and Breasts and Overall Risk in Chest CT When Using Organ-based Tube Current Modulation [J].
Franck, Caro ;
Smeets, Peter ;
Lapeire, Lore ;
Achten, Eric ;
Bacher, Klaus .
RADIOLOGY, 2018, 288 (01) :164-169